Evaluation of an a-Si EPID in direct detection configuration as a water-equivalent dosimeter for transit dosimetry.

PURPOSE A major problem associated with amorphous silicon (a-Si) electronic portal imaging devices (EPIDs) for transit dosimetry is the presence of a phosphor layer, which can introduce large deviations from water-equivalent behavior due to energy-dependent response and visible light scattering. In this study, an amorphous silicon EPID was modified to a direct detection configuration by removing the phosphor layer, and the accuracy of using it for transit dosimetry measurements was investigated for 6 and 18 MV treatment beams by comparison to ion-chamber in water measurements. METHODS Solid water and copper were both evaluated as buildup materials. Using the optimum buildup thickness in each case, effects of changes in radiation field size, source to detector distance, and patient/phantom thickness were investigated by comparison to reference measurements made by an ionization chamber on the central axis. The off-axis response of the imager was also investigated by comparison of EPID image profiles to dose profiles obtained by a scanning ionization chamber in a water tank with various thicknesses of slab phantoms, and an anthropomorphic phantom in the beam using Gamma evaluation (3%, 3 mm criteria). The imaging characteristics of the direct EPID were investigated by comparison to a commercial EPID using QC3V phantom, and by taking images of an anthropomorphic pelvic phantom containing fiducial gold markers. RESULTS Either 30 mm of solid water or 3.3 mm of copper were found to be the most suitable buildup thicknesses with solid water providing more accurate results. Using solid water buildup, the EPID response compared to the reference dosimeter within 2% for all conditions except phantom thicknesses larger than 25 cm in 6 MV beams, which was up to 6.5%. Gamma evaluation results comparing EPID profiles and reference ionization chamber profiles showed that for 6 and 18 MV beams, at least 91.8% and 90.9% of points had a Gamma <1 for all phantoms, respectively. But using copper buildup, the EPID response had more discrepancies from the ionization chamber reference measurements, including: More than 2% difference for small air gaps using 6 MV beams, up to 8% difference for phantom thicknesses larger than 25 cm in 6 MV beams, and large differences (up to 9.3%) for increasing phantom thicknesses in 18 MV beams. The percentage of points with Gamma <1 with copper buildup were at least 96.6% and 99.8% in 6 and 18 MV beams, respectively. CONCLUSIONS The direct EPID performs as an ion-chamber detector for transit dosimetry applications in all geometries studied except for small discrepancies at 6 MV for thick phantoms. This can be ameliorated by the calibration of the EPID to dose at an intermediate phantom thickness. The major current limitation of the direct EPID is poor quality of images compared with the clinical configuration, which could be overcome by a method to interchange between imaging and dosimetry setups.

[1]  F. Faccio,et al.  Study and realization of an amorphous silicon X-ray detector with effective energy sensitivity , 1992 .

[2]  M C Kirby,et al.  Measurement possibilities using an electronic portal imaging device. , 1993, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[3]  A Fenster,et al.  X-ray scatter in megavoltage transmission radiography: physical characteristics and influence on image quality. , 1994, Medical physics.

[4]  M C Schell,et al.  A technique of automating compensator design for lung inhomogeneity correction using an electron portal imaging device. , 1994, Medical physics.

[5]  D M Roback,et al.  Evaluation of electronic portal imaging device for missing tissue compensator design and verification. , 1995, Medical physics.

[6]  M C Kirby,et al.  The use of an electronic portal imaging device for exit dosimetry and quality control measurements. , 1995, International journal of radiation oncology, biology, physics.

[7]  Y. Zhu,et al.  Portal dosimetry using a liquid ion chamber matrix: dose response studies. , 1995, Medical physics.

[8]  P M Evans,et al.  The application of transit dosimetry to precision radiotherapy. , 1996, Medical physics.

[9]  N Papanikolaou,et al.  Calculation of portal dose using the convolution/superposition method. , 1996, Medical physics.

[10]  R. Boellaard,et al.  A convolution model to convert transmission dose images to exit dose distributions. , 1997, Medical physics.

[11]  J Yorkston,et al.  Initial performance evaluation of an indirect-detection, active matrix flat-panel imager (AMFPI) prototype for megavoltage imaging. , 1998, International journal of radiation oncology, biology, physics.

[12]  Evaluation of an electronic portal imaging device for transit dosimetry. , 1999, Acta oncologica.

[13]  K. Lam,et al.  Relative dosimetry using active matrix flat-panel imager (AMFPI) technology. , 1999, Medical physics.

[14]  A G Visser,et al.  Transit dosimetry with an electronic portal imaging device (EPID) for 115 prostate cancer patients. , 1999, International Journal of Radiation Oncology, Biology, Physics.

[15]  B. Heijmen,et al.  Testing of the stability of intensity modulated beams generated with dynamic multileaf collimation, applied to the MM50 racetrack microtron. , 2000, Medical physics.

[16]  C C Ling,et al.  Relative profile and dose verification of intensity-modulated radiation therapy. , 2000, International journal of radiation oncology, biology, physics.

[17]  B. McCurdy,et al.  Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging device. , 2001, Medical physics.

[18]  Richard Lee,et al.  An investigation of a new amorphous silicon electronic portal imaging device for transit dosimetry. , 2002, Medical physics.

[19]  Mike Partridge,et al.  A performance comparison of direct- and indirect-detection flat-panel imagers , 2002 .

[20]  B G Fallone,et al.  Dosimetric IMRT verification with a flat-panel EPID. , 2003, Medical physics.

[21]  Tom Depuydt,et al.  The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields. , 2004, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[22]  Peter B Greer,et al.  Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry. , 2005, Medical physics.

[23]  J. Moran,et al.  An Active Matrix Flat Panel Dosimeter (AMFPD) for in-phantom dosimetric measurements. , 2005, Medical physics.

[24]  M. Bucciolini,et al.  Pretreatment verification of IMRT absolute dose distributions using a commercial a-Si EPID. , 2006, Medical physics.

[25]  M. V. van Herk,et al.  Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification. , 2006, Medical physics.

[26]  Jean Pouliot,et al.  Calibration of an amorphous-silicon flat panel portal imager for exit-beam dosimetry. , 2006, Medical physics.

[27]  Philippe Lambin,et al.  Routine individualised patient dosimetry using electronic portal imaging devices. , 2007, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[28]  G J Budgell,et al.  Daily monitoring of linear accelerator beam parameters using an amorphous silicon EPID , 2007, Physics in medicine and biology.

[29]  P. Lambin,et al.  A global calibration model for a-Si EPIDs used for transit dosimetry. , 2007, Medical physics.

[30]  J. Moran,et al.  Performance of a direct-detection active matrix flat panel dosimeter (AMFPD) for IMRT measurements. , 2007, Medical physics.

[31]  Clive Baldock,et al.  Initial evaluation of a commercial EPID modified to a novel direct-detection configuration for radiotherapy dosimetry. , 2008, Medical physics.

[32]  P. Lambin,et al.  Transition from a simple to a more advanced dose calculation algorithm for radiotherapy of non-small cell lung cancer (NSCLC): implications for clinical implementation in an individualized dose-escalation protocol. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[33]  P. Lambin,et al.  A literature review of electronic portal imaging for radiotherapy dosimetry. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.