Two-Wired Active Spring-Loaded Dry Electrodes for EEG Measurements

Dry contact electrode-based EEG acquisition is one of the easiest ways to obtain neural information from the human brain, providing many advantages such as rapid installation, and enhanced wearability. However, high contact impedance due to insufficient electrical coupling at the electrode-scalp interface still remains a critical issue. In this paper, a two-wired active dry electrode system is proposed by combining finger-shaped spring-loaded probes and active buffer circuits. The shrinkable probes and bootstrap topology-based buffer circuitry provide reliable electrical coupling with an uneven and hairy scalp and effective input impedance conversion along with low input capacitance. Through analysis of the equivalent circuit model, the proposed electrode was carefully designed by employing off-the-shelf discrete components and a low-noise zero-drift amplifier. Several electrical evaluations such as noise spectral density measurements and input capacitance estimation were performed together with simple experiments for alpha rhythm detection. The experimental results showed that the proposed electrode is capable of clear detection for the alpha rhythm activation, with excellent electrical characteristics such as low-noise of 1.131 μVRMS and 32.3% reduction of input capacitance.

[1]  Daniel Sánchez Morillo,et al.  Dry EEG Electrodes , 2014, Sensors.

[2]  Conor O'Mahony,et al.  Microneedle-Based Electrodes with Integrated Through-Silicon via for Biopotential Recording , 2011 .

[3]  S. Coyle,et al.  Brain–computer interfaces: a review , 2003 .

[4]  Peter Achermann,et al.  A model of human sleep homeostasis based on EEG slow-wave activity: Quantitative comparison of data and simulations , 1993, Brain Research Bulletin.

[5]  Sang-Hoon Lee,et al.  A capacitive, biocompatible and adhesive electrode for long-term and cap-free monitoring of EEG signals. , 2013, Journal of neural engineering.

[6]  Yong Zhu,et al.  Nanomaterial-Enabled Dry Electrodes for Electrophysiological Sensing: A Review , 2016, JOM.

[7]  S. Smith EEG in the diagnosis, classification, and management of patients with epilepsy , 2005, Journal of Neurology, Neurosurgery & Psychiatry.

[8]  Penghai Li,et al.  Fabrication of chitosan/Au-TiO2 nanotube-based dry electrodes for electroencephalography recording. , 2017, Materials science & engineering. C, Materials for biological applications.

[9]  Chris Van Hoof,et al.  Soft, Comfortable Polymer Dry Electrodes for High Quality ECG and EEG Recording , 2014, Sensors.

[10]  Xiaowei Yang,et al.  A Motion Interference-Insensitive Flexible Dry Electrode , 2016, IEEE Transactions on Biomedical Engineering.

[11]  Sang-Hoon Lee,et al.  Self-Adhesive and Capacitive Carbon Nanotube-Based Electrode to Record Electroencephalograph Signals From the Hairy Scalp , 2016, IEEE Transactions on Biomedical Engineering.

[12]  Jens Haueisen,et al.  Contact Pressure and Flexibility of Multipin Dry EEG Electrodes , 2018, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[13]  T. Jung,et al.  Dry and Noncontact EEG Sensors for Mobile Brain–Computer Interfaces , 2012, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[14]  Chin-Teng Lin,et al.  Design, Fabrication and Experimental Validation of a Novel Dry-Contact Sensor for Measuring Electroencephalography Signals without Skin Preparation , 2011, Sensors.

[15]  Alexander J. Casson,et al.  3D Printed Dry EEG Electrodes , 2016, Sensors.

[16]  Roberto Guerrieri,et al.  Active Electrode IC for EEG and Electrical Impedance Tomography With Continuous Monitoring of Contact Impedance , 2015, IEEE Transactions on Biomedical Circuits and Systems.

[17]  Guangli Li,et al.  Novel passive ceramic based semi-dry electrodes for recording electroencephalography signals from the hairy scalp , 2016 .

