Low-Power Miniaturized Helium Dielectric Barrier Discharge Photoionization Detectors for Highly Sensitive Vapor Detection.

This paper presents the design, fabrication, and characterization of a microhelium dielectric barrier discharge photoionization detector (μHDBD-PID) on chip with dimensions of only ∼15 mm × ∼10 mm × ∼0.7 mm and weight of only ∼0.25 g. It offers low power consumption (<400 mW), low helium consumption (5.8 mL/min), rapid response (up to ∼60 ms at a flow rate of 1.5 mL/min), quick warm-up time (∼5 min), an excellent detection limit (a few picograms), a large linear dynamic range (>4 orders of magnitude), and maintenance-free operation. Furthermore, the μHDBD-PID can be driven with a miniaturized (∼5 cm × ∼2.5 cm × ∼2.5 cm), light (22 g), and low cost (∼$2) power supply with only 1.5 VDC input. The dependence of the μHDBD-PID performance on bias voltage, auxiliary helium flow rate, carrier gas flow rate, and temperature was also systematically investigated. Finally, the μHDBD-PID was employed to detect permanent gases and a sublist of the EPA 8260 standard reagents that include 51 analytes. The μHDBD-PID developed here can have a broad range of applications in portable and microgas chromatography systems for in situ, real-time, and sensitive gas analysis.

[1]  Rongrong Zhang,et al.  The generation characteristics of dielectric barrier glow discharge plasma in air , 2013 .

[2]  J. Franzke,et al.  Dielectric barrier discharges in analytical chemistry. , 2011, The Analyst.

[3]  W. M. Haynes CRC Handbook of Chemistry and Physics , 1990 .

[4]  Michael Keidar,et al.  The Effect of Tuning Cold Plasma Composition on Glioblastoma Cell Viability , 2014, PloS one.

[5]  S. Stearns,et al.  Pulsed discharge emission detector: an element-selective detector for gas chromatography. , 2000, Journal of chromatography. A.

[6]  J. Behnke,et al.  Modelling of the homogeneous barrier discharge in helium at atmospheric pressure , 2003 .

[7]  A. Bogaerts,et al.  Gas discharge plasmas and their applications , 2002 .

[8]  S. Wilkinson,et al.  Dielectric Barrier Discharge Plasma Actuators for Flow Control , 2010 .

[9]  C. Zheng,et al.  Dielectric barrier discharge carbon atomic emission spectrometer: universal GC detector for volatile carbon-containing compounds. , 2014, Analytical chemistry.

[10]  A. Yu,et al.  Helium direct current discharge ionization detector for gas chromatography , 1997 .

[11]  J. Luong,et al.  Universal and discriminative detection using a miniaturized pulsed discharge detector in comprehensive two-dimensional GC. , 2006, Journal of separation science.

[12]  Meng-Qiang Zhao,et al.  Development of a dielectric barrier discharge ion source for ambient mass spectrometry , 2007, Journal of the American Society for Mass Spectrometry.

[13]  Steven J Ray,et al.  Atmospheric pressure chemical ionization source. 1. Ionization of compounds in the gas phase. , 2008, Analytical chemistry.

[14]  C. Engelhard,et al.  Low-temperature plasma ionization differential ion mobility spectrometry. , 2015, Analytical chemistry.

[15]  X. Hou,et al.  Dielectric barrier discharge molecular emission spectrometer as gas chromatographic detector for amines , 2015 .

[16]  M. Agah,et al.  A micro-discharge photoionization detector for micro-gas chromatography , 2014, Microchimica Acta.

[17]  Shree Narayanan,et al.  Characterization of a micro-helium discharge detector for gas chromatography , 2015 .

[18]  S. Stearns,et al.  Pulsed discharge helium ionization detector Universal detector for inorganic and organic compounds at the low picogram level , 1994 .

[19]  Adam S. Pimentel,et al.  Pulsed Discharge Helium Ionization Detector for Highly Sensitive Aquametry , 2016, Analytical sciences : the international journal of the Japan Society for Analytical Chemistry.

[20]  J. Driscoll Evaluation of a new photoionization detector for organic compounds , 1977 .

[21]  Chao Zhang,et al.  Direct detection of explosives on solid surfaces by mass spectrometry with an ambient ion source based on dielectric barrier discharge. , 2007, Journal of mass spectrometry : JMS.

[22]  K. Kurabayashi,et al.  In situ calibration of micro-photoionization detectors in a multi-dimensional micro-gas chromatography system. , 2016, The Analyst.

[23]  E. Chen,et al.  Windowless pulsed-discharge photoionization detector application to qualitative analysis of volatile organic compounds , 1996 .

[24]  S. Stearns,et al.  Pulsed discharge helium ionization detector with multiple combined bias/collecting electrodes for gas chromatography. , 2013, Journal of chromatography. A.

[25]  M. Agah,et al.  GC-on-chip: integrated column and photoionization detector. , 2015, Lab on a chip.

[26]  L. Rao,et al.  Non-radioactive electron-capture detector , 1999 .

[27]  S. Stearns,et al.  Characterization of the pulsed discharge electron capture detector. , 1996, Analytical chemistry.

[28]  K. Kurabayashi,et al.  A fully automated portable gas chromatography system for sensitive and rapid quantification of volatile organic compounds in water , 2016 .

[29]  P. Mochalski,et al.  Evaluation of a pulse-discharge helium ionisation detector for the determination of neon concentrations by gas chromatography. , 2004, Journal of chromatography. A.

[30]  John R. Barnes,et al.  Inductors and Transformers , 2004 .

[31]  Komandoor E Achyuthan,et al.  Development of a Mesoscale Pulsed Discharge Helium Ionization Detector for Portable Gas Chromatography. , 2015, Analytical sciences : the international journal of the Japan Society for Analytical Chemistry.

[32]  M. Jafari Low-temperature plasma ionization ion mobility spectrometry. , 2011, Analytical chemistry.

[33]  J. Luong,et al.  Gas chromatographic applications with the dielectric barrier discharge detector. , 2006, Journal of chromatographic science.

[34]  Y. Mok,et al.  Temperature effect on hydrocarbon-enhanced nitric oxide conversion using a dielectric barrier discharge reactor , 2003 .

[35]  K. Kurabayashi,et al.  Flow-through microfluidic photoionization detectors for rapid and highly sensitive vapor detection. , 2015, Lab on a chip.

[36]  C. Zheng,et al.  Dielectric barrier discharge molecular emission spectrometer as multichannel GC detector for halohydrocarbons. , 2011, Analytical chemistry.

[37]  S. Stearns,et al.  Development and characterization of chlorine-selective pulsed discharge emission detector for gas chromatography. , 2000, Journal of chromatography. A.