A simplified method for manufacturing glass-insulated metal microelectrodes

A simplified method to manufacture durable, glass-insulated, tungsten microelectrodes with sufficient control of the final electrode impedance is described. This method requires only two instruments, an electrolytic etcher for wires and pipette puller, for manufacturing these electrodes. The manufacture of these electrodes involves 3 steps: (1) etching tungsten wire to sharpen the tip, (2) insulating the electrode by pulling a glass pipette over the sharpened tungsten wire and (3) assessing and adjusting the tip exposure and impedance of the electrode to meet recording requirements. Control over the electrode impedance is easily accomplished by varying the distance between the uppermost portion of the heating coil and the sharpened wire tip before a glass pipette is pulled over the wire tip. This distance determines the area of tip exposure and also the location where the glass insulation ends and the exposed electrode tip begins. A performance test of these electrodes in a chronically prepared monkey showed that they were strong enough to repeatedly penetrate thickened dura mater without significant changes in impedance and to isolate cortical neuronal activity after these multiple penetrations. Furthermore, the strength of these microelectrodes eliminated the need to remove reactive granular tissue from the dura overlying the recording site.

[1]  A microelectrode for depth recording in awake animals. , 1986, Electroencephalography and clinical neurophysiology.

[2]  G. Loeb,et al.  R. Lemon , 1985, Neuroscience.

[3]  A. Johnston,et al.  The controlled preparation of metal microelectrodes. , 1969, Israel journal of medical sciences.

[4]  G. Loeb,et al.  Parylene as a Chronically Stable, Reproducible Microelectrode Insulator , 1977, IEEE Transactions on Biomedical Engineering.

[5]  E. Marg A Rugged, Reliable and Sterilizable Microelectrode for recording Single Units from the Brain , 1964, Nature.

[6]  J. A. Freeman,et al.  A simple method for producing, in quantity, metal micro-electrodes with a desired taper and impedance. , 1969, Electroencephalography and clinical neurophysiology.

[7]  M L Wolbarsht,et al.  Glass Insulated Platinum Microelectrode , 1960, Science.

[8]  E. Chudler,et al.  Somatosensory, multisensory, and task-related neurons in cortical area 7b (PF) of unanesthetized monkeys. , 1994, Journal of neurophysiology.

[9]  J. Dostrovsky,et al.  An improved method for insulating tungsten micro-electrodes with glass [proceedings]. , 1977, The Journal of physiology.

[10]  P. Maclean,et al.  A platinum micro-electrode for intracerebral exploration with a chronically fixed stereotaxic device. , 1967, Electroencephalography and clinical neurophysiology.

[11]  D. Hubel Tungsten Microelectrode for Recording from Single Units. , 1957, Science.

[12]  J Y Lettvin,et al.  Glass-Coated Tungsten Microelectrodes , 1965, Science.

[13]  Insulating micro-electrodes by centrifuging. , 1966, Electroencephalography and clinical neurophysiology.

[14]  F. W. Srg A simple glass-coated, fire-polished tungsten electrode with conductance adjustment using hydrofluoridic acid , 1988 .

[15]  G. Rebec,et al.  A simple device for the reliable production of varnish-insulated, high-impedance tungsten microelectrodes , 1989, Journal of Neuroscience Methods.