Clinical and radiographic outcomes of minimally invasive percutaneous pedicle screw placement with intraoperative CT (O-arm) image guidance navigation.

OBJECT Intraoperative CT image-guided navigation (IGN) has been increasingly incorporated into minimally invasive spine surgery (MIS). The vast improvement in image resolution and virtual real-time images with CT-IGN has proven superiority over traditional fluoroscopic techniques. The authors describe their perioperative MIS technique using the O-arm with navigation, and they report their postoperative experience, accuracy results, and technical aspects. METHODS A retrospective review of 48 consecutive adult patients undergoing minimally invasive percutaneous posterior spinal fusion with intraoperative CT-IGN between July 2010 and August 2013 at Cedars-Sinai Medical Center was performed. Two surgeons assessed 290 screws in a blinded fashion on intraoperative O-arm images and postoperative CT scans for bony pedicle wall breach. Grade 1 breach was defined to be < 2 mm, Grade 2 breach to be between 2 and 4 mm, and a Grade 3 breach to be > 4 mm. Additionally, anterior vertebral body breach was recorded. RESULTS Of 290 pedicle screws placed, 280 (96.6%) were in an acceptable position without cortical wall or anterior breach. Of the 10 breaches (3.4%) 5 were lateral (50%), 4 were medial, and 1 was anterior; 90% of breaches were Grade 1-2 and all medial breaches were Grade 1. The one Grade 3 breach was lateral. No vascular or neurological complications were observed intraoperatively, and no significant postoperative complications were noted. The mean clinical follow-up period was 18 months (range 3-39 months). The overall clinical outcomes, measured using the visual analog scale (back pain scores), were improved significantly postoperatively at 3 months compared with preoperatively (visual analog score 6.35 vs 3.57; p < 0.0001). No revision surgery was performed for screw misplacement or neurological deterioration. CONCLUSIONS New CT-IGN with the mobile O-arm scanner has increased the accuracy of pedicle screw/instrumentation placement using MIS techniques. The authors' high (96.6%) accuracy rate in MIS compares favorably with historical published accuracy rates for fluoroscopy-based techniques. Additional advantages of CT-IGN over fluoroscopic imaging methods are lower occupational radiation exposure for the surgical team, reduced need for postoperative imaging, and decreased rates of revision surgery. For now, the authors simply conclude that use of intraoperative CT-IGN is safe and accurate.

[1]  J. Knox,et al.  Superior segment facet joint violation and cortical violation after minimally invasive pedicle screw placement. , 2011, The spine journal : official journal of the North American Spine Society.

[2]  Donald A. Smith,et al.  O-Arm-based image guidance in minimally invasive spine surgery: Technical note , 2013, Clinical Neurology and Neurosurgery.

[3]  R. Castelein,et al.  Complications of pedicle screws in lumbar and lumbosacral fusions in 105 consecutive primary operations , 2002, European Spine Journal.

[4]  A. Alho,et al.  Accuracy of pedicle screw insertion: A prospective CT study in 30 low back patients , 2005, European Spine Journal.

[5]  H. Richter,et al.  Percutaneous transarticular atlantoaxial screw fixation using a cannulated screw system and image guidance. , 2004, Minimally invasive neurosurgery : MIN.

[6]  P. Robertson,et al.  Radiation Exposure During Fluoroscopically Assisted Pedicle Screw Insertion in the Lumbar Spine , 2000, Spine.

[7]  David W. Polly,et al.  Pediatric Pedicle Screw Placement Using Intraoperative Computed Tomography and 3-Dimensional Image-Guided Navigation , 2012, Spine.

[8]  Charles Kuntz,et al.  Accuracy of Thoracic Vertebral Body Screw Placement Using Standard Fluoroscopy, Fluoroscopic Image Guidance, and Computed Tomographic Image Guidance: A Cadaver Study , 2003, Spine.

[9]  Ariel Roguin,et al.  Brain and neck tumors among physicians performing interventional procedures. , 2013, The American journal of cardiology.

[10]  Ping Zhou,et al.  Pedicle screw insertion accuracy with different assisted methods: a systematic review and meta-analysis of comparative studies , 2011, European Spine Journal.

[11]  Nicholas Theodore,et al.  Placement of percutaneous thoracic pedicle screws using neuronavigation. , 2010, World neurosurgery.

[12]  Vincent Arlet,et al.  Complications of Pedicle Screw Fixation in Scoliosis Surgery: A Systematic Review , 2010, Spine.

[13]  D. E. Casey Jones,et al.  Accuracy of Pedicle Screw Placement in Lumbar Fusions by Plain Radiographs and Computed Tomography , 1995, Spine.

[14]  D. Polly,et al.  Comparison of Cranial Facet Joint Violation Rates Between Open and Percutaneous Pedicle Screw Placement Using Intraoperative 3-D CT (O-arm) Computer Navigation , 2013, Spine.

[15]  P M Rommens,et al.  Radiation exposure to the hands and the thyroid of the surgeon during intramedullary nailing. , 1998, Injury.

[16]  X. Weng,et al.  [Complications associated with the technique of pedicle screw fixation]. , 2002, Zhongguo yi xue ke xue yuan xue bao. Acta Academiae Medicinae Sinicae.

[17]  W. Castro,et al.  Accuracy of Pedicle Screw Placement in Lumbar Vertebrae , 1996, Spine.

