Development of an eye tracking medical headlamp
暂无分享,去创建一个
This study investigates a medical headlamp which is designed to adjust LED light automatically on the X axis and Y axis plane as determined a surgeon during a surgery. The needs of surgeons are determined through a CMOS camera which detects the movement of the eye. The system is composed of a mechanical actuator and an electrical control center. The LED lamp actuator (50×24×21 mm3) was designed to be as small as possible. The actuator can be easily attached onto typical goggles and incurs no burden during long-term use due to its light weight of 40g. To determine the movement of the eye, a limited embedded CPU was designed while considering memory, the process speed, simplicity and accuracy. The designed system was tested by exercises such as moving the eye in a zigzag direction and in random up, down, left, and right directions with adjustments for different lightings. Each was tested 100 times (total 1500), showing satisfactory results with 98.72% accuracy.
[1] Robert J. K. Jacob,et al. Evaluation of eye gaze interaction , 2000, CHI.
[2] Jeff B. Pelz,et al. Building a lightweight eyetracking headgear , 2004, ETRA.
[3] Andrew T Duchowski,et al. A breadth-first survey of eye-tracking applications , 2002, Behavior research methods, instruments, & computers : a journal of the Psychonomic Society, Inc.