PURPOSE
The purpose of this study was to develop and validate a mathematical model to describe narrow-beam attenuation of kilovoltage x-ray beams for the intended applications of half-value layer (HVL) and quarter-value layer (QVL) estimations, patient organ shielding, and computer modeling.
METHODS
An empirical model, which uses the Lambert W function and represents a generalized Lambert-Beer law, was developed. To validate this model, transmission of diagnostic energy x-ray beams was measured over a wide range of attenuator thicknesses [0.49-33.03 mm Al on a computed tomography (CT) scanner, 0.09-1.93 mm Al on two mammography systems, and 0.1-0.45 mm Cu and 0.49-14.87 mm Al using general radiography]. Exposure measurements were acquired under narrow-beam geometry using standard methods, including the appropriate ionization chamber, for each radiographic system. Nonlinear regression was used to find the best-fit curve of the proposed Lambert W model to each measured transmission versus attenuator thickness data set. In addition to validating the Lambert W model, we also assessed the performance of two-point Lambert W interpolation compared to traditional methods for estimating the HVL and QVL [i.e., semi-logarithmic (exponential) and linear interpolation].
RESULTS
The Lambert W model was validated for modeling attenuation versus attenuator thickness with respect to the data collected in this study (R2 > 0.99). Furthermore, Lambert W interpolation was more accurate and less sensitive to the choice of interpolation points used to estimate the HVL and/or QVL than the traditional methods of semilogarithmic and linear interpolation.
CONCLUSIONS
The proposed Lambert W model accurately describes attenuation of both monoenergetic radiation and (kilovoltage) polyenergetic beams (under narrow-beam geometry).
[1]
R. Sloboda,et al.
Linear accelerator photon beam quality at off-axis points.
,
1997,
Medical physics.
[2]
B. Archer,et al.
On the measurement of half-value layer in film-screen mammography.
,
1990,
Medical physics.
[3]
Cynthia H McCollough,et al.
A method to generate equivalent energy spectra and filtration models based on measurement for multidetector CT Monte Carlo dosimetry simulations.
,
2009,
Medical physics.
[4]
H. Akaike.
A new look at the statistical model identification
,
1974
.
[5]
D. Marquardt.
An Algorithm for Least-Squares Estimation of Nonlinear Parameters
,
1963
.
[6]
Cynthia H McCollough,et al.
Measurement of half-value layer in x-ray CT: a comparison of two noninvasive techniques.
,
2000
.
[7]
S C Bushong,et al.
Diagnostic X-ray shielding design based on an empirical model of photon attenuation.
,
1983,
Health physics.
[8]
M. McKetty,et al.
The AAPM/RSNA physics tutorial for residents. X-ray attenuation.
,
1998,
Radiographics : a review publication of the Radiological Society of North America, Inc.
[9]
B. Bjärngard,et al.
Attenuation in high-energy x-ray beams.
,
1994,
Medical physics.
[10]
Gaston H. Gonnet,et al.
On the LambertW function
,
1996,
Adv. Comput. Math..
[11]
Robert M Corless,et al.
Some applications of the Lambert W function to physics
,
2000,
Canadian Journal of Physics.
[12]
C Kleinschmidt,et al.
Analytical considerations of beam hardening in medical accelerator photon spectra.
,
1999,
Medical physics.
[13]
M. McNitt-Gray,et al.
Precision of dosimetry-related measurements obtained on current multidetector computed tomography scanners.
,
2010,
Medical physics.