An initial investigation of a wireless patient radiation dosimeter for use in interventional radiology

Radiation exposure during interventional radiology (IR) procedures is a critical issue. We have developed a wireless real-time dosimeter for IR patients that use nontoxic phosphor (four sensors). We evaluated the basic performance parameters (such as dose linearity, batch uniformity, reproducibility, and wireless-communication conditions) of the developed system using an IR X-ray system. Further, we investigated the influence of noise from other medical equipment on our wireless real-time dosimeter in the IR X-ray room. Overall, our wireless system exhibited excellent performance in terms of uniformity, reproducibility, and linearity; moreover, the wireless communication performance was better. The developed system enabled real-time visualization of patient radiation dose, without noise contamination from other medical equipment. In addition, the wireless system can be easily installed in a location where the PC screen (display) can be readily viewed by the IR physician. Hence, we developed a wireless system that can display the patient radiation dose data in real time; the system performed satisfactorily upon application in radiation dosimetry. Therefore, our wireless system will facilitate the real-time monitoring/management of patient radiation dose during IR.

[1]  Masayuki Zuguchi,et al.  Real-time patient radiation dosimeter for use in interventional radiology. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[2]  K. Chida,et al.  A PHANTOM STUDY TO DETERMINE THE OPTIMAL PLACEMENT OF EYE DOSEMETERS ON INTERVENTIONAL CARDIOLOGY STAFF. , 2019, Radiation protection dosimetry.

[3]  S. Perrone,et al.  Risk of radiation-induced lens opacities among surgeons and interventional medical staff , 2018, Radiological Physics and Technology.

[4]  R. Sievert,et al.  Book Reviews : Recommendations of the International Commission on Radiological Protection (as amended 1959 and revised 1962). I.C.R.P. Publication 6. 70 pp. PERGAMON PRESS. Oxford, London and New York, 1964. £1 5s. 0d. [TB/54] , 1964 .

[5]  Masayuki Zuguchi,et al.  Effectiveness of a novel real-time dosimeter in interventional radiology: a comparison of new and old radiation sensors , 2018, Radiological Physics and Technology.

[6]  Wesley E. Bolch,et al.  Physical validation of UF‐RIPSA: A rapid in‐clinic peak skin dose mapping algorithm for fluoroscopically guided interventions , 2018, Journal of applied clinical medical physics.

[7]  Koichi Chida,et al.  New real-time patient radiation dosimeter for use in radiofrequency catheter ablation , 2019, Journal of radiation research.

[8]  E Gaxiola,et al.  Real-time measurement of skin radiation during cardiac catheterization. , 1998, Catheterization and cardiovascular diagnosis.

[9]  Eliseo Vano,et al.  Importance of a Patient Dosimetry and Clinical Follow-up Program in the Detection of Radiodermatitis After Long Percutaneous Coronary Interventions , 2013, CardioVascular and Interventional Radiology.

[10]  G. Reitz,et al.  ICRP Publication 123: Assessment of Radiation Exposure of Astronauts in Space , 2013 .

[11]  Masayuki Zuguchi,et al.  Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures. , 2006, AJR. American journal of roentgenology.

[12]  Masayuki Zuguchi,et al.  Radiation dose to the pediatric cardiac catheterization and intervention patient. , 2010, AJR. American journal of roentgenology.

[13]  J. Eccles,et al.  International electrotechnical commission , 1955, Journal of the American Institute of Electrical Engineers.

[14]  Masayuki Zuguchi,et al.  Evaluating the performance of a MOSFET dosimeter at diagnostic X-ray energies for interventional radiology , 2009, Radiological physics and technology.

[15]  Jack Valentin,et al.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. , 2007, Annals of the ICRP.

[16]  K. Chida,et al.  Red emission phosphor for real-time skin dosimeter for fluoroscopy and interventional radiology. , 2014, Medical physics.

[17]  Masayuki Zuguchi,et al.  Novel Dosimeter Using a Nontoxic Phosphor for Real-Time Monitoring of Patient Radiation Dose in Interventional Radiology. , 2015, AJR. American journal of roentgenology.

[18]  Koichi Chida,et al.  Evaluating the maximum patient radiation dose in cardiac interventional procedures. , 2011, Radiation protection dosimetry.

