Individualized gating windows based on four-dimensional CT information for respiration-gated radiotherapy

The purpose of this work is to relate the gating window and displacement of a moving tumor target and develop a systematic method to individualize the gating window for respiration-gated radiation therapy (RT). As the relationship between patient anatomy and respiration phase is contained in 4D images, we aim to quantify this information and utilize the data to guide gated treatment planning. After 4D image acquisition, the target and organs at risk were delineated manually on the selected gating phase. The contours were propagated automatically onto every phase-specific image set using a control volume-based contour mapping technique. The mean and maximum distances between the contours in the gating phase and each of other phases were evaluated in three dimensions. The gating window was determined in such a way that the residual movement of the target within the window is smaller or equal to the patient's setup error. The proposed method was applied to plan the gated treatments of 12 lung cancer patients. As a result of this work, a method to calculate patient-specific gating windows has been developed. The general reference drawn from this study is that, with the aide of 4D images and automated 4D contour propagation, it is feasible to individualize the gating widow selection. As compared with the current practice, the proposed technique has a potential to eliminate the guesswork involved in choosing a gating window and avoid dosimetric error in planning gated RT. In conclusion, individualization of gating windows reduces the subjectivity in respiration-gated RT and improves the treatment of moving targets.

[1]  K. Lam,et al.  Improvement of CT-based treatment-planning models of abdominal targets using static exhale imaging. , 1998, International journal of radiation oncology, biology, physics.

[2]  G. Christensen,et al.  A method for the reconstruction of four-dimensional synchronized CT scans acquired during free breathing. , 2003, Medical physics.

[3]  R M Henkelman,et al.  The double-fissure sign: a motion artifact on thin-section CT scans. , 1987, Radiology.

[4]  M. V. van Herk,et al.  Respiratory correlated cone beam CT. , 2005, Medical physics.

[5]  K. Ohara,et al.  Irradiation synchronized with respiration gate. , 1989, International journal of radiation oncology, biology, physics.

[6]  D J Conces,et al.  Motion artifacts on CT simulate bronchiectasis. , 1988, AJR. American journal of roentgenology.

[7]  Lei Xing,et al.  Image registration with auto-mapped control volumes , 2006 .

[8]  Icru Prescribing, recording, and reporting photon beam therapy , 1993 .

[9]  George T. Y. Chen,et al.  Four-dimensional image-based treatment planning: Target volume segmentation and dose calculation in the presence of respiratory motion. , 2005, International journal of radiation oncology, biology, physics.

[10]  Joseph O Deasy,et al.  Deblurring of breathing motion artifacts in thoracic PET images by deconvolution methods. , 2006, Medical physics.

[11]  A. Pevsner,et al.  Quantitation of respiratory motion during 4D-PET/CT acquisition. , 2004, Medical physics.

[12]  Lei Xing,et al.  Model-based image reconstruction for four-dimensional PET. , 2006, Medical physics.

[13]  C. Ling,et al.  Respiration-correlated spiral CT: a method of measuring respiratory-induced anatomic motion for radiation treatment planning. , 2002, Medical physics.

[14]  P. Munro,et al.  Four-dimensional cone-beam computed tomography using an on-board imager. , 2006, Medical physics.

[15]  R. Mohan,et al.  Acquiring a four-dimensional computed tomography dataset using an external respiratory signal. , 2003, Physics in medicine and biology.

[16]  Lei Xing,et al.  Reducing respiratory motion artifacts in positron emission tomography through retrospective stacking. , 2006, Medical physics.

[17]  C. J. Ritchie,et al.  Predictive respiratory gating: a new method to reduce motion artifacts on CT scans. , 1994, Radiology.

[18]  Suresh Senan,et al.  Benefit of respiration-gated stereotactic radiotherapy for stage I lung cancer: an analysis of 4DCT datasets. , 2005, International journal of radiation oncology, biology, physics.

[19]  Vladimir Pekar,et al.  Automated model-based organ delineation for radiotherapy planning in prostatic region. , 2004, International journal of radiation oncology, biology, physics.

[20]  C. Ramsey,et al.  Clinical efficacy of respiratory gated conformal radiation therapy. , 1999, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[21]  Lei Xing,et al.  MO‐E‐330A‐07: Knowledge‐Based Auto‐Contouring in 4D Radiation Therapy , 2006 .

[22]  Lei Xing,et al.  Automated contour mapping using sparse volume sampling for 4D radiation therapy. , 2007, Medical physics.

[23]  Quan Chen,et al.  Automatic re-contouring in 4D radiotherapy , 2006, Physics in medicine and biology.

[24]  George Starkschall,et al.  Dosimetric benefits of respiratory gating: a preliminary study , 2004, Journal of applied clinical medical physics.

[25]  H. Kubo,et al.  Respiration gated radiotherapy treatment: a technical study. , 1996, Physics in medicine and biology.

[26]  Gikas S. Mageras,et al.  Fluoroscopic evaluation of diaphragmatic motion reduction with a respiratory gated radiotherapy system , 2001, Journal of applied clinical medical physics.

[27]  S. K. Hilal,et al.  The tuning fork artifact in computerized tomography , 1979 .

[28]  M. V. van Herk,et al.  Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy. , 2002, International journal of radiation oncology, biology, physics.

[29]  T. Pan,et al.  4D-CT imaging of a volume influenced by respiratory motion on multi-slice CT. , 2004, Medical physics.