Deblurring of breathing motion artifacts in thoracic PET images by deconvolution methods.

In FDG-PET imaging of thoracic tumors, blurring due to breathing motion often significantly degrades the quality of the observed image, which then obscures the tumor boundary. We demonstrate a deblurring technique that combines patient-specific motion estimates of tissue trajectories with image deconvolution techniques, thereby partially eliminating breathing-motion induced artifacts. Two data sets were used to evaluate the methodology including mobile phantoms and clinical images. The clinical images consist of PET/CT co-registered images of patients diagnosed with lung cancer. A breathing motion model was used to locally estimate the location-dependent tissue location probability function (TLP) due to breathing. The deconvolution process is carried by an expectation-maximization (EM) iterative algorithm using the motion-based TLP. Several methods were used to improve the robustness of the deblurring process by mitigating noise amplification and compensating for motion estimate uncertainties. The mobile phantom study with controlled settings demonstrated significant reduction in underestimation error of concentration in high activity case without significant superiority between the different applied methods. In case of medium activity concentration (moderate noise levels), less improvement was reported (10%-15% reduction in underestimation error relative to 15%-20% reduction in high concentration). Residual denoising using wavelets offered the best performance for this case. In the clinical data case, the image spatial resolution was significantly improved, especially in the direction of greatest motion (cranio-caudal). The EM algorithm converged within 15 and 5 iterations in the large and small tumor cases, respectively. A compromise between a figure-of-merit and entropy minimization was suggested as a stopping criterion. Regularization techniques such as wavelets and Bayesian methods provided further refinement by suppressing noise amplification. Our initial results show that the proposed method provides a feasible framework for improving PET thoracic images, without the need for gated/4-D PET imaging, when 4-D CT is available to estimate tumor motion.

[1]  B. Sharma,et al.  PET/CT in oncology--a major advance. , 2005, Clinical radiology.

[2]  T. Fusco,et al.  Myopic deconvolution method for adaptive optics images of stellar fields , 1999 .

[3]  Sasa Mutic,et al.  Impact of FDG-PET on radiation therapy volume delineation in non-small-cell lung cancer. , 2004, International journal of radiation oncology, biology, physics.

[4]  K. Lam,et al.  Improvement of CT-based treatment-planning models of abdominal targets using static exhale imaging. , 1998, International journal of radiation oncology, biology, physics.

[5]  G. Christensen,et al.  A method for the reconstruction of four-dimensional synchronized CT scans acquired during free breathing. , 2003, Medical physics.

[6]  Nikolas P. Galatsanos,et al.  Hierarchical Bayesian image restoration from partially known blurs , 2000, IEEE Trans. Image Process..

[7]  P. Keall 4-dimensional computed tomography imaging and treatment planning. , 2004, Seminars in radiation oncology.

[8]  Roger Fulton,et al.  The design and implementation of a motion correction scheme for neurological PET. , 2003, Physics in medicine and biology.

[9]  T Kozuka,et al.  Lung cancer: intermittent irradiation synchronized with respiratory motion--results of a pilot study. , 1998, Radiology.

[10]  J. Dempsey,et al.  Novel breathing motion model for radiotherapy. , 2005, International journal of radiation oncology, biology, physics.

[11]  B.M.W. Tsui,et al.  A simulation study of the effect of gating scheme on respiratory motion bluffing in FDG Lung PET , 2002, 2002 IEEE Nuclear Science Symposium Conference Record.

[12]  R Mohan,et al.  Predicting respiratory motion for four-dimensional radiotherapy. , 2004, Medical physics.

[13]  R. Castillo,et al.  Attenuation correction of PET images with respiration-averaged CT images in PET/CT. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[14]  I. Johnstone,et al.  Adapting to Unknown Smoothness via Wavelet Shrinkage , 1995 .

[15]  T J Holmes,et al.  Blind deconvolution of quantum-limited incoherent imagery: maximum-likelihood approach. , 1992, Journal of the Optical Society of America. A, Optics and image science.

[16]  Gregory C Sharp,et al.  Speed and amplitude of lung tumor motion precisely detected in four-dimensional setup and in real-time tumor-tracking radiotherapy. , 2006, International journal of radiation oncology, biology, physics.

