Development and performance evaluation of the first model of 4-D CT-scanner

Four-dimensional (4-D) computed tomography (CT) is a dynamic volume imaging system of moving organs with an image quality comparable to conventional CT and it is realized by continuous and high-speed cone-beam CT. We have developed a novel two-dimensional (2-D) detector for 4-D CT, which is based on the present CT technology, and mounted it on the gantry frame of the state-of-the-art CT-scanner. In the present paper, we describe the design and the performance evaluation results of the first model of the 4-D CT-scanner. The X-ray detector for the 4-D CT-scanner is a discrete pixel detector in which pixel data are measured from independent detector elements. The numbers of elements are 912 (channels) /spl times/ 256 (segments) and the element size is approximately 1 mm /spl times/ 1 mm. Data sampling rate is 900 views (frames)/s and the dynamic range of the A/D converter is 16 bits. The rotation speed of the gantry is 1.0 s/rotation. The data transfer system between rotating and stationary parts in the gantry consists of laser diode and photodiode pairs and it achieves a net transfer speed of 5 Gbps. Volume data of 512/spl times/512/spl times/256 voxels are reconstructed with an Feldkamp-Davis-Kress (FDK) algorithm by parallel use of microprocessors. The image characteristics such as noise, uniformity, and spatial resolution were evaluated with stationary phantoms in a single rotation. The exposure dose to an object was measured with an extension of the standard measurement method of CT dose index (CTDI). Several volunteers were scanned to explore clinical potentials. For the 4-D CT-scanner, the image characteristics of stationary objects were almost the same as that of conventional CT, while CTDI of the 4-D CT was slightly higher than that of conventional CT. The isotropic resolving power of less than 0.5 mm was achieved for the stationary object.

[1]  Hiroshi Aradate,et al.  Large area 2-dimensional detector for real-time 3-dimensional CT (4D-CT) , 2001 .

[2]  S Kobayashi,et al.  Development of a 3D CT-scanner using a cone beam and video-fluoroscopic system. , 1998, Radiation medicine.

[3]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[4]  K. Satoh,et al.  Image characteristics and effective dose estimation of cone beam CT using video-fluoroscopic system , 1998, 1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255).

[5]  Takanori Tsunoo,et al.  Four-dimensional computed tomography (4D CT)--concepts and preliminary development. , 2003, Radiation medicine.

[6]  Richard A. Robb,et al.  The Dynamic Spatial Reconstructor: An X-Ray Video-Fluoroscopic CT Scanner for Dynamic Volume Imaging of Moving Organs , 1982, IEEE Transactions on Medical Imaging.

[7]  Hiroshi Aradate,et al.  Large-area two-dimensional detector for real-time three-dimensional CT (4D CT) , 2001, SPIE Medical Imaging.

[8]  C Picard,et al.  In vivo evaluation of a new system for 3D computerized angiography. , 1994, Physics in medicine and biology.

[9]  Takanori Tsunoo,et al.  4-dimensional computed tomography (4D CT) - its concepts and preliminary development , 2001, 2001 IEEE Nuclear Science Symposium Conference Record (Cat. No.01CH37310).