Lumen claims of the STERRAD 100NX sterilizer: testing performance limits when processing equipment containing long, narrow lumens.

BACKGROUND According to manufacturers information, the STERRAD 100NX sterilizer-a low temperature H(2)O(2) gas plasma sterilizer-can adequately process single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a maximum length of 500 mm using standard cycle sterilizing conditions. The aim of this study was to qualify the performance of this H(2)O(2) gas plasma sterilizer under different experimental settings representing worst case conditions. METHODS Inoculated carriers were placed at the midpoint position of specified lumens and then submitted to flex scope sterilizing conditions. To simulate insufficient cleaning or crystalline residues, we added organic and inorganic challenges to our inoculated carriers. RESULTS For experiments done with unchallenged carriers, quantitative analysis reached a log(10) reduction rate of ≥5.71, whereas qualitative results showed no growth in 24 out of 30 biologic indicators tested using flex scope half cycle conditions. Any additional kind of challenge significantly impaired the sterilization outcome. CONCLUSION The findings of our current study emphasize the importance of a thorough validated cleaning of medical devices as well as timing for cleaning and decontamination before being exposed to the H(2)O(2) sterilization process and, furthermore, the need for strict adherence to manufacturer's recommendations.

[1]  J. Babb,et al.  Endoscope decontamination: where do we go from here? , 1995, The Journal of hospital infection.

[2]  G. S. Oxborrow,et al.  Sterility testing validation of sterilization processes and sporicide testing , 1991 .

[3]  W. Rutala,et al.  Comparative evaluation of the sporicidal activity of new low-temperature sterilization technologies: ethylene oxide, 2 plasma sterilization systems, and liquid peracetic acid. , 1998, American journal of infection control.

[4]  R. Li,et al.  Parameterization of inoculum effect via mathematical modeling: aminoglycosides against Staphylococcus aureus and Escherichia coli. , 1998, Journal of chemotherapy.

[5]  Nancy S. Chu,et al.  The microbial flora of the gastrointestinal tract and the cleaning of flexible endoscopes. , 2000, Gastrointestinal endoscopy clinics of North America.

[6]  S. Ison,et al.  Disinfection, sterilization and preservation (5th ed) , 2002 .

[7]  H. Chan-Myers,et al.  Natural bioburden levels detected on rigid lumened medical devices before and after cleaning. , 1997, American journal of infection control.

[8]  N. Bachhofner,et al.  Re-evaluation of current A0 value recommendations for thermal disinfection of reusable human waste containers based on new experimental data. , 2010, The Journal of hospital infection.

[9]  D. Nelson Infection control during gastrointestinal endoscopy. , 2003, The Journal of laboratory and clinical medicine.

[10]  M. Alfa,et al.  Cleaning efficacy of medical device washers in North American healthcare facilities. , 2010, The Journal of hospital infection.

[11]  J. Babb,et al.  Endoscope decontamination: automated vs. manual. , 1995, The Journal of hospital infection.

[12]  J. Okpara-Hofmann,et al.  Comparison of low-temperature hydrogen peroxide gas plasma sterilization for endoscopes using various Sterrad models. , 2005, The Journal of hospital infection.

[13]  K. Sugano,et al.  Contamination with Hepatitis B Virus DNA in Gastrointestinal Endoscope Channels: Risk of Infection on Reuse after On-Site Cleaning , 2005, Endoscopy.

[14]  C. Alvarado,et al.  Multi-society guideline for reprocessing flexible gastrointestinal endoscopes , 2004, Gastrointestinal endoscopy.

[15]  G. Ridgway,et al.  Assessment of the efficacy of a low temperature hydrogen peroxide gas plasma sterilization system. , 1995, The Journal of hospital infection.

[16]  S. Block,et al.  Disinfection, sterilization, and preservation , 1977 .

[17]  Carol Petersen,et al.  Multi-society guideline for reprocessing flexible gastrointestinal endoscopes , 2003, Infection Control & Hospital Epidemiology.