Implementation of capillary blood glucose monitoring in a teaching hospital and determination of program requirements to maintain quality testing.

PURPOSE To study the implementation of bedside capillary glucose monitoring using a hospital-wide quality control (QC) program. METHODS A prospective study of QC performance in 7 outpatient and 39 inpatient treatment units was performed in a large teaching hospital over a 2-year period. Approximately 800 nurses were trained to perform bedside capillary glucose monitoring (Accu-Chek II, Boehringer-Mannheim, Indianapolis, IN). An eight-point QC program was instituted including proficiency testing, instrument maintenance, performance of daily controls, storage of reagent strips and supplies, instrument calibration, and documentation procedures. RESULTS Comparison of laboratory and bedside test results (split-sample proficiency testing) revealed Y = 1.004X + 7.26, r = 0.95, with a mean percent difference of -4.2% (p < 0.001). Less than 7% of results fell outside +/- 20% of the laboratory results. QC scores (0 = worst to 4 = best), based on adherence to the QC program, improved from 0 on the first inspection to 3.7 +/- 0.17 by the 11th inspection. The most common QC deficiencies were failure to perform split-sample testing (41.4%) and failure to perform instrument maintenance (30.2%). Significant differences were noted in the QC performance of different types of medical services. During the 2-year study period, the total number of glucose assays performed in the clinical laboratories decreased by 22.2% concurrent with initiation of bedside testing. The number of instruments in the hospital increased from 10 to 46. CONCLUSIONS Bedside capillary glucose assays can be widely implemented in large hospitals with an acceptable degree of accuracy. QC programs with frequent inspections are necessary to identify units that function inadequately, and a formal disciplinary policy is required to ensure compliance with the program.

[1]  D. Lezotte,et al.  Comparison of Blood Glucose Testing Using Reagent Strips With and Without a Meter (Chemstrips bG and Dextrostix/Dextrometer) , 1981, Diabetes Care.

[2]  A. Schiffrin,et al.  Evaluation of Two Methods of Self Blood Glucose Monitoring by Trained Insulin-dependent Diabetic Adolescents Outside the Hospital , 1983, Diabetes Care.

[3]  S. Leichter,et al.  Bedside Capillary Glucose Monitoring in the General Hospital , 1984, Diabetes Care.

[4]  M. Burritt,et al.  Evaluation of the One Touch Glucose Meter for Use by a Phlebotomy Team , 1990 .

[5]  R. Clements,et al.  An Evaluation of Patient Performance of and Their Satisfaction with Various Rapid Blood Glucose Measurement Systems , 1983, Diabetes Care.

[6]  A. Schiffrin,et al.  Multiple daily self-glucose monitoring: its essential role in long-term glucose control in insulin-dependent diabetic patients treated with pump and multiple subcutaneous injections. , 1982, Diabetes care.

[7]  D. Nathan,et al.  Bedside capillary glucose measurement by staff nurses in a general hospital. , 1986, The American journal of medicine.

[8]  E. Bruck,et al.  National Committee for Clinical Laboratory Standards. , 1980, Pediatrics.

[9]  R. Tattersall,et al.  Patient self-monitoring of blood glucose and refinements of conventional insulin treatment. , 1981, The American journal of medicine.

[10]  S. Chu,et al.  Evaluation of the Reliability of Bedside Glucose Testing , 1989 .

[11]  C. Ting,et al.  Evaluation of the quality of bedside monitoring of the blood glucose level in a teaching hospital. , 1988, CMAJ : Canadian Medical Association journal = journal de l'Association medicale canadienne.