Credibility Evidence for Computational Patient Models Used in the Development of Physiological Closed-Loop Controlled Devices for Critical Care Medicine

Physiological closed-loop controlled medical devices automatically adjust therapy delivered to a patient to adjust a measured physiological variable. In critical care scenarios, these types of devices could automate, for example, fluid resuscitation, drug delivery, mechanical ventilation, and/or anesthesia and sedation. Evidence from simulations using computational models of physiological systems can play a crucial role in the development of physiological closed-loop controlled devices; but the utility of this evidence will depend on the credibility of the computational model used. Computational models of physiological systems can be complex with numerous non-linearities, time-varying properties, and unknown parameters, which leads to challenges in model assessment. Given the wide range of potential uses of computational patient models in the design and evaluation of physiological closed-loop controlled systems, and the varying risks associated with the diverse uses, the specific model as well as the necessary evidence to make a model credible for a use case may vary. In this review, we examine the various uses of computational patient models in the design and evaluation of critical care physiological closed-loop controlled systems (e.g., hemodynamic stability, mechanical ventilation, anesthetic delivery) as well as the types of evidence (e.g., verification, validation, and uncertainty quantification activities) presented to support the model for that use. We then examine and discuss how a credibility assessment framework (American Society of Mechanical Engineers Verification and Validation Subcommittee, V&V 40 Verification and Validation in Computational Modeling of Medical Devices) for medical devices can be applied to computational patient models used to test physiological closed-loop controlled systems.

[1]  Rodney A. Kennedy,et al.  Robust control design for automatic regulation of blood pressure , 2013 .

[2]  Mauro Ursino,et al.  Interaction between carotid baroregulation and the pulsating heart: a mathematical model. , 1998, American journal of physiology. Heart and circulatory physiology.

[3]  Karolos M. Grigoriadis,et al.  Robust linear parameter-varying control of blood pressure using vasoactive drugs , 2015, Int. J. Control.

[4]  Derek A. Linkens,et al.  Fuzzy logic for auditory evoked response monitoring and control of depth of anaesthesia , 1998, Fuzzy Sets Syst..

[5]  Mike Rivington,et al.  Validation of biophysical models: issues and methodologies. A review , 2011, Agronomy for Sustainable Development.

[6]  A. Visioli,et al.  On the Identification of the Propofol PK/PD Model Using BIS Measurements , 2017 .

[7]  ShiNung Ching,et al.  Robust control of burst suppression for medical coma , 2015, Journal of neural engineering.

[8]  R. Millard,et al.  On using a self-tuning controller for blood pressure regulation during surgery in man. , 1987, Computers in biology and medicine.

[9]  J.W. Huang,et al.  Multiple-drug hemodynamic control using fuzzy decision theory , 1998, IEEE Transactions on Biomedical Engineering.

[10]  Youqing Wang,et al.  An enriched simulation environment for evaluation of closed-loop anesthesia , 2014, Journal of Clinical Monitoring and Computing.

[11]  E.A. Woodruff,et al.  A model for the design and evaluation of algorithms for closed-loop cardiovascular therapy , 1997, IEEE Transactions on Biomedical Engineering.

[12]  Guido Avanzolini,et al.  A comprehensive simulator of the human respiratory system: Validation with experimental and simulated data , 2007, Annals of Biomedical Engineering.

[13]  Giuseppe Baselli,et al.  Modelling and disentangling physiological mechanisms: linear and nonlinear identification techniques for analysis of cardiovascular regulation , 2009, Philosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences.

[14]  J Schüttler,et al.  Population Pharmacokinetics of Propofol: A Multicenter Study , 2000, Anesthesiology.

[15]  J. Urzua,et al.  A model of internal control may improve the response time of an automatic arterial pressure controller , 2000, Proceedings of the 22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society (Cat. No.00CH37143).

[16]  Carolyn L. Beck,et al.  Design of L1 adaptive controllers for human patient anesthesia , 2015 .

[17]  Eann A Patterson,et al.  A framework to establish credibility of computational models in biology. , 2017, Progress in biophysics and molecular biology.

[18]  Patrick J. Roache,et al.  Perspective: Validation—What Does It Mean? , 2009 .

[19]  Emery N Brown,et al.  A closed-loop anesthetic delivery system for real-time control of burst suppression , 2013, Journal of neural engineering.

[20]  Sandy Weininger,et al.  Regulatory Considerations for Physiological Closed-Loop Controlled Medical Devices Used for Automated Critical Care: Food and Drug Administration Workshop Discussion Topics , 2017, Anesthesia and analgesia.

[21]  Karolos M. Grigoriadis,et al.  Automatic Control of Arterial Pressure for Hypotensive Patients using Phenylephrine , 2014 .

[22]  Maxime Cannesson,et al.  Closed-Loop Fluid Resuscitation: Robustness Against Weight and Cardiac Contractility Variations , 2013, Anesthesia and analgesia.

[23]  N. T. Smith,et al.  Improved safety and efficacy in adaptive control of arterial blood pressure through the use of a supervisor , 1992, IEEE Transactions on Biomedical Engineering.

[24]  Mahdi Mahfouf,et al.  Generalized predictive control with feedforward (GPCF) for multivariable anaesthesia , 1992 .

[25]  Chang-Sei Kim,et al.  A Comparative Data-Based Modeling Study on Respiratory CO2 Gas Exchange during Mechanical Ventilation , 2016, Front. Bioeng. Biotechnol..

[26]  Guy A. Dumont,et al.  PATIENT VARIABILITY AND UNCERTAINTY QUANTIFICATION IN ANESTHESIA: PART II – PKPD UNCERTAINTY , 2006 .

[27]  L Finkelstein,et al.  Validation of simple and complex models in physiology and medicine. , 1984, The American journal of physiology.

[28]  Luis Rodrigues,et al.  Controller design for neuromuscular blockade level tracking based on optimal control , 2017 .

[29]  Mahdi Mahfouf,et al.  Unconstrained and constrained generalised predictive control of depth of anaesthesia during surgery , 2003 .

[30]  Gerard L. Coté,et al.  Nonlinear closed-loop control system for intracranial pressure regulation , 2006, Annals of Biomedical Engineering.

[31]  Manfred Morari,et al.  A new paradigm for the closed-loop intraoperative administration of analgesics in humans , 2002, IEEE Transactions on Biomedical Engineering.

[32]  M. Rivington,et al.  Validation of Biophysical Models: Issues and Methodologies , 2011 .

[33]  Carolyn L. Beck,et al.  On the use of multivariable piecewise-linear models for predicting human response to anesthesia , 2004, IEEE Transactions on Biomedical Engineering.

[34]  Carolyn L. Beck Modeling and control of pharmacodynamics , 2015, Eur. J. Control.

[35]  T. G. Coleman,et al.  Validation of a computational platform for the analysis of the physiologic mechanisms of a human experimental model of hemorrhage. , 2009, Resuscitation.

[36]  Michel M R F Struys,et al.  Performance Evaluation of Two Published Closed-loop Control Systems Using Bispectral Index Monitoring: A Simulation Study , 2004, Anesthesiology.

[37]  Saltelli Andrea,et al.  Global Sensitivity Analysis: The Primer , 2008 .

[38]  Hiromitsu Ohmori,et al.  Adaptive regulation of arterial gas pressures by using robust adaptive control scheme , 1988, Proceedings of the 27th IEEE Conference on Decision and Control.

[39]  Guy A. Dumont,et al.  PATIENT VARIABILITY AND UNCERTAINTY QUANTIFICATION IN ANESTHESIA: PART I – PKPD MODELING AND IDENTIFICATION , 2006 .

[40]  Torbjörn Wigren,et al.  Nonlinear Identification of a Minimal Neuromuscular Blockade Model in Anesthesia , 2012, IEEE Transactions on Control Systems Technology.

