Comparison of two measuring instruments, B-pro and SphygmoCor system as reference, to evaluate central systolic blood pressure and radial augmentation index

A comparison is made of central aortic systolic pressure (CASP) and the radial augmentation index (rAIx) estimated with the B-Pro device and SphygmoCor (as reference) in 104 healthy Caucasians without drug treatment, together with an analysis of the relationship between CASP and rAIx, and arterial stiffness. Peripheral and central blood pressure, and the rAIx were measured with B-pro and SphygmoCor, with determination of the central augmentation index (CAIx), pulse wave velocity (PWV), carotid intima-media thickness (IMT) and the ankle-brachial index (ABI). rAIx as determined with B-Pro was greater than measured with SphygmoCor (5.85; 95%CI: 1.75–9.96), in the same way as CASP, estimated from the transfer function (1.47; 95%CI: 0.47–2.47 mm Hg) and with the second peak of the radial wave (4.46; 95%CI: 2.80–6.12 mm Hg). The Pearson correlation coefficient for CASP with B-Pro and SphygmoCor was r=0.937 (P<0.01), with an intraclass correlation of 0.972 (95%CI: 0.959–0.981). In the case of rAIx, the correlation coefficient was r=0.436 (P<0.01), with an intraclass correlation of 0.599 (95% CI: 0.409–0.728). The correlation of CASP (B-pro) with PWV was r=0.558 (P<0.01), with CAIx r=0.253 (P<0.01) and with carotid IMT r=0.442 (P<0.01). The correlation of rAIx (B-Pro) with age was r=0.369 (r<0.01), and with CAIx r=0.463 (P<0.001). Central arterial pressure estimated with B-Pro in healthy Caucasians without drug treatment offers adequate validity vs. the reference standard (SphygmoCor). However, in the estimation of rAIx, some differences with respect to the reference standard have been detected, probably related to measurement of the second peak of the radial wave.

[1]  Ira Tabas,et al.  Letter by Williams and Tabas regarding article "atherosclerosis 2005: recent discoveries and novel hypotheses". , 2006, Circulation.

[2]  Alice Stanton,et al.  Superiority of Ambulatory Over Clinic Blood Pressure Measurement in Predicting Mortality: The Dublin Outcome Study , 2005, Hypertension.

[3]  Luis García-Ortiz,et al.  Physical exercise, fitness and dietary pattern and their relationship with circadian blood pressure pattern, augmentation index and endothelial dysfunction biological markers: EVIDENT study protocol , 2010 .

[4]  I. Meredith,et al.  ‘Generalizability’ of a radial-aortic transfer function for the derivation of central aortic waveform parameters , 2007, Journal of hypertension.

[5]  K. Kario,et al.  Maximum Value of Home Blood Pressure: A Novel Indicator of Target Organ Damage in Hypertension , 2011, Hypertension.

[6]  M. Woodward,et al.  The effects of blood pressure reduction and of different blood pressure-lowering regimens on major cardiovascular events according to baseline blood pressure: meta-analysis of randomized trials , 2011, Journal of hypertension.

[7]  Katsuhiko Kohara,et al.  Radial augmentation index: a useful and easily obtainable parameter for vascular aging. , 2004, American journal of hypertension.

[8]  Alice Stanton,et al.  Differential Impact of Blood Pressure–Lowering Drugs on Central Aortic Pressure and Clinical Outcomes: Principal Results of the Conduit Artery Function Evaluation (CAFE) Study , 2006, Circulation.

[9]  Blood pressure and the global burden of disease 2000. Part II: Estimates of attributable burden , 2006, Journal of hypertension.

[10]  M. Kähönen,et al.  Letter by Nieminen et al regarding article, "Differential impact of blood pressure-lowering drugs on central aortic pressure and clinical outcomes: principal results of the Conduit Artery Function Evaluation (CAFE) study". , 2006, Circulation.

[11]  E. Lakatta,et al.  Central versus ambulatory blood pressure in the prediction of all-cause and cardiovascular mortalities , 2011, Journal of hypertension.

