An Ex Vivo Evaluation of Tomographic 3-D Ultrasound, B-Mode Ultrasound, CT And MR Imaging to Measure Artery Diameter, Length and Wall Volume.

Precise measurement of luminal diameter in arteries is important when planning interventional vascular procedures in patients. Measuring wall volume may be important in detecting early artery disease and in the assessment of treatments to prevent atherosclerosis. An ex vivo phantom using porcine arteries was used to evaluate the accuracy with which (i) B-mode ultrasound, (ii) 3-D tomographic ultrasound (tUS), (iii) computed tomography (CT) and (iv) magnetic resonance imaging (MRI) measured length, diameters and volume. The mean error in inner-to-inner diameter measurements by B mode, tUS, CT and MRI were 0.08 ± 0.26, -0.73 ± 0.96 mm, 0.09 ± 0.55 and 0.60 ± 1.01 mm, respectively. The mean error in outer-to-outer diameter measurements by B mode, tUS, CT and MRI were -1.33 ± 0.61, -1.03 ± 0.35, 0.02 ± 1.00 and -0.47 ± 1.32 mm, respectively. The mean error in volume measurements by B mode, tUS, CT and MRI were -0.54 ± 0.62, -0.06 ± 0.09, 0.01 ± 0.18 and -0.20 ± 0.32 cm3, respectively. Errors in length and diameters remain within clinically acceptable thresholds where MRI was the least accurate. tUS was the most accurate method of volume measurement.

[1]  J. Powell,et al.  Limitations of ultrasonography in surveillance of small abdominal aortic aneurysms , 1991, The British journal of surgery.

[2]  D. King,et al.  Evaluation of in vitro measurement accuracy of a three‐dimensional ultrasound scanner. , 1991, Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine.

[3]  O. Gilja,et al.  In vivo comparison of 3D ultrasonography and magnetic resonance imaging in volume estimation of human kidneys. , 1995, Ultrasound in medicine & biology.

[4]  James A. Evans,et al.  Reproducibility of ultrasound measurement of the abdominal aorta , 2011, The British journal of surgery.

[5]  E. Katsogridakis,et al.  Carotid plaque volume in patients undergoing carotid endarterectomy , 2018, The British journal of surgery.

[6]  Nassir Navab,et al.  Reliability of a freehand three-dimensional ultrasonic device allowing anatomical orientation "at a glance": Study protocol for 3D measurements with Curefab CS® , 2012 .

[7]  V. Fuster,et al.  Subclinical Atherosclerosis Burden by 3D Ultrasound in Mid-Life: The PESA Study. , 2017, Journal of the American College of Cardiology.

[8]  Damian M H Sin,et al.  Evaluation of the accuracy and reliability of two 3‐dimensional sonography methods in volume measurement of small structures: An in vitro phantom study , 2009, Journal of clinical ultrasound : JCU.

[9]  S. Allender,et al.  European cardiovascular disease statistics , 2008 .

[10]  S. Haulon,et al.  Management of abdominal aortic aneurysms clinical practice guidelines of the European society for vascular surgery. , 2011, European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery.

[11]  Lu Gao,et al.  Predictive value for cardiovascular events of common carotid intima media thickness and its rate of change in individuals at high cardiovascular risk – Results from the PROG-IMT collaboration , 2018, PloS one.

[12]  Ankur Kapoor,et al.  Navigation systems for ablation. , 2010, Journal of vascular and interventional radiology : JVIR.

[13]  Aaron Fenster,et al.  3D Ultrasound Measurement of Change in Carotid Plaque Volume: A Tool for Rapid Evaluation of New Therapies , 2005, Stroke.

[14]  T. Nelson,et al.  Three‐dimensional ultrasound: accuracy of distance and volume measurements , 1996, Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology.

[15]  R. Zotz,et al.  In Vitro Measurement Accuracy of Three‐Dimensional Ultrasound , 2001, Echocardiography.

[16]  S. Hughes Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects , 2005 .

[17]  V. Fuster,et al.  Prevalence, Vascular Distribution, and Multiterritorial Extent of Subclinical Atherosclerosis in a Middle-Aged Cohort: The PESA (Progression of Early Subclinical Atherosclerosis) Study , 2015, Circulation.

[18]  E. Arnesen,et al.  The difference between ultrasound and computed tomography (CT) measurements of aortic diameter increases with aortic diameter: analysis of axial images of abdominal aortic and common iliac artery diameter in normal and aneurysmal aortas. The Tromsø Study, 1994-1995. , 2004, European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery.