A magnetic induction tomography system for samples with conductivities below 10 S m−1

A 16-channel magnetic induction tomography (MIT) system has been constructed for imaging samples with low conductivities (<10 S m−1) such as biological tissues or ionized water in pipelines. The system has a fixed operating frequency of 10 MHz and employs heterodyne downconversion of the received signals, to 10 kHz, to reduce phase instabilities during signal distribution and processing. The real and imaginary components of the received signal, relative to a synchronous reference, are measured using a digital lock-in amplifier. Images are reconstructed using a linearized reconstruction method based on inversion of a sensitivity matrix with Tikhonov regularization. System performance measurements and images of a pipeline phantom and a human leg in vivo are presented. The average phase precision of the MIT system is 17 millidegrees.

[1]  Nevzat G. Gencer,et al.  Electrical conductivity imaging via contactless measurements: an experimental study , 2003, IEEE Transactions on Medical Imaging.

[2]  H Griffiths,et al.  Magnetic induction tomography: phase versus vector-voltmeter measurement techniques , 2003, Physiological measurement.

[3]  Jannicke Hilland,et al.  Simple sensor system for measuring the dielectric properties of saline solutions , 1997 .

[4]  H Griffiths,et al.  A primary field compensation scheme for planar array magnetic induction tomography. , 2004, Physiological measurement.

[5]  C Gabriel,et al.  The dielectric properties of biological tissues: I. Literature survey. , 1996, Physics in medicine and biology.

[6]  David S. Holder,et al.  Electrical Impedance Tomography : Methods, History and Applications , 2004 .

[7]  H Scharfetter,et al.  A multifrequency magnetic induction tomography system using planar gradiometers: data collection and calibration. , 2006, Physiological measurement.

[8]  Per Christian Hansen,et al.  REGULARIZATION TOOLS: A Matlab package for analysis and solution of discrete ill-posed problems , 1994, Numerical Algorithms.

[9]  H Griffiths,et al.  Residual capacitive coupling and the measurement of permittivity in magnetic induction tomography , 2007, Physiological measurement.

[10]  H. Griffiths,et al.  A numerical model for magnetic induction tomographic measurements in biological tissues. , 2001, Physiological measurement.

[11]  D. Geselowitz An application of electrocardiographic lead theory to impedance plethysmography. , 1971, IEEE transactions on bio-medical engineering.

[12]  Peter R. Johnston,et al.  Selecting the corner in the L-curve approach to Tikhonov regularization , 2000, IEEE Transactions on Biomedical Engineering.

[13]  O Dössel,et al.  Design and performance of a planar-array MIT system with normal sensor alignment. , 2005, Physiological measurement.

[14]  A. J. Peyton,et al.  A state of the art review of electromagnetic tomography. , 2003 .

[15]  H. Griffiths Magnetic induction tomography , 2001 .

[16]  M. L. Meade Lock-in amplifiers : principles and applications , 1983 .

[17]  A Korjenevsky,et al.  Magnetic induction tomography: experimental realization. , 2000, Physiological measurement.

[18]  O Dössel,et al.  A comparison of sensors for minimizing the primary signal in planar-array magnetic induction tomography. , 2005, Physiological measurement.

[19]  Patricia Brunner,et al.  Reconstruction of the shape of conductivity spectra using differential multi-frequency magnetic induction tomography , 2006, Physiological measurement.

[20]  Anthony J. Peyton,et al.  Addressing the difficulties in using inductive methods to evaluating human body composition. , 2003 .