[18]  Bin Yang,et al.  A Novel Bristle-Shaped Semi-Dry Electrode With Low Contact Impedance and Ease of Use Features for EEG Signal Measurements , 2020, IEEE Transactions on Biomedical Engineering.

[19]  Rasoul Dehghani,et al.  Two-Wired Current Modulator Active Electrode for Ambulatory Biosignal Recording , 2019, IEEE Transactions on Biomedical Circuits and Systems.

[20]  Anil Kumar,et al.  Dry Electrode-Based Fully Isolated EEG/fNIRS Hybrid Brain-Monitoring System , 2019, IEEE Transactions on Biomedical Engineering.

[21]  An Application Guide for Op Amps , 1995 .

[22]  José Higino Correia,et al.  New dry electrodes based on iridium oxide (IrO) for non-invasive biopotential recordings and stimulation , 2010 .

[23]  Reza Moghimi Low Noise Signal Conditioning for Sensor-Based Circuits , 2010 .

[24]  K. Knuth,et al.  What Is a Question , 2004, physics/0403089.

[25]  W. Zhou,et al.  A Mini Review of Microneedle Array Electrode for Bio-Signal Recording: A Review , 2020, IEEE Sensors Journal.

[26]  G. Ruffinia,et al.  First human trials of a dry electrophysiology sensor using a carbon nanotube array interface , 2008 .

[27]  Yong Gyu Lim,et al.  Comparison of pre-amplifier topologies for use in brain-computer interface with capacitively-coupled EEG electrodes , 2013 .

[28]  Refet Firat Yazicioglu,et al.  Active Electrodes for Wearable EEG Acquisition: Review and Electronics Design Methodology , 2017, IEEE Reviews in Biomedical Engineering.

[29]  C. Neuper,et al.  Combining Brain–Computer Interfaces and Assistive Technologies: State-of-the-Art and Challenges , 2010, Front. Neurosci..

[30]  C. Grozea,et al.  Bristle-sensors—low-cost flexible passive dry EEG electrodes for neurofeedback and BCI applications , 2011, Journal of neural engineering.

[31]  Enrique Mario Spinelli,et al.  A Two-Wired Ultra-High Input Impedance Active Electrode , 2018, IEEE Transactions on Biomedical Circuits and Systems.

[32]  Tzyy-Ping Jung,et al.  Dry-Contact and Noncontact Biopotential Electrodes: Methodological Review , 2010, IEEE Reviews in Biomedical Engineering.

[33]  Heinz Jäckel,et al.  Low-Noise Two-Wired Buffer Electrodes for Bioelectric Amplifiers , 2007, IEEE Transactions on Biomedical Engineering.

[34]  T. J. Sullivan,et al.  A Low-Noise, Non-Contact EEG/ECG Sensor , 2007, 2007 IEEE Biomedical Circuits and Systems Conference.

[35]  Gert Cauwenberghs,et al.  Ultra-High Input Impedance, Low Noise Integrated Amplifier for Noncontact Biopotential Sensing , 2011, IEEE Journal on Emerging and Selected Topics in Circuits and Systems.

[36]  Peter Enoksson,et al.  Micromachined electrodes for biopotential measurements , 2001 .

[37]  Heinz Jäckel,et al.  A pseudodifferential amplifier for bioelectric events with DC-offset compensation using two-wired amplifying electrodes , 2006, IEEE Transactions on Biomedical Engineering.

[38]  S. Lanyi The noise of input stages with low parasitic capacitance , 2001 .

[39]  Bor-Shyh Lin,et al.  Novel Active Comb-Shaped Dry Electrode for EEG Measurement in Hairy Site , 2015, IEEE Transactions on Biomedical Engineering.

[40]  Yong Gyu Lim,et al.  Brain–computer interfaces using capacitive measurement of visual or auditory steady-state responses , 2013, Journal of neural engineering.