Microflow fields in the hinge region of the CarboMedics bileaflet mechanical heart valve design.

OBJECTIVE The design of bileaflet mechanical heart valves includes some degree of leakage flow on valve closure for the reverse flow to wash the hinge and pivot region of the valve. It is believed that this reverse flow helps to prevent areas of stasis and inhibit microthrombus formation. However, the magnitude of this retrograde flow may also give rise to unacceptable levels of blood element damage and lead to platelet activation or hemolysis as a result of the increased flow velocities through the hinge region. The purpose of this study was to evaluate the hinge flow dynamics of a 23-mm CarboMedics bileaflet mechanical valve (Sulzer CarboMedics Inc, Austin, Tex) and then to compare the results with those of the St Jude Medical 23-mm Regent (St Jude Medical Inc, Minneapolis, Minn) and Medtronic Parallel (Medtronic, Inc, Minneapolis, Minn) valves studied earlier. This comparison allows new insight into the microflow fields within the hinge region of the CarboMedics bileaflet mechanical valve, which have not been previously assessed during its clinical history. METHODS Two-dimensional laser Doppler velocimetry was used to measure the velocity and turbulent shear stress fields in the hinge regions. To conduct these measurements, exact dimensional models of the bileaflet hinge regions were cast or machined from transparent plastic materials. The experiment was conducted in a pulsatile flow loop with measurements taken at different levels within the pivot and hinge regions. RESULTS In the 23-mm CarboMedics valve hinge, the phase-averaged forward velocity obtained at the flat level and levels of 190 microm and 390 microm above flat and 1 mm below flat were 0.54 m/s, 0.77 m/s, 0.3 m/s, and 1.0 m/s, respectively. Corresponding values of the peak phase-averaged leakage velocities were 3.17 m/s, 2.91 m/s, 2.52 m/s, and 0.5 m/s, respectively. Corresponding turbulent shear stresses were 5510 dyne/cm(2), 5640 dyne/cm(2), 4380 dyne/cm(2), and 4810 dyne/cm(2), respectively. CONCLUSIONS The hinge flow dynamics of the CarboMedics bileaflet design lie somewhere in between those of the St Jude Medical and the Medtronic Parallel valve designs. The fluid dynamics of the investigated valve were found to be similar to those of the St Jude Medical valves, although with slightly higher leakage velocities and turbulent shear stresses. This discrepancy may be a result of the sharper corners associated with the hinge design of the CarboMedics valve. It could also be due to the incremental enlargement of the internal orifice area of the St Jude Medical Regent design.

[1]  S P Sutera,et al.  Deformation and fragmentation of human red blood cells in turbulent shear flow. , 1975, Biophysical journal.

[2]  A P Yoganathan,et al.  A comparison of the hinge and near-hinge flow fields of the St Jude medical hemodynamic plus and regent bileaflet mechanical heart valves. , 2000, The Journal of thoracic and cardiovascular surgery.

[3]  H Reul,et al.  Leakage flow at mechanical heart valve prostheses: improved washout or increased blood damage? , 1999, The Journal of heart valve disease.

[4]  J. Monro,et al.  Mid-term results with 1,503 CarboMedics mechanical valve implants. , 2000, The Journal of heart valve disease.

[5]  C. E. Rasmussen,et al.  Twelve years' clinical experience with the CarboMedics prosthetic heart valve. , 2001, Journal of Heart Valve Disease.

[6]  A. Yoganathan,et al.  Pulsatile flow velocity and shear stress measurements on the St. Jude bileaflet valve prosthesis. , 1986, Scandinavian journal of thoracic and cardiovascular surgery.

[7]  Brandon R. Travis,et al.  An In Vitro Study of the Hinge and Near-Field Forward Flow Dynamics of the St. Jude Medical® Regent™ Bileaflet Mechanical Heart Valve , 2004, Annals of Biomedical Engineering.

[8]  A P Yoganathan,et al.  Velocity measurements and flow patterns within the hinge region of a Medtronic Parallel bileaflet mechanical valve with clear housing. , 1996, The Journal of heart valve disease.

[9]  J A Rooney,et al.  Hemolysis Near an Ultrasonically Pulsating Gas Bubble , 1970, Science.

[10]  C. Akins,et al.  Results with mechanical cardiac valvular prostheses. , 1995, The Annals of thoracic surgery.

[11]  R M Hochmuth,et al.  Shear-induced aggregation and lysis of platelets. , 1976, Transactions - American Society for Artificial Internal Organs.

[12]  L Zuckerman,et al.  Shear-induced activation of platelets. , 1979, Journal of biomechanics.

[13]  P. Lu,et al.  A reevaluation and discussion on the threshold limit for hemolysis in a turbulent shear flow. , 2001, Journal of biomechanics.

[14]  H N Sabbah,et al.  Measured Turbulence and Its Effect on Thrombus Formation , 1974, Circulation research.

[15]  U. Steinseifer,et al.  In vitro comparison of bileaflet aortic heart valve prostheses. St. Jude Medical, CarboMedics, modified Edwards-Duromedics, and Sorin-Bicarbon valves. , 1993, The Journal of thoracic and cardiovascular surgery.

[16]  D B Geselowitz,et al.  Estimation of Reynolds stresses within the Penn State left ventricular assist device. , 1990, ASAIO transactions.

[17]  Ajit P. Yoganathan,et al.  Visualization of the hinge flow in a 5:1 scaled model of the medtronic parallel bileaflet heart valve prosthesis , 1998 .

[18]  A P Yoganathan,et al.  A microstructural flow analysis within a bileaflet mechanical heart valve hinge. , 1996, The Journal of heart valve disease.

[19]  S. Deutsch,et al.  Laser Doppler velocimetry and flow visualization studies in the regurgitant leakage flow region of three mechanical mitral valves. , 2001, Artificial organs.

[20]  J. Zhang,et al.  CarboMedics mechanical prosthesis: performance at eight years. , 2000, The Journal of heart valve disease.

[21]  Jeffrey Thomas Ellis An in vitro investigation of the leakage and hinge flow fields through bileaflet mechanical heart valves and their relevance to thrombogenesis , 1999 .

[22]  N H Hwang,et al.  Pressure and flow fields in the hinge region of bileaflet mechanical heart valves. , 1999, The Journal of heart valve disease.

[23]  S H Chu,et al.  Turbulence characteristics downstream of bileaflet aortic valve prostheses. , 2000, Journal of biomechanical engineering.

[24]  A P Yoganathan,et al.  An in vitro investigation of the retrograde flow fields of two bileaflet mechanical heart valves. , 1996, The Journal of heart valve disease.

[25]  N H Hwang,et al.  Human red blood cell hemolysis in a turbulent shear flow: contribution of Reynolds shear stresses. , 1984, Biorheology.