Tendon Transfer Fixation in the Foot and Ankle: A Biomechanical Study Evaluating Two Sizes of Pilot Holes for Bioabsorbable Screws

The purpose of this study was to compare the initial fixation strengths of bioabsorbable screws for tendon transfers in the foot and ankle when the pilot hole size varied. A 7times20 mm screw was used with 5.5 mm and 6.5 mm drill holes, and a 5times20 mm screw was used with 3.9 mm and 4.5 mm drill holes. Biomechanical testing was performed on each tendon transfer in cadaver specimens. A paired t-test showed no significant difference in pullout strength when pilot hole size varied between 79 to 93% of the screw size for the 7 mm screw and 78 to 90% of the screw size for the 5 mm screw. Previous studies have found a critical value of tendon tension equaling 50 N with passive dorsiflexion of the foot. With an average value of approximately 170 N, the 7 mm screw provided three times the requisite strength. The 5 mm screw provided 1.5 times the requisite strength, but the transfer was technically more difficult.

[1]  M. Laval-jeantet,et al.  Dual-energy X-ray absorptiometry of the calcaneus: Comparison with vertebral dual-energy X-ray absorptiometry and quantitative computed tomography , 2004, Calcified Tissue International.

[2]  C. Ambrose,et al.  Biomechanical Testing of Cadaveric Specimens: Importance of Bone Mineral Density Assessment , 2002, Foot & ankle international.

[3]  L. Paulos,et al.  Bioscrew fixation of patellar tendon autografts. , 2000, Biomaterials.

[4]  V. De Maertelaer,et al.  Quantitative ultrasound and dual X-ray absorptiometry measurements of the calcaneus in patients on maintenance hemodialysis. , 2000, Bone.

[5]  M. Walton Absorbable and metal interference screws: comparison of graft security during healing. , 1999, Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association.

[6]  R. Speller,et al.  The effect of region of interest selection on dual energy X-ray absorptiometry measurements of the calcaneus in 55 post-menopausal women. , 1999, The British journal of radiology.

[7]  W. H. Warden,et al.  Magnetic Resonance Imaging of Bioabsorbable Polylactic Acid Interference Screws During the First 2 Years After Anterior Cruciate Ligament Reconstruction , 1999 .

[8]  L. Paulos,et al.  Bioabsorbable interference screws for graft fixation in anterior cruciate ligament reconstruction. , 1999, Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association.

[9]  K. Humer,et al.  Serial magnetic resonance imaging evaluation of a bioabsorbable interference screw and the adjacent bone. , 1999, Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association.

[10]  F. Barber Tripled semitendinosus-cancellous bone anterior cruciate ligament reconstruction with bioscrew fixation. , 1999, Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association.

[11]  R Seil,et al.  Ligament graft initial fixation strength using biodegradable interference screws. , 1999, Journal of biomedical materials research.

[12]  R. Ordidge,et al.  In Vivo MRI Measurements of Bone Quality in the Calcaneus: A Comparison with DXA and Ultrasound , 1999, Osteoporosis International.

[13]  R Seil,et al.  Comparison of Initial Fixation Strength Between Biodegradable and Metallic Interference Screws and a Press-Fit Fixation Technique in a Porcine Model , 1998, The American journal of sports medicine.

[14]  Y. An,et al.  Fixation of Osteotomies Using Bioabsorbable Screws in the Canine Femur , 1998, Clinical orthopaedics and related research.

[15]  H. Nicholson,et al.  Determination of a standard site for the measurement of bone mineral density of the human calcaneus , 1998, Journal of anatomy.

[16]  J. Hui,et al.  Biomechanical Study of Tibialis Anterior Tendon Transfer , 1998, Clinical orthopaedics and related research.

[17]  W. Walsh,et al.  Initial fixation strength of polylactic acid interference screws in anterior cruciate ligament reconstruction. , 1998, Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association.

[18]  A. Weiler,et al.  Biodegradable Interference Screw Fixation Exhibits Pull-Out Force and Stiffness Similar to Titanium Screws , 1998, The American journal of sports medicine.

[19]  A. Weiler,et al.  Hamstring tendon fixation using interference screws: a biomechanical study in calf tibial bone. , 1998, Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association.

[20]  G. Finerman,et al.  Morphology and Matrix Composition During Early Tendon to Bone Healing , 1997, Clinical orthopaedics and related research.

[21]  J. Nyland,et al.  Biomechanical comparison between BioScrew and titanium alloy interference screws for bone-patellar tendon-bone graft fixation in anterior cruciate ligament reconstruction. , 1997, Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association.

[22]  S Rupp,et al.  Fixation strength of a biodegradable interference screw and a press-fit technique in anterior cruciate ligament reconstruction with a BPTB graft. , 1997, Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association.

[23]  S. Majumdar,et al.  High Resolution Magnetic Resonance Imaging of the Calcaneus: Age-Related Changes in Trabecular Structure and Comparison with Dual X-Ray Absorptiometry Measurements , 1997, Calcified Tissue International.

[24]  L. L. Johnson,et al.  Metal and biodegradable interference screws: comparison of failure strength. , 1996, Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association.

[25]  Greg A. Brown,et al.  Comparison of Failure Strength Between Metallic and Absorbable Interference Screws , 1996, The American journal of sports medicine.

[26]  M. Järvinen,et al.  Fixation strength of a biodegradable screw in anterior cruciate ligament reconstruction. , 1995, The Journal of bone and joint surgery. British volume.

[27]  K. Bose,et al.  Biomechanical Study on Tibialis Posterior Tendon Transfers , 1995, Clinical orthopaedics and related research.

[28]  J. M. Ballard,et al.  Comparative Evaluation of Bone Suture Anchor to Bone Tunnel Fixation of Tibialis Anterior Tendon in Cadaveric Cuboid Bone: A Biomechanical Investigation , 1995, Foot & ankle international.

[29]  L. Paulos,et al.  Preliminary results of an absorbable interference screw. , 1995, Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association.

[30]  P. Kurzweil,et al.  Tibial interference screw removal following anterior cruciate ligament reconstruction. , 1995, Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association.

[31]  M. Ito,et al.  Dual energy X-ray absorptiometry of the calcaneus: comparison with other techniques to assess bone density and value in predicting risk of spine fracture. , 1994, AJR. American journal of roentgenology.

[32]  I. Vuori,et al.  Precision of dual-energy x-ray absorptiometry in determining bone mineral density and content of various skeletal sites. , 1992, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[33]  J. Andrish,et al.  A biomechanical comparison of different surgical techniques of graft fixation in anterior cruciate ligament reconstruction , 1987, The American journal of sports medicine.