Accuracy of electromagnetic tracking with a prototype field generator in an interventional OR setting.

PURPOSE The authors have studied the accuracy and robustness of a prototype electromagnetic window field generator (WFG) in an interventional radiology suite with a robotic C-arm. The overall purpose is the development of guidance systems combining real-time imaging with tracking of flexible instruments for bronchoscopy, laparoscopic ultrasound, endoluminal surgery, endovascular therapy, and spinal surgery. METHODS The WFG has a torus shape, which facilitates x-ray imaging through its centre. The authors compared the performance of the WFG to that of a standard field generator (SFG) under the influence of the C-arm. Both accuracy and robustness measurements were performed with the C-arm in different positions and poses. RESULTS The system was deemed robust for both field generators, but the accuracy was notably influenced as the C-arm was moved into the electromagnetic field. The SFG provided a smaller root-mean-square position error but was more influenced by the C-arm than the WFG. The WFG also produced smaller maximum and variance of the error. CONCLUSIONS Electromagnetic (EM) tracking with the new WFG during C-arm based fluoroscopy guidance seems to be a step forward, and with a correction scheme implemented it should be feasible.

[1]  Rajnikant V. Patel,et al.  Data fusion for catheter tracking using Kalman filtering: applications in robot-assisted catheter insertion , 2011, Medical Imaging.

[2]  Nadine Abi-Jaoudeh,et al.  Electromagnetic navigation for thoracic aortic stent-graft deployment: a pilot study in swine. , 2010, Journal of vascular and interventional radiology : JVIR.

[3]  Nassir Navab,et al.  Magneto-Optical Tracking of Flexible Laparoscopic Ultrasound: Model-Based Online Detection and Correction of Magnetic Tracking Errors , 2009, IEEE Transactions on Medical Imaging.

[4]  Ziv Yaniv,et al.  Electromagnetic tracking in the clinical environment. , 2009, Medical physics.

[5]  Makoto Hashizume,et al.  Intraoperative Magnetic Tracker Calibration Using a Magneto-Optic Hybrid Tracker for 3-D Ultrasound-Based Navigation in Laparoscopic Surgery , 2008, IEEE Transactions on Medical Imaging.

[6]  Guang-Zhong Yang,et al.  Freehand Cocalibration of Optical and Electromagnetic Trackers for Navigated Bronchoscopy , 2004, MIAR.

[7]  Berthold K. P. Horn,et al.  Closed-form solution of absolute orientation using unit quaternions , 1987 .

[8]  T Langø,et al.  Navigated ultrasound in laparoscopic surgery , 2009, Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy.

[9]  Ralf Eberhardt,et al.  Electromagnetic navigation diagnostic bronchoscopy in peripheral lung lesions. , 2007, Chest.

[10]  Takayuki Kitasaka,et al.  Bronchoscopy Navigation beyond Electromagnetic Tracking Systems: A Novel Bronchoscope Tracking Prototype , 2011, MICCAI.

[11]  Philippe Cinquin,et al.  3D Tracking of Laparoscopic Instruments Using Statistical and Geometric Modeling , 2011, MICCAI.

[12]  Terry M. Peters,et al.  A hardware and software protocol for the evaluation of electromagnetic tracker accuracy in the clinical environment: a multi-center study , 2007, SPIE Medical Imaging.

[13]  Volodymyr V. Kindratenko,et al.  A survey of electromagnetic position tracker calibration techniques , 2005, Virtual Reality.

[14]  J. Remy,et al.  Electromagnetic navigation diagnostic bronchoscopy for small peripheral lung lesions , 2007, European Respiratory Journal.

[15]  S. Ommedal,et al.  Side-Branched AAA Stent Graft Insertion Using Navigation Technology: A Phantom Study , 2007, European Surgical Research.

[16]  C Bürk,et al.  Prototype of an online navigation system for laparoscopic radiofrequency ablation. , 2009, Hepato-gastroenterology.

[17]  Douglas L. Jones,et al.  Mapping of Cardiac Electrophysiology Onto a Dynamic Patient-Specific Heart Model , 2009, IEEE Transactions on Medical Imaging.