Regional-surface-based registration for image-guided neurosurgery: effects of scan modes on registration accuracy

PurposeThe conventional surface-based method only registers the facial zone with preoperative point cloud, resulting in low accuracy away from the facial area. Acquiring a point cloud of the entire head for registration can improve registration accuracy in all parts of the head. However, it takes a long time to collect a point cloud of the entire head. It may be more practical to selectively scan part of the head to ensure high registration accuracy in the surgical area of interest. In this study, we investigate the effects of different scan regions on registration errors in different target areas when using a surface-based registration method.MethodsWe first evaluated the correlation between the laser scan resolution and registration accuracy to determine an appropriate scan resolution. Then, with the appropriate resolution, we explored the effects of scan modes on registration error in computer simulation experiments, phantom experiments and two clinical cases. The scan modes were designed based on different combinations of five zones of the head surface, i.e., the sphenoid-frontal zone, parietal zone, left temporal zone, right temporal zone and occipital zone. In the phantom experiment, a handheld scanner was used to acquire a point cloud of the head. A head model containing several tumors was designed, enabling us to calculate the target registration errors deep in the brain to evaluate the effect of regional-surface-based registration.ResultThe optimal scan modes for tumors located in the sphenoid-frontal, parietal and temporal areas are mode 4 (i.e., simultaneously scanning the sphenoid-frontal zone and the temporal zone), mode 4 and mode 6 (i.e., simultaneously scanning the sphenoid-frontal zone, the temporal zone and the parietal zone), respectively. For the tumor located in the occipital area, no modes were able to achieve reliable accuracy.ConclusionThe results show that selecting an appropriate scan resolution and scan mode can achieve reliable accuracy for use in sphenoid-frontal, parietal and temporal area surgeries while effectively reducing the operation time.

[1]  Robert J. Maciunas,et al.  Registration of head volume images using implantable fiducial markers , 1997, IEEE Transactions on Medical Imaging.

[2]  Zhijian Song,et al.  A new robust markerless method for automatic image-to-patient registration in image-guided neurosurgery system , 2017, Computer assisted surgery.

[3]  V. Seifert,et al.  Laser Surface Scanning for Patient Registration in Intracranial Image-guided Surgery , 2002, Neurosurgery.

[4]  B M Dawant,et al.  Laser range scanning for image-guided neurosurgery: investigation of image-to-physical space registrations. , 2008, Medical physics.

[5]  Robert L. Galloway,et al.  Cortical surface registration for image-guided neurosurgery using laser-range scanning , 2003, IEEE Transactions on Medical Imaging.

[6]  P. Schönemann,et al.  A generalized solution of the orthogonal procrustes problem , 1966 .

[7]  R Bucholz,et al.  Automated rejection of contaminated surface measurements for improved surface registration in image guided neurosurgery. , 2000, Studies in health technology and informatics.

[8]  C. Mascott In vivo accuracy of image guidance performed using optical tracking and optimized registration. , 2006, Journal of neurosurgery.

[9]  Man Ning Wang,et al.  Properties of the target registration error for surface matching in neuronavigation , 2011, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[10]  Jean-Christophe Sol,et al.  Quantification of True In Vivo (Application) Accuracy in Cranial Image-guided Surgery: Influence of Mode of Patient Registration , 2006, Neurosurgery.

[11]  Zhijian Song,et al.  Distribution templates of the fiducial points in image-guided neurosurgery. , 2010, Neurosurgery.

[12]  Franz Kainberger,et al.  Comparison of laser surface scanning and fiducial marker-based registration in frameless stereotaxy. Technical note. , 2007, Journal of neurosurgery.

[13]  S Hassfeld,et al.  Automated laser registration in image-guided surgery: evaluation of the correlation between laser scan resolution and navigation accuracy. , 2004, International journal of oral and maxillofacial surgery.

[14]  P. Woerdeman,et al.  Application accuracy in frameless image-guided neurosurgery: a comparison study of three patient-to-image registration methods. , 2007, Journal of neurosurgery.

[15]  Song Zhijian,et al.  Distribution Templates of the Fiducial Points in Image‐Guided Neurosurgery , 2010 .

[16]  K. Paulsen,et al.  Intraoperative fiducial-less patient registration using volumetric 3D ultrasound: a prospective series of 32 neurosurgical cases. , 2015, Journal of neurosurgery.

[17]  T. Peters Image-guidance for surgical procedures , 2006, Physics in medicine and biology.

[18]  S Hassfeld,et al.  High-Resolution Laser Surface Scanning for Patient Registration in Cranial Computer-Assisted Surgery , 2004, Minimally invasive neurosurgery : MIN.

[19]  K. S. Arun,et al.  Least-Squares Fitting of Two 3-D Point Sets , 1987, IEEE Transactions on Pattern Analysis and Machine Intelligence.

[20]  D. Paraskevopoulos,et al.  Comparative study of application accuracy of two frameless neuronavigation systems: experimental error assessment quantifying registration methods and clinically influencing factors , 2011, Neurosurgical Review.

[21]  K. Paulsen,et al.  Intraoperative patient registration using volumetric true 3D ultrasound without fiducials. , 2012, Medical physics.

[22]  Zhijian Song,et al.  A new markerless patient-to-image registration method using a portable 3D scanner. , 2014, Medical physics.