Generation and verification of QFix kVue Calypso‐compatible couch top model for a dedicated stereotactic linear accelerator with FFF beams

This study details the generation, verification, and implementation of a treatment planning system (TPS) couch top model for patient support system used in conjunction with a dedicated stereotactic linear accelerator. Couch top model was created within the TPS using CT simulation images of the kVue Calpyso‐compatible couchtop (with rails). Verification measurements were compared to TPS dose prediction for different energies (6 MV FFF and 10 MV FFF) and rail configurations (rails in and rails out) using: 1) central axis point‐dose measurements with pinpoint chamber in water‐equivalent phantom at 42 gantry angles for various field sizes (2×2 cm2,4×4 cm2,10×10 cm2); and 2) Gafchromic EBT3 film parallel to beam in acrylic slab to assess changes in surface and percent depth doses in PA geometry. To assess sensitivity of delivered dose to variations in patient lateral position, measurements at central axis using the pinpoint chamber geometry were taken at lateral couch displacements of 2, 5, and 10 mm for 6 MV FFF. The maximum percent difference for point‐dose measurements was 3.24% (6 MV FFF) and 2.30% (10 MV FFF). The average percent difference for point‐dose measurements was less than 1.10% for all beam energies and rail geometries. The maximum percent difference between calculated and measured dose can be as large as 13.0% if no couch model is used for dose calculation. The presence of the couch structures also impacts surface dose and PDD, which was evaluated with Gafchromic film measurements. The upstream shift in the depth of dose maximum (dmax) was found to be 10.5 mm for 6 MV FFF and 5.5 mm for 10 MV FFF for ‘Rails In’ configuration. Transmission of the treatment beam through the couch results in an increase in surface dose (absolute percentage) of approximately 50% for both photon energies (6 MV FFF and 10 MV FFF). The largest sensitivity to lateral shifts occurred at the lateral boundary of the rail structures. The mean magnitude (standard deviation) of the deviation between shifted and centered measurements over all field sizes tested was 0.61% (0.61%) for 2 mm shifts, 0.46% (0.67%) for 5 mm shifts, and 0.86% (1.46%) for 10 mm shifts. PACS numbers: 87.56.‐v, 87.56.Da, 87.56.Fc

[1]  Darby Erler,et al.  A Comparison of Two Immobilization Systems for Stereotactic Body Radiation Therapy (SBRT) of Lung Tumors , 2009 .

[2]  M S Muthuswamy,et al.  A method of beam-couch intersection detection. , 1999, Medical physics.

[3]  Martin Fuss,et al.  Implications of a high-definition multileaf collimator (HD-MLC) on treatment planning techniques for stereotactic body radiation therapy (SBRT): a planning study. , 2009, Radiation oncology.

[4]  Emiliano Spezi,et al.  Dosimetric characteristics of the Siemens IGRT carbon fiber tabletop. , 2007, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[5]  [Evaluation of a new table, Exact IGRT(Varian), on the treatment beam attenuation and the image quality]. , 2010, Cancer radiotherapie : journal de la Societe francaise de radiotherapie oncologique.

[6]  J. Seppälä,et al.  Increased beam attenuation and surface dose by different couch inserts of treatment tables used in megavoltage radiotherapy , 2011, Journal of applied clinical medical physics.

[7]  Matthias Guckenberger,et al.  Precision of Image-Guided Radiotherapy (IGRT) in Six Degrees of Freedom and Limitations in Clinical Practice , 2007, Strahlentherapie und Onkologie.

[8]  Hui Yan,et al.  Dosimetric characteristics of novalis Tx system with high definition multileaf collimator. , 2008, Medical physics.

[9]  Jennifer Diffey,et al.  The effect of gantry angle on megavoltage photon beam attenuation by a carbon fiber couch insert. , 2005, Medical physics.

[10]  Assessment and management of radiotherapy beam intersections with the treatment couch , 2010, Journal of applied clinical medical physics.

[11]  David Wilkins,et al.  Investigating treatment dose error due to beam attenuation by a carbon fiber tabletop , 2006, Journal of applied clinical medical physics.

[12]  E. Moros,et al.  Modeling of carbon fiber couch attenuation properties with a commercial treatment planning system. , 2008, Medical physics.

[13]  T. Kataria,et al.  Impact of 6MV photon beam attenuation by carbon fiber couch and immobilization devices in IMRT planning and dose delivery , 2006, Journal of medical physics.

[14]  J. Wong,et al.  Flat-panel cone-beam computed tomography for image-guided radiation therapy. , 2002, International journal of radiation oncology, biology, physics.

[15]  The effect of carbon fiber couch inserts on surface dose with beam size variation. , 2001, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[16]  Emiliano Spezi,et al.  Evaluating the influence of the Siemens IGRT carbon fibre tabletop in head and neck IMRT. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[17]  J Pouliot,et al.  Electronic portal imaging device detection of radioopaque markers for the evaluation of prostate position during megavoltage irradiation: a clinical study. , 1997, International journal of radiation oncology, biology, physics.

[18]  M. V. van Herk,et al.  Physical aspects of a real-time tumor-tracking system for gated radiotherapy. , 2000, International journal of radiation oncology, biology, physics.

[19]  Évaluation d’une nouvelle table, Exact® IGRT (Varian), sur l’atténuation du faisceau de traitement et la qualité des images , 2010 .

[20]  Radhe Mohan,et al.  Dosimetric properties of photon beams from a flattening filter free clinical accelerator , 2005, Physics in medicine and biology.

[21]  David Followill,et al.  The clinical impact of the couch top and rails on IMRT and arc therapy , 2011, Physics in medicine and biology.

[22]  H Alasti,et al.  Portal imaging for evaluation of daily on-line setup errors and off-line organ motion during conformal irradiation of carcinoma of the prostate. , 2001, International journal of radiation oncology, biology, physics.

[23]  M. Mills,et al.  MO‐E‐230A‐03: Journal of Applied Clinical Medical Physics , 2006 .

[24]  Heng Li,et al.  Dosimetric effects caused by couch tops and immobilization devices: report of AAPM Task Group 176. , 2014, Medical physics.

[25]  Frederik Wenz,et al.  Phantom and in-vivo measurements of dose exposure by image-guided radiotherapy (IGRT): MV portal images vs. kV portal images vs. cone-beam CT. , 2007, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[26]  Ian Poon,et al.  A comparison of two immobilization systems for stereotactic body radiation therapy of lung tumors. , 2009, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[27]  H. Vorwerk,et al.  Treatment Couch Modeling in the Treatment Planning System Eclipse , 2010 .

[28]  E. Yorke,et al.  Deep inspiration breath hold and respiratory gating strategies for reducing organ motion in radiation treatment. , 2004, Seminars in radiation oncology.

[29]  P. Keall 4-dimensional computed tomography imaging and treatment planning. , 2004, Seminars in radiation oncology.

[30]  David W Smith,et al.  Dosimetric characterization of the iBEAM evo carbon fiber couch for radiotherapy. , 2010, Medical physics.

[31]  Tommy Knöös,et al.  Current status and future perspective of flattening filter free photon beams. , 2011, Medical physics.

[32]  Armand Djouguela,et al.  The Effect of a Carbon-Fiber Couch on the Depth-Dose Curves and Transmission Properties for Megavoltage Photon Beams , 2007, Strahlentherapie und Onkologie.