Procedures for Calibrating Neutron Personnel Dosimeters

Procedures are given for routine testing and calibration of neutron dosimeters and remmeters with radioactive neutron sources. The issues addressed include: the choice of neutron source; phantom construction; fluence to dose eouivalent conversion; and corrections for air scatter, room return, and anisotropic-neutron emission. Explicit, semi-empirical , analytic expressions are given for the room return correction, and calculated numerical values are given for air scatter. Key WDrds: air scatter; calibration; californium; dose equivalent; dosimeter; neutron; rerrmeter; room return.

[1]  R. B. Schwartz,et al.  Neutron-dosimeter performance and associated calibrations at nuclear power plants , 1982 .

[2]  R. B. Schwartz,et al.  Measurement of neutrons reflected form the surfaces of a calibration room. , 1982, Health physics.

[3]  C. Eisenhauer,et al.  Difficulties of calculating spectrum-averaged values of a neutron dose equivalent. , 1981, Health Physics.

[4]  K. Harrison The calibration of neutron detectors with spherical moderators , 1981 .

[5]  W. Alberts,et al.  Neutron fluence-to-dose equivalent conversion factors. , 1981, Health Physics.

[6]  W. Mannhart Progress in Integral Data and Their Accuracy: Average Neutron Cross Sections in the Californium-252 Benchmark Field , 1981 .

[7]  A. Nash,et al.  Effect of badge position on the response of an albedo neutron dosimeter. , 1980, Health Physics.

[8]  D. E. Hankins Phantoms for calibrating albedo neutron dosimeters. , 1980, Health Physics.

[9]  R. C. McCall,et al.  Transport of Accelerator Produced Neutrons in a Concrete Room , 1978, IEEE Transactions on Nuclear Science.

[10]  D. E. Hankins Neutron dosimetry studies at the Lawrence Livermore Laboratory , 1977 .

[11]  H. Ing,et al.  Calculated spectra for the dosimetry of D2O-moderated neutrons. , 1977, Health physics.

[12]  H. T. Heaton,et al.  A Californium-252 Fission Spectrum Irradiation Facility for Neutron Reaction Rate Measurements , 1977 .

[13]  D. E. Hankins Energy dependence measurements of remmeters and albedo neutron dosimeters at neutron energies of thermal and between 2 keV and 5. 67 MeV , 1977 .

[14]  C. Eisenhauer,et al.  Fission spectrum neutrons for cross section validation and neutron flux transfer , 1975 .

[15]  G. L. Simmons Neutron dose and fluence distributions in an infinite air medium , 1973 .

[16]  E. Axton The effective centre of a moderating sphere when used as an instrument for fast neutron flux measurement , 1972 .

[17]  C. Eisenhauer An image source technique for calculation reflection of gamma rays or neutrons. , 1965, Health Physics.

[18]  F. J. Allen,et al.  DEPENDENCE OF NEUTRON ALBEDOS UPON HYDROGEN CONTENT OF A SHIELD , 1963 .