Accuracy of coronary CT angiography using a submillisievert fraction of radiation exposure: comparison with invasive coronary angiography.

BACKGROUND Coronary computed tomography angiography (CTA) is increasingly being used for evaluation of coronary artery disease (CAD). As a result of the widely reported potential of carcinogenic risk from x-ray based examinations, many strategies have been developed for dose reduction with CTA. OBJECTIVES The purpose of this study was to assess the diagnostic accuracy of CTA acquired with a submillisievert fraction of effective radiation dose reconstructed with a model-based iterative reconstruction (MBIR) using invasive coronary angiography (ICA) as a standard of reference. METHODS In 36 patients (body mass index range 17 to 39 kg/m(2)) undergoing ICA for CAD evaluation, a CTA was acquired using very low tube voltage (80 to 100 kV) and current (150 to 210 mA) and was reconstructed with MBIR. CAD (defined as ≥50% luminal narrowing) was assessed on CTA and on ICA. RESULTS CTA resulted in an estimated radiation dose exposure of 0.29 ± 0.12 mSv (range 0.16 to 0.53 mSv), yielding 96.9% (436 of 450) interpretable segments. On an intention-to-diagnose basis, no segment was excluded, and vessels with at least 1 nonevaluable segment and no further finding were classified as false positive. This resulted in a sensitivity, specificity, positive, and negative predictive value and accuracy of 100%, 74%, 77%, 100%, and 86% per patient and 85%, 86%, 56%, 96%, and 85% per vessel, respectively. CONCLUSIONS The use of MBIR reconstruction allows accurate noninvasive diagnosis of CAD with CTA at a submillisievert fraction of effective radiation dose comparable with a chest x-ray in 2 views.

[1]  C. Unterberg-Buchwald,et al.  Head-to-head comparison of prospectively triggered vs retrospectively gated coronary computed tomography angiography: Meta-analysis of diagnostic accuracy, image quality, and radiation dose. , 2013, American heart journal.

[2]  L. Shaw,et al.  Impact of Ethnicity and Gender Differences on Angiographic Coronary Artery Disease Prevalence and In-Hospital Mortality in the American College of Cardiology–National Cardiovascular Data Registry , 2008, Circulation.

[3]  H. Shim,et al.  Image quality of Adaptive Iterative Dose Reduction 3D of coronary CT angiography of 640-slice CT: comparison with filtered back-projection , 2013, The International Journal of Cardiovascular Imaging.

[4]  B. Hamm,et al.  Noninvasive Coronary Angiography by 320-Row Computed Tomography With Lower Radiation Exposure and Maintained Diagnostic Accuracy: Comparison of Results With Cardiac Catheterization in a Head-to-Head Pilot Investigation , 2009, Circulation.

[5]  A. Einstein,et al.  Cardiac imaging: does radiation matter? , 2012, European heart journal.

[6]  Fuminari Tatsugami,et al.  Accuracy of low-dose computed tomography coronary angiography using prospective electrocardiogram-triggering: first clinical experience. , 2008, European heart journal.

[7]  D. Andreini,et al.  Feasibility and diagnostic accuracy of a low radiation exposure protocol for prospective ECG-triggering coronary MDCT angiography. , 2012, Clinical radiology.

[8]  Volkmar Falk,et al.  Guidelines on Myocardial Revascularization the Task Force on Myocardial Revascularization of the European Society of Cardiology (esc) and the European Association for Cardio-thoracic Surgery (eacts) Developed with the Special Contribution of the European Association for Percutaneous Cardiovascular I , 2022 .

[9]  Masaki Katsura,et al.  Model-Based Iterative Reconstruction Technique for Ultralow-Dose Chest CT: Comparison of Pulmonary Nodule Detectability With the Adaptive Statistical Iterative Reconstruction Technique , 2013, Investigative radiology.

[10]  J. Stehli,et al.  First experience with monochromatic coronary computed tomography angiography from a 64-slice CT scanner with Gemstone Spectral Imaging (GSI). , 2013, Journal of cardiovascular computed tomography.

[11]  Jeroen J. Bax,et al.  Cardiac computed tomography: indications, applications, limitations, and training requirements: report of a Writing Group deployed by the Working Group Nuclear Cardiology and Cardiac CT of the European Society of Cardiology and the European Council of Nuclear Cardiology. , 2008, European heart journal.

[12]  Tobias A. Fuchs,et al.  Coronary computed tomography angiography with model-based iterative reconstruction using a radiation exposure similar to chest X-ray examination , 2014, European heart journal.

[13]  Sana Boudabbous,et al.  Computed tomography of the chest with model-based iterative reconstruction using a radiation exposure similar to chest X-ray examination: preliminary observations , 2013, European Radiology.

[14]  L. Appel,et al.  How many patients are necessary to assess test performance? , 1990, JAMA.

[15]  Stefan Ulzheimer,et al.  Accuracy of prospectively ECG-triggered very low-dose coronary dual-source CT angiography using iterative reconstruction for the detection of coronary artery stenosis: comparison with invasive catheterization. , 2014, European heart journal cardiovascular Imaging.

[16]  Allen J. Taylor,et al.  Effect of hybrid iterative reconstruction technique on quantitative and qualitative image analysis at 256-slice prospective gating cardiac CT , 2012, European Radiology.

[17]  V. Treyer,et al.  First head-to-head comparison of effective radiation dose from low-dose 64-slice CT with prospective ECG-triggering versus invasive coronary angiography , 2009, Heart.

[18]  Volkmar Falk,et al.  Guidelines on myocardial revascularization. , 2010, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[19]  J. Knuuti,et al.  Ionizing radiation risks of cardiac imaging: estimates of the immeasurable. , 2011, European heart journal.

[20]  M. Vembar,et al.  Diagnostic accuracy of low-dose 256-slice multi-detector coronary CT angiography using iterative reconstruction in patients with suspected coronary artery disease , 2013, European Radiology.

[21]  M. Budoff,et al.  Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter ACCURACY (Assessment by Coronary Computed Tomographic Angiography of Indi , 2008, Journal of the American College of Cardiology.

[22]  I. Burger,et al.  Validation of a new contrast material protocol adapted to body surface area for optimized low-dose CT coronary angiography with prospective ECG-triggering , 2010, The International Journal of Cardiovascular Imaging.

[23]  Michael Fiechter,et al.  CT coronary angiography: impact of adapted statistical iterative reconstruction (ASIR) on coronary stenosis and plaque composition analysis , 2013, The International Journal of Cardiovascular Imaging.

[24]  J. Stehli,et al.  Image quality in low-dose coronary computed tomography angiography with a new high-definition CT scanner , 2013, The International Journal of Cardiovascular Imaging.

[25]  V. Treyer,et al.  Low-dose computed tomography coronary angiography with prospective electrocardiogram triggering: feasibility in a large population. , 2011, Journal of the American College of Cardiology.

[26]  Michael Fiechter,et al.  Inter-scan variability of coronary artery calcium scoring assessed on 64-multidetector computed tomography vs. dual-source computed tomography: a head-to-head comparison. , 2011, European heart journal.

[27]  N. Buderer,et al.  Statistical methodology: I. Incorporating the prevalence of disease into the sample size calculation for sensitivity and specificity. , 1996, Academic emergency medicine : official journal of the Society for Academic Emergency Medicine.

[28]  N. Paul,et al.  Perioperative β-Blockers : Use With Caution Perioperative β Blockers in Patients Having Non-Cardiac Surgery : A Meta-Analysis , 2010 .

[29]  F. Eberli,et al.  Functionally relevant coronary artery disease: comparison of 64-section CT angiography with myocardial perfusion SPECT. , 2008, Radiology.

[30]  J. Leipsic,et al.  Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CT angiography. , 2010, AJR. American journal of roentgenology.

[31]  W. Huda,et al.  Effective doses in radiology and diagnostic nuclear medicine: a catalog. , 2008, Radiology.

[32]  W. Kalender,et al.  Image quality of ultra-low radiation exposure coronary CT angiography with an effective dose <0.1 mSv using high-pitch spiral acquisition and raw data-based iterative reconstruction , 2013, European Radiology.

[33]  B. Wall,et al.  Radiation exposure of the UK population from medical and dental X-ray examinations. , 2001 .

[34]  Jeroen J. Bax,et al.  2013 ESC guidelines on the management of stable coronary artery disease: the Task Force on the management of stable coronary artery disease of the European Society of Cardiology. , 2013, European heart journal.

[35]  U. Schoepf,et al.  Detection of coronary artery stenosis with sub-milliSievert radiation dose by prospectively ECG-triggered high-pitch spiral CT angiography and iterative reconstruction , 2013, European Radiology.

[36]  S. Achenbach,et al.  Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition. , 2010, European heart journal.

[37]  R. Frye,et al.  A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. , 1975, Circulation.

[38]  Jörg Hausleiter,et al.  Estimated radiation dose associated with cardiac CT angiography. , 2009, JAMA.

[39]  Mathias Prokop,et al.  Diagnostic accuracy of 64-slice computed tomography coronary angiography: a prospective, multicenter, multivendor study. , 2008, Journal of the American College of Cardiology.

[40]  Scott L. Thomas,et al.  Multilevel Modeling of Categorical Outcomes Using IBM SPSS , 2012 .

[41]  Fuminari Tatsugami,et al.  Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. , 2007, European heart journal.