An impedance pneumography signal quality index: Design, assessment and application to respiratory rate monitoring

Abstract Impedance pneumography (ImP) is widely used for respiratory rate (RR) monitoring. However, ImP-derived RRs can be imprecise. The aim of this study was to develop a signal quality index (SQI) for the ImP signal, and couple it with a RR algorithm, to improve RR monitoring. An SQI was designed which identifies candidate breaths and assesses signal quality using: the variation in detected breath durations, how well peaks and troughs are defined, and the similarity of breath morphologies. The SQI categorises 32 s signal segments as either high or low quality. Its performance was evaluated using two critical care datasets. RRs were estimated from high-quality segments using a RR algorithm, and compared with reference RRs derived from manual annotations. The SQI had a sensitivity of 77.7 %, and specificity of 82.3 %. RRs estimated from segments classified as high quality were accurate and precise, with mean absolute errors of 0.21 and 0.40 breaths per minute (bpm) on the two datasets. Clinical monitor RRs were significantly less precise. The SQI classified 34.9 % of real-world data as high quality. In conclusion, the proposed SQI accurately identifies high-quality segments, and RRs estimated from those segments are precise enough for clinical decision making. This SQI may improve RR monitoring in critical care. Further work should assess it with wearable sensor data.

[1]  Jaques Reifman,et al.  Algorithms to qualify respiratory data collected during the transport of trauma patients , 2006, Physiological measurement.

[2]  B. Debaene,et al.  Accuracy of respiratory rate monitoring by capnometry using the Capnomask(R) in extubated patients receiving supplemental oxygen after surgery. , 2012, British journal of anaesthesia.

[3]  Antti Vehkaoja,et al.  Comparison of simple algorithms for estimating respiration rate from electrical impedance pneumography signals in wearable devices , 2017 .

[4]  Q. Mcnemar Note on the sampling error of the difference between correlated proportions or percentages , 1947, Psychometrika.

[5]  Gorjan Alagic,et al.  #p , 2019, Quantum information & computation.

[6]  Willis J. Tompkins,et al.  Electrode Motion Artifacts in Electrical Impedance Pneumography , 1985, IEEE Transactions on Biomedical Engineering.

[7]  Gerard Cybulski,et al.  Assessment of calibration methods on impedance pneumography accuracy , 2016, Biomedizinische Technik. Biomedical engineering.

[8]  G. G. Stokes "J." , 1890, The New Yale Book of Quotations.

[9]  David A. Clifton,et al.  Breathing Rate Estimation From the Electrocardiogram and Photoplethysmogram: A Review , 2017, IEEE Reviews in Biomedical Engineering.

[10]  Sa , 2021, Encyclopedic Dictionary of Archaeology.

[11]  Peter Szolovits,et al.  MIMIC-III, a freely accessible critical care database , 2016, Scientific Data.

[12]  Lionel Tarassenko,et al.  Experiences implementing a system for widespread recording of patient physiology , 2015 .

[13]  Gerard Cybulski,et al.  Decomposition of the Cardiac and Respiratory Components from Impedance Pneumography Signals , 2017, BIOSIGNALS.

[14]  Paris B Lovett,et al.  The vexatious vital: neither clinical measurements by nurses nor an electronic monitor provides accurate measurements of respiratory rate in triage. , 2005, Annals of emergency medicine.

[15]  N H Lovell,et al.  Signal quality measures for pulse oximetry through waveform morphology analysis , 2011, Physiological measurement.

[16]  Mit Critical Data Secondary Analysis of Electronic Health Records , 2016 .

[17]  R. Speck,et al.  Comparison of Acoustic Respiration Rate, Impedance Pneumography and Capnometry Monitors for Respiration Rate Accuracy and Apnea Detection during GI Endoscopy Anesthesia , 2013 .

[18]  Kayvan Najarian,et al.  Motion Artifact Suppression in Impedance Pneumography Signal for Portable Monitoring of Respiration: An Adaptive Approach , 2017, IEEE Journal of Biomedical and Health Informatics.

[19]  R.G. Mark,et al.  A signal abnormality index for arterial blood pressure waveforms , 2006, 2006 Computers in Cardiology.

[20]  L. Tarassenko,et al.  An assessment of algorithms to estimate respiratory rate from the electrocardiogram and photoplethysmogram , 2016, Physiological measurement.

[21]  黄亚明 PhysioBank , 2009 .

[22]  Gerard Cybulski,et al.  Motion artifact detection in respiratory signals based on Teager energy operator and accelerometer signals , 2017 .

[23]  Axel Schäfer,et al.  Estimation of Breathing Rate from Respiratory Sinus Arrhythmia: Comparison of Various Methods , 2008, Annals of Biomedical Engineering.

[24]  J Viik,et al.  A method for suppressing cardiogenic oscillations in impedance pneumography , 2011, Physiological measurement.

[25]  D. Altman,et al.  Agreement Between Methods of Measurement with Multiple Observations Per Individual , 2007, Journal of biopharmaceutical statistics.

[26]  Jeffrey M. Hausdorff,et al.  Physionet: Components of a New Research Resource for Complex Physiologic Signals". Circu-lation Vol , 2000 .

[27]  David A. Clifton,et al.  Signal-Quality Indices for the Electrocardiogram and Photoplethysmogram: Derivation and Applications to Wireless Monitoring , 2015, IEEE Journal of Biomedical and Health Informatics.

[28]  A. Granholm,et al.  Respiratory rates measured by a standardised clinical approach, ward staff, and a wireless device , 2016, Acta anaesthesiologica Scandinavica.

[29]  Peter Charlton,et al.  Continuous respiratory rate monitoring to detect clinical deteriorations using wearable sensors , 2017 .

[30]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[31]  Kayvan Najarian,et al.  Epsilon-Tube Filtering: Reduction of High-Amplitude Motion Artifacts From Impedance Plethysmography Signal , 2015, IEEE Journal of Biomedical and Health Informatics.

[32]  David A. Clifton,et al.  Probabilistic Estimation of Respiratory Rate from Wearable Sensors , 2015 .

[33]  J. Freundlich,et al.  Electrical impedance pneumography for simple nonrestrictive continuous monitoring of respiratory rate, rhythm and tidal volume for surgical patients. , 1974, Chest.

[34]  Jaques Reifman,et al.  CAN WE IMPROVE THE CLINICAL UTILITY OF RESPIRATORY RATE AS A MONITORED VITAL SIGN? , 2009, Shock.

[35]  L. Tarassenko,et al.  Extraction of respiratory signals from the electrocardiogram and photoplethysmogram: technical and physiological determinants , 2017, Physiological measurement.

[36]  Lionel Tarassenko,et al.  Data fusion for estimating respiratory rate from a single-lead ECG , 2013, Biomed. Signal Process. Control..

[37]  K. Hillman,et al.  The objective medical emergency team activation criteria: a case-control study. , 2007, Resuscitation.

[38]  B. Preckel,et al.  Insights into postoperative respiration by using continuous wireless monitoring of respiratory rate on the postoperative ward: a cohort study , 2019, Journal of Clinical Monitoring and Computing.

[39]  S. Braun Respiratory Rate and Pattern , 1990 .