Measurement of intraocular pressure with pressure phosphene tonometry in children.

PURPOSE To study the accuracy and acceptability of intraocular pressure (IOP) measurement by the pressure phosphene tonometer, non-contact tonometer, and Goldmann tonometer in children. METHODS Fifty children (5 to 14 years old) participated in this prospective comparative study. IOP was measured with the pressure phosphene tonometer, non-contact tonometer, and Goldmann tonometer by three different examiners who were masked to the results. The children were also asked to grade the degree of discomfort from 0 to 5 (0 = no discomfort; 5 = most discomfort). RESULTS The mean IOPs measured by the Goldmann tonometer, pressure phosphene tonometer, and non-contact tonometer were 15.9 mm Hg (standard deviation [SD]: = 5.5 mm Hg; range: 10 to 36 mm Hg), 16.0 mm Hg (SD: 2.9 mm Hg; range: 12 to 25 mm Hg), and 15.7 mm Hg (SD = 5.1 mm Hg; range: 8 to 32 mm Hg), respectively (P = .722). The mean difference between pressure phosphene tonometer and Goldmann tonometer readings was 2.9 mm Hg and that between non-contact tonometer and Goldmann tonometer readings was 2.1 mm Hg. The 95% confidence interval of the mean difference between pressure phosphene tonometer and Goldmann tonometer readings was -1.07 and 1.19, and that between non-contact tonometer and Goldmann tonometer readings was -1.07 and 0.53. The mean discomfort ratings for the pressure phosphene tonometer, non-contact tonometer, and Goldmann tonometer were 0.6, 2.0, and 2.3, respectively (P < .001). CONCLUSION Although the pressure phosphene tonometer was less accurate than the non-contact tonometer compared with Goldmann tonometer, it gave a reasonably close estimate and had a high specificity of raised IOP. In addition, measurement by the pressure phosphene tonometer is most acceptable to children. The pressure phosphene tonometer can be considered as an alternative method of IOP measurement in children.

[1]  J. Schuman,et al.  Corneal thickness measurement in the management of primary open-angle glaucoma: a report by the American Academy of Ophthalmology. , 2007, Ophthalmology.

[2]  A. Şahin,et al.  Effect of central corneal thickness and radius of the corneal curvature on intraocular pressure measured with the Tono-Pen and noncontact tonometer in healthy schoolchildren. , 2007, Journal of pediatric ophthalmology and strabismus.

[3]  R. Fechtner,et al.  Evolving global risk assessment of ocular hypertension to glaucoma , 2007, Current opinion in ophthalmology.

[4]  P. Debry,et al.  Comparison of Proview phosphene tonometry with Goldmann applanation tonometry. , 2006, Canadian journal of ophthalmology. Journal canadien d'ophtalmologie.

[5]  P. Gunvant,et al.  Effect of Proview self‐tonometry on pharmaceutical compliance , 2006, Clinical and experimental optometry.

[6]  W. Swanson,et al.  A Clinical Evaluation of Proview Pressure Phosphene Tonometry in Children , 2006, Optometry and vision science : official publication of the American Academy of Optometry.

[7]  Barry YM Yeung,et al.  Comparison between central corneal thickness measurements by ultrasound pachymetry and optical coherence tomography , 2006, Clinical & experimental ophthalmology.

[8]  D. Y. Leung,et al.  Pressure phosphene self‐tonometry in a patient with Posner–Schlossman syndrome , 2006, Clinical & experimental ophthalmology.

[9]  P. Herse,et al.  The Proview Eye Pressure Monitor: influence of clinical factors on accuracy and agreement with the Goldmann tonometer , 2005, Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians.

[10]  J. Ng,et al.  Ocular hypertensive and anti‐inflammatory responses to different dosages of topical dexamethasone in children: a randomized trial , 2005, Clinical & experimental ophthalmology.

[11]  G. Sanderson,et al.  The pressure phosphene tonometer—a clinical evaluation , 2005, Eye.

[12]  T. Wong,et al.  Pressure phosphene self-tonometry: a comparison with goldmann tonometry in glaucoma patients. , 2004, Investigative ophthalmology & visual science.

[13]  R. Ritch,et al.  The Proview phosphene tonometer fails to measure ocular pressure accurately in clinical practice. , 2004, Ophthalmology.

[14]  G. Gamble,et al.  Comparison of the Proview pressure phosphene tonometer performed by the patient and examiner with the Goldmann applanation tonometer , 2004, Clinical & experimental ophthalmology.

[15]  C. Pang,et al.  Ocular-hypertensive and anti-inflammatory response to rimexolone therapy in children. , 2003, Archives of ophthalmology.

[16]  L. Tychsen,et al.  The effect of an eyelid speculum on intraocular pressure measurement in children. , 2002, American journal of ophthalmology.

[17]  D. Lam,et al.  A prospective study on ocular hypertensive and antiinflammatory response to different dosages of fluorometholone in children. , 2001, Ophthalmology.

[18]  R. Ritch,et al.  Attempted eyelid closure affects intraocular pressure measurement. , 2001, American journal of ophthalmology.

[19]  D. Lam,et al.  Ocular hypertensive response to topical dexamethasone in children: a dose-dependent phenomenon. , 2000, Ophthalmology.

[20]  M. Snir,et al.  Intraocular pressure fluctuations during strabismus operations and the postoperative period. , 1999, Ophthalmic surgery and lasers.

[21]  B. B. Fresco A new tonometer--the pressure phosphene tonometer: clinical comparison with Goldman tonometry. , 1998, Ophthalmology.

[22]  M. Tso,et al.  Ocular-hypertensive response to topical steroids in children. , 1997, Ophthalmology.

[23]  T. Hirose,et al.  Tonometry in pediatric patients: a comparative study among Tono-pen, Perkins, and Schiötz tonometers. , 1995, Journal of pediatric ophthalmology and strabismus.

[24]  M. Jaafar,et al.  Normal intraocular pressure in children: a comparative study of the Perkins applanation tonometer and the pneumatonometer. , 1993, Journal of pediatric ophthalmology and strabismus.

[25]  P. Nucci,et al.  Noncontact tonometry in children. , 1991, Optometry and Vision Science.

[26]  O. Grüsser,et al.  Responses of retinal ganglion cells to eyeball deformation: A neurophysiological basis for “pressure phosphenes” , 1989, Vision Research.

[27]  P. Lichter,et al.  Clinical comparison of the Oculab Tono-Pen to the Goldmann applanation tonometer. , 1987, Ophthalmology.

[28]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[29]  C. Krakau,et al.  EFFECTS OF LOADING OF THE EYE ON THE INTRAOCULAR PRESSURE AND ON THE EPISCLERAL VENOUS PRESSURE , 1974, Acta ophthalmologica.

[30]  R. Moses The Goldmann applanation tonometer. , 1958, American journal of ophthalmology.

[31]  L. E. Lipetz The X Ray and Radium Phosphenes * , 1955, The British journal of ophthalmology.

[32]  G. Brindley,et al.  The site of electrical excitation of the human eye , 1955, The Journal of physiology.

[33]  R. Sihota,et al.  Distribution and determinants of intraocular pressure in a normal pediatric population. , 2006, Journal of pediatric ophthalmology and strabismus.

[34]  M. Kass Standardizing the measurement of intraocular pressure for clinical research. Guidelines from the Eye Care Technology Forum. , 1996, Ophthalmology.

[35]  S. Vernon,et al.  Deterioration in the accuracy of the Pulsair non-contact tonometer with use: Need for regular calibration , 1992, Eye.