[17]  A. M. Suárez,et al.  La tomografía por emisión de positrones (PET) en la patología oncológica de cabeza y cuello , 2004 .

[18]  Raymond F. Muzic,et al.  A method to correct for scatter, spillover, and partial volume effects in region of interest analysis in PET , 1998, IEEE Transactions on Medical Imaging.

[19]  C C Ling,et al.  The deep inspiration breath-hold technique in the treatment of inoperable non-small-cell lung cancer. , 2000, International journal of radiation oncology, biology, physics.

[20]  Fionn Murtagh,et al.  Deconvolution in Astronomy: A Review , 2002 .

[21]  J. M. Ollinger,et al.  Positron Emission Tomography , 2018, Handbook of Small Animal Imaging.

[22]  A. Pevsner,et al.  Quantitation of respiratory motion during 4D-PET/CT acquisition. , 2004, Medical physics.

[23]  Wolfgang A. Tome,et al.  TU‐C‐J‐6B‐03: Deformable Fusion of 4D PET Images , 2005 .

[24]  G J Kutcher,et al.  Deep inspiration breath-hold technique for lung tumors: the potential value of target immobilization and reduced lung density in dose escalation. , 1999, International journal of radiation oncology, biology, physics.

[25]  Yuji Nakamoto,et al.  Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT , 2003, European Journal of Nuclear Medicine and Molecular Imaging.

[26]  R. Lecomte,et al.  Respiratory gating for 3-dimensional PET of the thorax: feasibility and initial results. , 2004, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[27]  S. Mustafovic,et al.  Image reconstruction of motion corrected sinograms , 2003, 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515).

[28]  E. Larsen,et al.  A method for incorporating organ motion due to breathing into 3D dose calculations. , 1999, Medical physics.

[29]  S. Nehmeh,et al.  PET/CT in non-small-cell lung cancer: value of respiratory-gated PET. , 2005, Chang Gung medical journal.

[30]  Harry Keller,et al.  Application of the spirometer in respiratory gated radiotherapy. , 2003, Medical physics.

[31]  F. Courbin,et al.  Deconvolution with Correct Sampling , 1997, astro-ph/9704059.

[32]  J. Karp,et al.  CHAPTER 10 – PET Systems , 2004 .

[33]  J. Wong,et al.  The use of active breathing control (ABC) to reduce margin for breathing motion. , 1999, International journal of radiation oncology, biology, physics.

[34]  Yu Wang,et al.  A simple respiration gating technique and its application in high-resolution PET camera , 2005, IEEE Transactions on Nuclear Science.

[35]  D Strul,et al.  Robustness of Anatomically Guided Pixel-by-Pixel Algorithms for Partial Volume Effect Correction in Positron Emission Tomography , 1999, Journal of cerebral blood flow and metabolism : official journal of the International Society of Cerebral Blood Flow and Metabolism.

[36]  Osama Mawlawi,et al.  PET/CT imaging artifacts. , 2005, Journal of nuclear medicine technology.

[37]  Aggelos K. Katsaggelos,et al.  Image identification and restoration based on the expectation-maximization algorithm , 1990 .

[38]  M E Phelps,et al.  Quantitation in Positron Emission Computed Tomography: 6. Effect of Nonuniform Resolution , 1982, Journal of computer assisted tomography.

[39]  C. Ling,et al.  Technical aspects of the deep inspiration breath-hold technique in the treatment of thoracic cancer. , 2000, International journal of radiation oncology, biology, physics.

[40]  Miles N. Wernick,et al.  CHAPTER 21 – Iterative Image Reconstruction , 2004 .

[41]  Kenneth E Rosenzweig,et al.  Reduction of respiratory motion artifacts in PET imaging of lung cancer by respiratory correlated dynamic PET: methodology and comparison with respiratory gated PET. , 2003, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[42]  M. V. van Herk,et al.  Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy. , 2002, International journal of radiation oncology, biology, physics.

[43]  Ja-Yong Koo,et al.  Wavelet deconvolution , 2002, IEEE Trans. Inf. Theory.

[44]  Sasa Mutic,et al.  Quantitation of the reconstruction quality of a four-dimensional computed tomography process for lung cancer patients. , 2005, Medical physics.