[41]  F. Tehrani,et al.  Mathematical analysis and computer simulation of the respiratory system in the newborn infant , 1993, IEEE Transactions on Biomedical Engineering.

[42]  Mark C Anderson What is predictive capability , 2010 .

[43]  Benoit Tavernier,et al.  Evaluation of a novel closed-loop fluid-administration system based on dynamic predictors of fluid responsiveness: an in silico simulation study , 2011, Critical care.

[44]  Teresa Mendonça,et al.  Controlling the depth of anesthesia by a novel positive control strategy , 2014, Comput. Methods Programs Biomed..

[45]  Berno J. E. Misgeld,et al.  Optimal learning control of oxygen saturation using a policy iteration algorithm and a proof-of-concept in an interconnecting three-tank system , 2017 .

[46]  C. Minto,et al.  Pharmacokinetic models of propofol for TCI , 2008, Anaesthesia.

[47]  R.J. Roy,et al.  Multiple drug hemodynamic control by means of a supervisory-fuzzy rule-based adaptive control system: validation on a model , 1995, IEEE Transactions on Biomedical Engineering.

[48]  Manfred Morari,et al.  Improving regulation of mean arterial blood pressure during anesthesia through estimates of surgery effects , 2000, IEEE Transactions on Biomedical Engineering.

[49]  Luis A. Paz,et al.  Performance of an Adaptive Controller for the Neuromuscular Blockade Based on Inversion of a Wiener Model , 2015 .

[50]  Alexander Medvedev,et al.  Nonlinear dynamics in closed-loop anesthesia: Pharmacokinetic/pharmacodynamic model under PID-feedback , 2014, 2014 American Control Conference.

[51]  C. Cobelli,et al.  In Silico Preclinical Trials: A Proof of Concept in Closed-Loop Control of Type 1 Diabetes , 2009, Journal of diabetes science and technology.

[52]  R. Northrop,et al.  A new approach to the modeling and control of postoperative pain , 1990, IEEE Transactions on Biomedical Engineering.

[53]  S. J. Silva,et al.  Mean arterial pressure-embedded model for real-time simulation and controller validation , 2017 .

[54]  Vincent C. Rideout,et al.  Mathematical and Computer Modeling of Physiological Systems , 1991 .

[55]  S L Shafer,et al.  Response Surface Model for Anesthetic Drug Interactions , 2000, Anesthesiology.

[56]  Chi-Ngon Nguyen,et al.  The benefits of using Guyton's model in a hypotensive control system , 2008, Comput. Methods Programs Biomed..

[57]  Timothy G. Trucano,et al.  Predictive Capability Maturity Model for computational modeling and simulation. , 2007 .

[58]  S L Shafer,et al.  The influence of age on propofol pharmacodynamics. , 1999, Anesthesiology.

[59]  Pras Pathmanathan,et al.  Applying a Computational Model Credibility Framework to Physiological Closed-Loop Controlled Medical Device Testing , 2018, 2018 IEEE Life Sciences Conference (LSC).

[60]  Jin-Oh Hahn,et al.  Development and In Silico Evaluation of a Model-Based Closed-Loop Fluid Resuscitation Control Algorithm , 2019, IEEE Transactions on Biomedical Engineering.

[61]  L. Praly,et al.  Indirect adaptive nonlinear control , 1988, Proceedings of the 27th IEEE Conference on Decision and Control.

[62]  Farida Cheriet,et al.  Simulations for Mechanical Ventilation in Children: Review and Future Prospects , 2013, Critical care research and practice.

[63]  T. G. Coleman,et al.  Systems biology and integrative physiological modelling , 2011, The Journal of physiology.

[64]  Ralph C. Smith,et al.  Uncertainty Quantification: Theory, Implementation, and Applications , 2013 .

