Variables affecting electronic root canal measurement.

This study was conducted in two parts. In the first part, 20 single-rooted teeth that had been scheduled for extraction were investigated. The electronic root canal lengths were measured in vivo with a Dental Sono-Explorer type Y-III, and the actual canal lengths were measured after extraction of the teeth. The rate of agreement of the two measurements was 77.5% within a range of +/- 0.5 mm, while it was 100% at +/- 2.0 mm, which is acceptable clinically. In the second part, there were 19 simulated canals whose lengths and apical foramen sizes were known beforehand. Experiments revealed a negative correlation between the areas of the apical foramina and the difference between the electronic and the actual root canal lengths. This relationship was shown by the linear regression equation: ŷ = 0.6-1.6x. With the exception of the smallest areas of foramina, electronic root canal length measurements were less than the actual lengths.

[1]  A F Fouad,et al.  An in vitro comparison of five root canal length measuring instruments. , 1989, Journal of endodontics.

[2]  Y. Kuttler,et al.  Microscopic investigation of root apexes. , 1955, Journal of the American Dental Association.

[3]  L. Huang,et al.  An experimental study of the principle of electronic root canal measurement. , 1987, Journal of endodontics.