Noise characterization of supercontinuum sources for low-coherence interferometry applications.

We examine the noise properties of supercontinuum light sources when used in low-coherence interferometry applications. The first application is a multiple-scattering low-coherence interferometry (ms2/LCI) system, where high power and long image acquisition times are required to image deep into tissue. For this system, we compare the noise characteristics of two supercontinuum sources from different suppliers. Both sources have long-term drift that limits the amount of time over which signal averaging is advantageous for reducing noise. The second application is a high-resolution optical coherence tomography system, where broadband light is needed for high axial resolution. For this system, we compare the noise performance of the two supercontinuum sources and a light source based on four superluminescent diodes (SLD) using imaging contrast as a comparative metric. We find that the NKT SuperK has superior noise performance compared with the Fianium SC-450-4, but neither meets the performance of the SLD.

[1]  A. Fercher,et al.  Performance of fourier domain vs. time domain optical coherence tomography. , 2003, Optics express.

[2]  Vladimir Shidlovski,et al.  Ultrahigh resolution optical coherence tomography imaging with a broadband superluminescent diode light source. , 2004, Optics express.

[3]  Maciej Szkulmowski,et al.  Averaging techniques for OCT imaging. , 2013, Optics express.

[4]  Adam Wax,et al.  Design and validation of an angle-resolved low-coherence interferometry fiber probe for in vivo clinical measurements of depth-resolved nuclear morphology. , 2011, Journal of biomedical optics.

[5]  R. Spaide,et al.  Enhanced depth imaging spectral-domain optical coherence tomography. , 2008, American journal of ophthalmology.

[6]  Jennifer K. Barton,et al.  OPTICAL COHERENCE TOMOGRAPHY FOR BIODIAGNOSTICS , 1997 .

[7]  R N Graf,et al.  Parallel frequency-domain optical coherence tomography scatter-mode imaging of the hamster cheek pouch using a thermal light source. , 2008, Optics letters.

[8]  B. Bouma,et al.  Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography. , 2003, Optics letters.

[9]  J. Méteau,et al.  Supercontinuum sources in optical coherence tomography: A state of the art and the application to scan-free time domain correlation techniques and depth dependant dispersion compensation , 2012 .

[10]  Timothy A. Birks,et al.  Supercontinuum generation in photonic crystal fibers and optical fiber tapers: a novel light source , 2002 .

[11]  O. Bang,et al.  Power dependence of supercontinuum noise in uniform and tapered PCFs. , 2012, Optics express.

[12]  James C. Wyant,et al.  White light interferometry , 2002, SPIE Defense + Commercial Sensing.

[13]  A. Fercher,et al.  Submicrometer axial resolution optical coherence tomography. , 2002, Optics letters.

[14]  A. Wax,et al.  Fourier domain multispectral multiple scattering low coherence interferometry. , 2013, Applied optics.

[15]  William J. Brown,et al.  Deep tissue imaging using spectroscopic analysis of multiply scattered light , 2014 .

[16]  J. Izatt,et al.  Optimal interferometer designs for optical coherence tomography. , 1999, Optics letters.

[17]  Maciej Wojtkowski,et al.  Quality improvement for high resolution in vivo images by spectral domain optical coherence tomography with supercontinuum source , 2005 .

[18]  Changhuei Yang,et al.  Sensitivity advantage of swept source and Fourier domain optical coherence tomography. , 2003, Optics express.

[19]  Ruikang K. Wang,et al.  Digital phase stabilization to improve detection sensitivity for optical coherence tomography , 2007 .

[20]  R. Windeler,et al.  Fundamental noise limitations to supercontinuum generation in microstructure fiber. , 2002, Physical review letters.