[18]  S. Esses,et al.  Complications associated with the technique of pedicle screw fixation. A selected survey of ABS members. , 1993, Spine.

[19]  Kevin T Foley,et al.  Minimally Invasive Lumbar Fusion , 2003, Spine.

[20]  R. Härtl,et al.  Minimal Access Versus Open Transforaminal Lumbar Interbody Fusion: Meta-Analysis of Fusion Rates , 2010, Spine.

[21]  C. Krettek,et al.  Navigated minimally invasive thoracolumbar pedicle screw placement with flat panel 3-D imaging. A feasibility study. , 2010, Technology and health care : official journal of the European Society for Engineering and Medicine.

[22]  M. O’Brien,et al.  Radiation Exposure During Pedicle Screw Placement in Adolescent Idiopathic Scoliosis: Is Fluoroscopy Safe? , 2006, Spine.

[23]  S. Gertzbein,et al.  Accuracy of Pedicular Screw Placement In Vivo , 1990, Spine.

[24]  R. Härtl,et al.  Pedicle screw navigation: a systematic review and meta-analysis of perforation risk for computer-navigated versus freehand insertion. , 2012, Journal of neurosurgery. Spine.

[25]  Rani Nasser,et al.  Clinical Assessment of Percutaneous Lumbar Pedicle Screw Placement Using the O-Arm Multidimensional Surgical Imaging System , 2012, Neurosurgery.

[26]  Gordon Singer,et al.  Occupational Radiation Exposure to the Surgeon , 2005, The Journal of the American Academy of Orthopaedic Surgeons.

[27]  H. Hashidate,et al.  Accuracy of Multilevel Registration in Image-Guided Pedicle Screw Insertion for Adolescent Idiopathic Scoliosis , 2010, Spine.

[28]  S. Chung,et al.  Comparison of Multifidus Muscle Atrophy and Trunk Extension Muscle Strength: Percutaneous Versus Open Pedicle Screw Fixation , 2005, Spine.

[29]  Boram Lee,et al.  Measurements of Surgeons' Exposure to Ionizing Radiation Dose During Intraoperative Use of C-Arm Fluoroscopy , 2012, Spine.

[30]  S. Hovsepian,et al.  The relationship between occupational radiation exposure and thyroid nodules , 2012, Journal of research in medical sciences : the official journal of Isfahan University of Medical Sciences.

[31]  P. Huddleston Facet Violation With the Placement of Percutaneous Pedicle Screws , 2012 .

[32]  M. Gutberlet,et al.  Computer tomography assessment of pedicle screw placement in lumbar and sacral spine: comparison between free-hand and O-arm based navigation techniques , 2011, European Spine Journal.

[33]  S. Dhall,et al.  Clinical and radiographic comparison of mini-open transforaminal lumbar interbody fusion with open transforaminal lumbar interbody fusion in 42 patients with long-term follow-up. , 2008, Journal of neurosurgery. Spine.

[34]  Jean-Paul Wolinsky,et al.  The efficacy of minimally invasive discectomy compared with open discectomy: a meta-analysis of prospective randomized controlled trials. , 2012, Journal of neurosurgery. Spine.

[35]  I. Lieberman,et al.  Radiation Exposure to the Surgeon During Percutaneous Pedicle Screw Placement , 2011, Journal of spinal disorders & techniques.

[36]  J. Rueger,et al.  Comparison of open versus percutaneous pedicle screw insertion in a sheep model , 2008, European Spine Journal.

[37]  Y. Rampersaud,et al.  Radiation Exposure to the Spine Surgeon During Fluoroscopically Assisted Pedicle Screw Insertion , 2000, Spine.

[38]  W. Truong,et al.  The Accuracy of Intraoperative O-arm Images for the Assessment of Pedicle Screw Postion , 2012, Spine.

[39]  R. Härtl,et al.  Navigated guide tube for the placement of mini-open pedicle screws using stereotactic 3D navigation without the use of K-wires: technical note. , 2013, Journal of neurosurgery. Spine.

[40]  The Removal of a Transdural Pedicle Screw Placed for Thoracolumbar Spine Fracture , 1996, Spine.

[41]  H. Zeeb,et al.  Testicular cancer risk associated with occupational radiation exposure: a systematic literature review , 2010, Journal of radiological protection : official journal of the Society for Radiological Protection.

[42]  F. Costa,et al.  A Prospective Multicenter Registry on the Accuracy of Pedicle Screw Placement in the Thoracic, Lumbar, and Sacral Levels With the Use of the O-arm Imaging System and StealthStation Navigation , 2012, Spine.

[43]  N. Theocharopoulos,et al.  Occupational exposure from common fluoroscopic projections used in orthopaedic surgery. , 2003, The Journal of bone and joint surgery. American volume.

[44]  S. Chung,et al.  Validation of the Korean Version of the Oswestry Disability Index , 2005, Spine.

[45]  V. Schreiber,et al.  Clinical and methodological precision of spinal navigation assisted by 3D intraoperative O-arm radiographic imaging. , 2011, Journal of neurosurgery. Spine.

[46]  Yan Wang,et al.  Application of Intraoperative Computed Tomography With or Without Navigation System in Surgical Correction of Spinal Deformity: A Preliminary Result of 59 Consecutive Human Cases , 2012, Spine.