[12]  Yutaka Imai,et al.  Practice guidelines of the European Society of Hypertension for clinic, ambulatory and self blood pressure measurement. , 2005, Journal of hypertension.

[13]  Bryan Williams,et al.  Development and validation of a novel method to derive central aortic systolic pressure from the radial pressure waveform using an n-point moving average method. , 2011, Journal of the American College of Cardiology.

[14]  I. Meredith,et al.  Effect of non-invasive calibration of radial waveforms on error in transfer-function-derived central aortic waveform characteristics. , 2004, Clinical science.

[15]  D. Nair,et al.  The use of ambulatory tonometric radial arterial wave capture to measure ambulatory blood pressure: the validation of a novel wrist-bound device in adults , 2008, Journal of Human Hypertension.

[16]  C. H. Chen,et al.  Estimation of central aortic pressure waveform by mathematical transformation of radial tonometry pressure. Validation of generalized transfer function. , 1997, Circulation.

[17]  Jaspal Kooner,et al.  Estimation of central aortic pressure by SphygmoCor requires intra-arterial peripheral pressures. , 2003, Clinical science.

[18]  M. Band,et al.  Peripheral blood pressure measurement is as good as applanation tonometry at predicting ascending aortic blood pressure , 2003, Journal of hypertension.

[19]  Rodney A. White,et al.  ACC/AHA 2005 Practice Guidelines for the management of patients with peripheral arterial disease (lower extremity, renal, mesenteric, and abdominal aortic): a collaborative report from the American Association for Vascular Surgery/Society for Vascular Surgery, Society for Cardiovascular Angiography , 2006, Circulation.

[20]  Hidehiko Komine,et al.  Relationship between augmentation index obtained from carotid and radial artery pressure waveforms , 2007, Journal of hypertension.

[21]  H. Struijker‐Boudier,et al.  Expert consensus document on arterial stiffness: methodological issues and clinical applications. , 2006, European heart journal.

[22]  M. O'Rourke,et al.  Prospective Evaluation of a Method for Estimating Ascending Aortic Pressure From the Radial Artery Pressure Waveform , 2001, Hypertension.

[23]  Hao-Min Cheng,et al.  Estimation of central systolic blood pressure using an oscillometric blood pressure monitor , 2010, Hypertension Research.

[24]  P. Verdonck,et al.  Noninvasive Assessment of Local Pulse Pressure: Importance of Brachial-to-Radial Pressure Amplification , 2005, Hypertension.

[25]  E. Vicaut,et al.  Mannheim Carotid Intima-Media Thickness Consensus (2004–2006) , 2006, Cerebrovascular Diseases.

[26]  N. N. Available World medical association declaration of Helsinki , 2000, Chinese Journal of Integrative Medicine.

[27]  Hao-Min Cheng,et al.  Central or peripheral systolic or pulse pressure: which best relates to target organs and future mortality? , 2009, Journal of hypertension.

[28]  E. Rodríguez-Sánchez,et al.  Relationship between intima-media thickness of the common carotid artery and arterial stiffness in subjects with and without type 2 diabetes: a case-series report , 2011, Cardiovascular diabetology.

[29]  Philip Chowienczyk,et al.  Peripheral Augmentation Index Defines the Relationship Between Central and Peripheral Pulse Pressure , 2008, Hypertension.

[30]  G. Grandes,et al.  Targeting physical activity promotion in general practice: Characteristics of inactive patients and willingness to change , 2008, BMC public health.

[31]  Jaspal S. Kooner,et al.  Estimation of central aortic pressure by SphygmoCor requires intra-arterial peripheral pressures. , 2003 .

[32]  Recommendations guiding physicians in biomedical research involving human subjects. World Medical Association Declaration of Helsinki. , 1994, Le Journal medical libanais. The Lebanese medical journal.

[33]  Azra Mahmud,et al.  Ambulatory arterial stiffness index, pulse wave velocity and augmentation index – interchangeable or mutually exclusive measures? , 2007, Journal of hypertension.

[34]  B. Fetics,et al.  Estimation of Central Aortic Pressure Waveform by Mathematical Transformation of Radial Tonometry Pressure Data , 1998 .