[65]  Laurent Papazian,et al.  High level of burnout in intensivists: prevalence and associated factors. , 2007, American journal of respiratory and critical care medicine.

[66]  Carolyn L. Beck,et al.  Modeling and Control of Anesthetic Pharmacodynamics , 2007 .

[67]  Carolyn L. Beck,et al.  Control of Patient Response to Anesthesia using ℒ 1 Adaptive Methods , 2012 .

[68]  Clara M. Ionescu,et al.  Lessons learned from closed loops in engineering: towards a multivariable approach regulating depth of anaesthesia , 2014, Journal of Clinical Monitoring and Computing.

[69]  T Bouillon,et al.  Model-based control of mechanical ventilation: design and clinical validation. , 2004, British journal of anaesthesia.

[70]  Bahram Parvinian,et al.  Control-oriented physiological modeling of hemodynamic responses to blood volume perturbation. , 2018, Control engineering practice.

[71]  Mehrdad Saif,et al.  Oxygen Therapy Control of Neonates: Part II - Evaluating Manual, PID and Fuzzy Logic controller Designs , 2008, Control. Intell. Syst..

[72]  Mario G. Iobbi,et al.  Oximetry feedback flow control simulation for oxygen therapy , 2007, Journal of Clinical Monitoring and Computing.

[73]  CM Friedrich,et al.  A model qualification method for mechanistic physiological QSP models to support model‐informed drug development , 2016, CPT: pharmacometrics & systems pharmacology.

[74]  M Cannesson,et al.  First closed-loop goal directed fluid therapy during surgery: a pilot study. , 2014, Annales francaises d'anesthesie et de reanimation.

[75]  R. Millard,et al.  Self-tuning control of hypotension during ENT surgery using a volatile anaesthetic , 1988 .

[76]  M.M. Silva,et al.  Local identifiability and sensitivity analysis of neuromuscular blockade and depth of hypnosis models , 2014, Comput. Methods Programs Biomed..

[77]  Koji Kashihara,et al.  Adaptive Predictive Control of Arterial Blood Pressure Based on a Neural Network During Acute Hypotension , 2004, Annals of Biomedical Engineering.

[78]  Teresa Mendonça,et al.  Nonlinear controller for bispectral index tracking: Robustness and on-line retuning , 2017 .

[79]  Nandu Goswami,et al.  Bridging different perspectives of the physiological and mathematical disciplines. , 2012, Advances in physiology education.

[80]  Meng Joo Er,et al.  An intelligent adaptive control scheme for postsurgical blood pressure regulation , 2005, IEEE Transactions on Neural Networks.

[81]  Marios M. Polycarpou,et al.  Indirect adaptive nonlinear control of drug delivery systems , 1998, IEEE Trans. Autom. Control..

[82]  A. M. Zbinden,et al.  Presented at 2nd MATHMOD Vienna Feb 5-7 1997 MODELING FOR CONTROL OF MEAN ARTERIAL BLOOD PRESSURE (MAP) DURING ANESTHESIA , 1997 .

[83]  Mauro Ursino,et al.  Short-term autonomic control of cardiovascular function: a mini-review with the help of mathematical models. , 2003, Journal of integrative neuroscience.

[84]  Pras Pathmanathan,et al.  Validation and Trustworthiness of Multiscale Models of Cardiac Electrophysiology , 2018, Front. Physiol..

[85]  M. Mealer,et al.  Burnout Syndrome in the Intensive Care Unit. Future Directions for Research. , 2016, Annals of the American Thoracic Society.

[86]  Pras Pathmanathan,et al.  Applicability Analysis of Validation Evidence for Biomedical Computational Models , 2017 .

[87]  José Antonio Reboso,et al.  Adaptive fuzzy predictive controller for anesthesia delivery , 2016 .

[88]  Qinglin Sun,et al.  Modeling and Simulation of an Electronic Oxygen Regulator Based on All-Coefficient Adaptive Control , 2016 .

[89]  Manfred Morari,et al.  Modeling and closed-loop control of hypnosis by means of bispectral index (BIS) with isoflurane , 2001, IEEE Transactions on Biomedical Engineering.

[90]  Berno J. E. Misgeld,et al.  ROBUST CONTROL OF END-TIDAL CO 2 USING THE H ∞ LOOP-SHAPING APPROACH , 2013 .

[91]  F T Tehrani,et al.  A mathematical model of the human respiratory system. , 1983, Journal of biomedical engineering.

[92]  R J Roy,et al.  A circulatory model for combined nitroprusside-dopamine therapy in acute heart failure. , 1990, Medical progress through technology.

[93]  Jose Salinas,et al.  Closed-loop control of fluid therapy for treatment of hypovolemia. , 2008, The Journal of trauma.

[94]  Fleur T. Tehrani,et al.  A closed-loop system for control of the fraction of inspired oxygen and the positive end-expiratory pressure in mechanical ventilation , 2012, Comput. Biol. Medicine.

[95]  G. Dumont,et al.  System Identification and Closed-Loop Control of End-Tidal CO$_{\bf 2}$ in Mechanically Ventilated Patients , 2012, IEEE Transactions on Information Technology in Biomedicine.

[96]  Guy A. Dumont,et al.  Coordinated semi‐adaptive closed‐loop control for infusion of two interacting medications , 2018 .

[97]  David Bracco,et al.  Burn fluid resuscitation: let the autopilot do it! , 2011, Critical care medicine.

[98]  B. Wayne Bequette,et al.  Simultaneous Regulation of Hemodynamic and Anesthetic States: A Simulation Study , 2004, Annals of Biomedical Engineering.

[99]  Richard N Upton,et al.  A Physiologically Based, Recirculatory Model of the Kinetics and Dynamics of Propofol in Man , 2005, Anesthesiology.

[100]  J. S. Packer,et al.  A closed-loop system for controlling blood oxygen and carbon dioxide levels in mechanically ventilated patients , 1995 .

[101]  Jin-Oh Hahn,et al.  Closed-Loop Fluid Resuscitation Control Via Blood Volume Estimation , 2016 .

[102]  Chin-Te Chen,et al.  Adaptive control of arterial blood pressure with a learning controller based on multilayer neural networks , 1997, IEEE Transactions on Biomedical Engineering.

[103]  Guy A. Dumont,et al.  Robust control of depth of anesthesia , 2008 .

[104]  Gade Pandu Rangaiah,et al.  A Comprehensive Evaluation of PID, Cascade, Model-Predictive, and RTDA Controllers for Regulation of Hypnosis† , 2009 .

[105]  Jeffrey Shuren,et al.  An FDA Viewpoint on Unique Considerations for Medical-Device Clinical Trials. , 2017, The New England journal of medicine.

[106]  Miguel Teixeira,et al.  Automatic control of the NMB level in general anaesthesia with a switching total system mass control strategy , 2013, Journal of Clinical Monitoring and Computing.

[107]  Hector Puebla,et al.  A Cascade Feedback Control Approach for Hypnosis , 2005, Annals of Biomedical Engineering.

[108]  Isabelle Queinnec,et al.  Biology and Control Theory: Current Challenges , 2007 .

[109]  Richard N Upton,et al.  Pharmacokinetic-pharmacodynamic modelling of the cardiovascular effects of drugs – method development and application to magnesium in sheep , 2005, BMC pharmacology.

[110]  R. W. Evans,et al.  Automatic control of blood oxygen saturation in premature infants , 1993, Proceedings of IEEE International Conference on Control and Applications.

[111]  W. G. He,et al.  Improvement in Arteral Oxygen Control Using Multiple-Model Adaptive Control Procedures , 1987, IEEE Transactions on Biomedical Engineering.

[112]  J. Severinghaus Simple, accurate equations for human blood O2 dissociation computations. , 1979, Journal of applied physiology: respiratory, environmental and exercise physiology.