[45]  A. Evans,et al.  Correction for partial volume effects in PET: principle and validation. , 1998, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[46]  Sasa Mutic,et al.  PET-guided IMRT for cervical carcinoma with positive para-aortic lymph nodes-a dose-escalation treatment planning study. , 2003, International journal of radiation oncology, biology, physics.

[47]  M. V. van Herk,et al.  Physical aspects of a real-time tumor-tracking system for gated radiotherapy. , 2000, International journal of radiation oncology, biology, physics.

[48]  S. Seliounine,et al.  Application of expectation maximization algorithms for image resolution improvement in a small animal PET system , 2005, IEEE Transactions on Nuclear Science.

[49]  Ming Jiang,et al.  Blind deblurring of spiral CT images , 2003, Conference Record of Thirty-Fifth Asilomar Conference on Signals, Systems and Computers (Cat.No.01CH37256).

[50]  Shree K. Nayar,et al.  Motion-based motion deblurring , 2004, IEEE Transactions on Pattern Analysis and Machine Intelligence.

[51]  L. Adler,et al.  Simultaneous recovery of size and radioactivity concentration of small spheroids with PET data. , 1999, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[52]  V Z Mesarović,et al.  Iterative linear minimum mean-square-error image restoration from partially known blur. , 2000, Journal of the Optical Society of America. A, Optics, image science, and vision.

[53]  Anil K. Jain Fundamentals of Digital Image Processing , 2018, Control of Color Imaging Systems.

[54]  R. Mohan,et al.  Acquiring a four-dimensional computed tomography dataset using an external respiratory signal. , 2003, Physics in medicine and biology.

[55]  Issam El-Naqa,et al.  Compensation of breathing motion artifacts in thoracic PET images by wavelet-based deconvolution , 2006, 3rd IEEE International Symposium on Biomedical Imaging: Nano to Macro, 2006..

[56]  Michael Potmesil,et al.  Modeling motion blur in computer-generated images , 1983, SIGGRAPH.

[57]  C. Ling,et al.  Effect of respiratory gating on quantifying PET images of lung cancer. , 2002, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[58]  M. Victor Wickerhauser,et al.  Adapted wavelet analysis from theory to software , 1994 .

[59]  Joseph O. Deasy,et al.  Automated 4D lung computed tomography reconstruction during free breathing for conformal radiation therapy , 2004, SPIE Medical Imaging.

[60]  T. R. Gosnell,et al.  Digital Image Reconstruction: Deblurring and Denoising , 2005 .

[61]  Michael E. Phelps,et al.  Quantitation in Positron Emission Computed Tomography , 1980 .

[62]  Kenneth E Rosenzweig,et al.  Tumor Motion Control in the Treatment of Non Small Cell Lung Cancer , 2005, Cancer investigation.

[63]  H. Biersack,et al.  FDG-PET in monitoring therapy of breast cancer , 2004, European Journal of Nuclear Medicine and Molecular Imaging.

[64]  T. Yen,et al.  Utility of 2-Fluoro-2-Deoxy-D-Glucose Positron Emission Tomography in Managing Patients of Colorectal Cancer With Unexplained Carcinoembryonic Antigen Elevation at Different Levels , 2005, Diseases of the colon and rectum.

[65]  Lei Xing,et al.  Model-based image reconstruction for four-dimensional PET. , 2006, Medical physics.

[66]  E Yorke,et al.  Four-dimensional (4D) PET/CT imaging of the thorax. , 2004, Medical physics.

[67]  PotmesilMichael,et al.  Modeling motion blur in computer-generated images , 1983 .

[68]  I El Naqa,et al.  A comparison of Monte Carlo dose calculation denoising techniques , 2005, Physics in medicine and biology.

[69]  Joseph O Deasy,et al.  Accelerating Monte Carlo simulations of radiation therapy dose distributions using wavelet threshold de-noising. , 2002, Medical physics.

[70]  F. Turkheimer,et al.  Kinetic modeling in positron emission tomography. , 2002, The quarterly journal of nuclear medicine : official publication of the Italian Association of Nuclear Medicine (AIMN) [and] the International Association of Radiopharmacology.