Detection Of Hydrogen From Dissolved Gases In Transformer Oil By Pd/Pt Misfet
暂无分享,去创建一个
This paper presents a MISFET type hydrogen sensor for detecting dissolved hydrogen in transformer isolation oil. Sensing and reference FETs, a buried resistor type heater, and a p-n junction diode for the temperature control are integrated on a single chip. Measurements are carried out at various hydrogen concentrations in transformer isolation oil. The results show that this detector can be used as a practical hydrogen sensor with an advantage of in-situ real time gas analyzing.
[1] C. Svensson,et al. A hydrogen-sensitive Pd-gate MOS transistor , 1975 .
[2] Ingemar Lundström,et al. Hydrogen sensitive mos-structures: Part 1: Principles and applications , 1981 .
[3] Ingemar Lundström,et al. Hydrogen sensitive mos-structures part 2: characterization , 1981 .
[4] I. Lundström,et al. Fish scales as biosensors , 1990 .