BACKGROUND
The U.S. Food and Drug Administration (FDA) evaluates medical devices by two main pathways. The more stringent Premarket Approval (PMA) review requires clinical trials, and the Premarket Notification 510(k) process generally exempts devices from clinical trials if they prove to be substantially equivalent to existing devices. We hypothesized that orthopaedic devices are more likely to be cleared through the 510(k) process and thus are more susceptible to being recalled.
METHODS
Using the FDA's public database, we searched for the following: PMA and 510(k) clearances for orthopaedics and non-orthopaedic specialties, including General & Plastic Surgery, Gastroenterology/Urology, Obstetrics/Gynecology, and Ear Nose & Throat, from 1992 to 2012. Additionally, we searched for all device recall events from 2002 to 2012. For the top-twenty recall companies, we calculated the odds ratio that compares the likelihood of recall for 510(k)-approved devices with that for PMA-approved devices.
RESULTS
From 1992 to 2012, the proportion of non-orthopaedic devices cleared via the 510(k) process decreased from 91% to 53%. However, that of orthopaedic devices decreased only from 94% to 88%. Furthermore, we found that from 2002 to 2012, the percentage of recalled devices was 17.8% for 510(k)-cleared devices and 1.6% for PMA-approved devices. When stratified on the basis of recall class, the odds ratios were 3.5 for class-I devices, 13.2 for class-II devices, and 8.5 for class-III devices.
CONCLUSIONS
Given that 510(k)-cleared devices were 11.5 times more likely to be recalled than PMA-approved devices, it is concerning that most orthopaedic devices are cleared through the 510(k) process with limited clinical trials data.
CLINICAL RELEVANCE
When orthopaedic surgeons are considering using a new device clinically in their patients, it is important for them to consider how the new device was approved by the FDA. If the device was approved by the 510(k) pathway, then it may have been approved without additional clinical studies confirming efficacy or safety.
[1]
Mark McClellan,et al.
Ensuring safe and effective medical devices.
,
2003,
The New England journal of medicine.
[2]
Daniel B. Kramer,et al.
FDA approval of cardiac implantable electronic devices via original and supplement premarket approval pathways, 1979-2012.
,
2014,
JAMA.
[3]
J J Gartland,et al.
Orthopaedic clinical research. Deficiencies in experimental design and determinations of outcome.
,
1988,
The Journal of bone and joint surgery. American volume.
[4]
R. Redberg,et al.
FDA regulation of cardiovascular devices and opportunities for improvement
,
2013,
Journal of Interventional Cardiac Electrophysiology.
[5]
P. Spencer,et al.
Medical device recalls.
,
2001,
Health law in Canada.
[6]
R. Barrack,et al.
An analysis of Food and Drug Administration medical device reports relating to total joint components.
,
1997,
The Journal of arthroplasty.
[7]
Larry J. Brindza,et al.
What is a premarket notification 510(k)
,
1980
.
[8]
T. Brennan,et al.
Improving the Postmarket Surveillance of Total Joint Arthroplasty Devices
,
2008,
The open rheumatology journal.
[9]
Stephen E. Graves,et al.
The 510(k) ancestry of a metal-on-metal hip implant.
,
2013,
The New England journal of medicine.
[10]
D. Zuckerman,et al.
Medical device recalls and the FDA approval process.
,
2011,
Archives of internal medicine.
[11]
J. Samet,et al.
Food and Drug Administration
,
2007,
BMJ : British Medical Journal.