[19]  Masayuki Zuguchi,et al.  A cross‐sectional study of the radiation dose and image quality of X‐ray equipment used in IVR , 2016, Journal of applied clinical medical physics.

[20]  K. Chida,et al.  Performance of the DOSIRIS™ eye lens dosimeter , 2019, Journal of radiological protection : official journal of the Society for Radiological Protection.

[21]  Masayuki Zuguchi,et al.  Comparison of dose at an interventional reference point between the displayed estimated value and measured value , 2011, Radiological physics and technology.

[22]  Masayuki Zuguchi,et al.  Occupational eye dose in interventional cardiology procedures , 2017, Scientific Reports.

[23]  B Geiger,et al.  ICRP Publication 120: Radiological Protection in Cardiology , 2013, Annals of the ICRP.

[24]  Masayuki Zuguchi,et al.  Fundamental study of a real-time occupational dosimetry system for interventional radiology staff , 2014, Journal of radiological protection : official journal of the Society for Radiological Protection.

[25]  K. Nishizawa,et al.  Dose measurement for patients and physicians using a glass dosemeter during endovascular treatment for brain disease. , 2003, Radiation protection dosimetry.

[26]  J. Valentin,et al.  Abstract: Avoidance of radiation injuries from medical interventional procedures, ICRP Publication 85 , 2000 .

[27]  Jack Valentin,et al.  News of the world , 2000 .

[28]  Masayuki Zuguchi,et al.  Total entrance skin dose: an effective indicator of maximum radiation dose to the skin during percutaneous coronary intervention. , 2007, AJR. American journal of roentgenology.

[29]  Adrian K. Dixon,et al.  Benefits and costs, an eternal balance , 2007 .

[30]  Masayuki Zuguchi,et al.  Does digital acquisition reduce patients' skin dose in cardiac interventional procedures? An experimental study. , 2004, AJR. American journal of roentgenology.

[31]  T. Ishibashi,et al.  Radiation dose and radiation protection for patients and physicians during interventional procedure. , 2010, Journal of radiation research.

[32]  K. Chida,et al.  Clarifying and visualizing sources of staff-received scattered radiation in interventional procedures. , 2011, AJR. American journal of roentgenology.

[33]  Masayuki Zuguchi,et al.  Radiation dose of interventional radiology system using a flat-panel detector. , 2009, AJR. American journal of roentgenology.

[34]  T. Ishibashi,et al.  Evaluation of patient radiation dose during cardiac interventional procedures: What is the most effective method? , 2009, Acta radiologica.

[35]  Koichi Chida,et al.  The necessity of follow-up for radiation skin injuries in patients after percutaneous coronary interventions: radiation skin injuries will often be overlooked clinically , 2012, Acta radiologica.

[36]  H. Toyoshima,et al.  OCCUPATIONAL RADIATION EXPOSURE DOSE OF THE EYE IN DEPARTMENT OF CARDIAC ARRHYTHMIA PHYSICIAN. , 2019, Radiation protection dosimetry.

[37]  Masayuki Zuguchi,et al.  Usefulness of non-lead aprons in radiation protection for physicians performing interventional procedures. , 2008, Radiation protection dosimetry.

[38]  E. Vañó,et al.  Patient dose values in interventional radiology. , 1995, The British journal of radiology.

[39]  Koichi Chida,et al.  Estimation of the Dose of Radiation Received by Patient and Physician During a Videofluoroscopic Swallowing Study , 2016, Dysphagia.

[40]  Masayuki Zuguchi,et al.  Occupational dose in interventional radiology procedures. , 2013, AJR. American journal of roentgenology.

[41]  H. Toyoshima,et al.  OCCUPATIONAL RADIATION EXPOSURE OF THE EYE IN NEUROVASCULAR INTERVENTIONAL PHYSICIAN. , 2019, Radiation protection dosimetry.

[42]  N. Yanagawa,et al.  Evaluation of scattered radiation from fluoroscopy using small OSL dosimeters , 2019, Radiological Physics and Technology.

[43]  J. Valentin,et al.  Don't step on it , 2000, Annals of the ICRP.

[44]  Masayuki Zuguchi,et al.  Comparison of the radiation dose in a cardiac IVR X-ray system. , 2011, Radiation protection dosimetry.