Mechanical cavopulmonary assistance of a patient-specific Fontan physiology: numerical simulations, lumped parameter modeling, and suction experiments.

This study investigated the performance of a magnetically levitated, intravascular axial flow blood pump for mechanical circulatory support of the thousands of Fontan patients in desperate need of a therapeutic alternative. Four models of the extracardiac, total cavopulmonary connection (TCPC) Fontan configuration were evaluated to formulate numerical predictions: an idealized TCPC, a patient-specific TCPC per magnetic resonance imaging data, and each of these two models having a blood pump in the inferior vena cava (IVC). A lumped parameter model of the Fontan physiology was used to specify boundary conditions. Pressure-flow characteristics, energy gain calculations, scalar stress levels, and blood damage estimations were executed for each model. Suction limitation experiments using the Sylgard elastomer tubing were also conducted. The pump produced pressures of 1-16 mm Hg for 2000-6000 rpm and flow rates of 0.5-4.5 L/min. The pump inlet or IVC pressure was found to decrease at higher rotational speeds. Maximum scalar stress estimations were 3 Pa for the nonpump models and 290 Pa for the pump-supported cases. The blood residence times for the pump-supported cases were shorter (0.9 s) as compared with the nonsupported configurations (2.5 s). However, the blood damage indices were higher (1.5%) for the anatomic model with pump support. The pump successfully augmented pressure in the TCPC junction and increased the hydraulic energy of the TCPC as a function of flow rate and rotational speed. The suction experiments revealed minimal deformation (<3%) at 9000 rpm. The findings of this study support the continued design and development of this blood pump.

[1]  Yi Wu,et al.  Effects of Muscle Pump on Rotary Blood Pumps in Dynamic Exercise: A Computer Simulation Study , 2008, Cardiovascular engineering.

[2]  D. Barron,et al.  Factors influencing early and late outcome following the Fontan procedure in the current era. The 'Two Commandments'? , 2007, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[3]  James F. Antaki,et al.  Quantitative Evaluation of Blood Damage in a Centrifugal VAD by Computational Fluid Dynamics , 2004 .

[4]  Giancarlo Pennati,et al.  Management of a stenotic right ventricle-pulmonary artery shunt early after the Norwood procedure. , 2009, The Annals of thoracic surgery.

[5]  J. Deanfield,et al.  Four decades of Fontan palliation , 2010, Nature Reviews Cardiology.

[6]  A. Throckmorton,et al.  Pediatric Circulatory Support: Current Strategies and Future Directions. Biventricular and Univentricular Mechanical Assistance , 2008, ASAIO journal.

[7]  R. Opitz,et al.  A Couette viscometer for short time shearing of blood. , 1980, Biorheology.

[8]  R Pietrabissa,et al.  A mathematical model of circulation in the presence of the bidirectional cavopulmonary anastomosis in children with a univentricular heart. , 1997, Medical engineering & physics.

[9]  Xinwei Song,et al.  Computational fluid dynamics prediction of blood damage in a centrifugal pump. , 2003, Artificial organs.

[10]  A. Throckmorton,et al.  Mechanical axial flow blood pump to support cavopulmonary circulation , 2008, The International journal of artificial organs.

[11]  R. Paul,et al.  Shear stress related blood damage in laminar couette flow. , 2003, Artificial organs.

[12]  Amy L Throckmorton,et al.  Intravascular mechanical cavopulmonary assistance for patients with failing Fontan physiology. , 2009, Artificial organs.

[13]  Alexandrina Untaroiu,et al.  Numerical Design and Experimental Hydraulic Testing of an Axial Flow Ventricular Assist Device for Infants and Children , 2007, ASAIO journal.

[14]  R Fumero,et al.  Use of mathematical model to predict hemodynamics in cavopulmonary anastomosis with persistent forward flow. , 2000, The Journal of surgical research.

[15]  Ajit P. Yoganathan,et al.  Importance of Accurate Geometry in the Study of the Total Cavopulmonary Connection: Computational Simulations and In Vitro Experiments , 2001, Annals of Biomedical Engineering.

[16]  A. Throckmorton,et al.  Interaction of an idealized cavopulmonary circulation with mechanical circulatory assist using an intravascular rotary blood pump. , 2010, Artificial organs.

[17]  C. Bertram,et al.  Transitional flow field characterization inside an arteriovenous graft-to-vein anastomosis under pulsatile flow conditions , 2005 .

[18]  Bartley P Griffith,et al.  Computational and experimental evaluation of the fluid dynamics and hemocompatibility of the CentriMag blood pump. , 2006, Artificial organs.

[19]  C Bludszuweit,et al.  Three-dimensional numerical prediction of stress loading of blood particles in a centrifugal pump. , 1995, Artificial organs.

[20]  Amy L Throckmorton,et al.  Design of a protective cage for an intravascular axial flow blood pump to mechanically assist the failing Fontan. , 2009, Artificial organs.

[21]  F. Migliavacca,et al.  Multiscale modeling of the cardiovascular system: application to the study of pulmonary and coronary perfusions in the univentricular circulation. , 2005, Journal of biomechanics.

[22]  F. Migliavacca,et al.  Modeling of the Norwood circulation: effects of shunt size, vascular resistances, and heart rate. , 2001, American journal of physiology. Heart and circulatory physiology.

[23]  Christopher M. Haggerty,et al.  Numerical, Hydraulic, and Hemolytic Evaluation of an Intravascular Axial Flow Blood Pump to Mechanically Support Fontan Patients , 2010, Annals of Biomedical Engineering.