Tilt and decentration tolerance of intraocular lenses: measurements with an improved mechanical model eye

Cataract, a clouding of the crystalline eye lens, is the leading cause of blindness. It can effectively be treated by cataract surgery, where the clouded lens is replaced by an artificial intraocular lens (IOL). Postoperative healing processes can cause a displacement of the IOL, which further leads to the fact that the quality of vision is deteriorated. Studies have shown that the imaging quality of high sophisticated IOL designs is more sensitive to lens displacements than simpler designs. The effects of IOL displacements are not well represented and tested within the current IOL test standard ISO 11979-2. This fact leads to the necessity to develope new test standards for novel and more sophisticated IOL designs. In this paper we present an improved model eye, which extends the current standard in three main aspects: First, the eye-model is very close to the physiology of the human eye. Second, electromechanic drives allow an automatic and precise simulation of postoperative lens tilts and decentrations, and finally in addition to standard conform MTF analysis, in the proposed setup also wavefront aberrations are measured. The latter reveals specific image aberrations caused by lens displacements. The model eye allows to objectively analyze the displacement tolerance of various IOL designs. The functionality of this novel setup is tested by measuring a spherical and an aspheric IOL design. Additionally, for comparison, IOLs that were already investigated with a previous version of the presented model eye are used for analysis. Measurements results reveal improvements compared to the previous version of the model eye and a functional prototype for wavefront measurement.

[1]  Sverker Norrby,et al.  Sources of error in intraocular lens power calculation , 2008, Journal of cataract and refractive surgery.

[2]  Pablo Artal,et al.  Theoretical comparison of aberration-correcting customized and aspheric intraocular lenses. , 2007, Journal of refractive surgery.

[3]  Achim Langenbucher,et al.  Effect of decentration and tilt on the image quality of aspheric intraocular lens designs in a model eye , 2009, Journal of cataract and refractive surgery.

[4]  Kirsten Lux,et al.  Reproducibility analysis of measurements with a mechanical semiautomatic eye model for evaluation of intraocular lenses , 2014, Photonics West - Biomedical Optics.

[5]  Mathias Forjan,et al.  Semi-automatic evaluation of intraocular lenses (IOL) using a mechanical eye model , 2013, Photonics West - Biomedical Optics.

[6]  R. Klein,et al.  Incidence of age-related cataract: the Beaver Dam Eye Study. , 1998, Archives of ophthalmology.

[7]  D. Pascolini,et al.  Global estimates of visual impairment: 2010 , 2011, British Journal of Ophthalmology.

[8]  R. Moses,et al.  Adler's Physiology of the Eye: , 1970 .

[9]  O. Findl Intraocular Lens Materials and Design , 2022 .

[10]  Stephen S. Lane,et al.  Optical performance of 3 intraocular lens designs in the presence of decentration , 2005, Journal of cataract and refractive surgery.

[11]  N A Brennan,et al.  Anatomically accurate, finite model eye for optical modeling. , 1997, Journal of the Optical Society of America. A, Optics, image science, and vision.

[12]  Natascha Bayer,et al.  Improvement of Optics, Mechanics and the Usability of a Mechanical Eye Model for Vision Quality Evaluation of IOLs.∗ , 2015 .

[13]  Hans Limburg,et al.  Visual impairment and blindness in Hungary , 2018, Acta ophthalmologica.

[14]  David Allen,et al.  Cataract and surgery for cataract , 2006, BMJ : British Medical Journal.

[15]  Edwin J Sarver,et al.  Visual Acuity as a Function of Zernike Mode and Level of Root Mean Square Error , 2003, Optometry and vision science : official publication of the American Academy of Optometry.

[16]  Erik Beckert,et al.  Implementation of a capsular bag model to enable sufficient lens stabilization within a mechanical eye model , 2015, Photonics West - Biomedical Optics.