Evaluation of Two Types of Biosensors for Immunoassay of Botulinum Toxin

Immunoassay of botulinum toxin (BTX) B type was investigated using two typed of biosensors: light addressable potentiometric sensor (LAPS) and surface plasmon resonance (SPR) sensor. Urease-tagged and immuno-filtration capture method have been used for LAPS. Tag-free and direct binding real-time detection method have been used for SPR sensor. The detection limit of sandwich assay format with LAPS was 10 ng/ml, which was the lowest among methods tested. SPR has the advantage of being more convenient because tag-free direct binding assay can be used and reaction time was reduced, regardless of low sensitivity. This result shows that sandwich assay format with LAPS can be used as an alternative method of BTX mouse bioassay which is known as the most sensitive method for the detection of BTX.