Computational Fluid Dynamics Characterization of Blood Flow in Central Aorta to Pulmonary Artery Connections: Importance of Shunt Angulation as a Determinant of Shear Stress-Induced Thrombosis

The central aortic shunt, consisting of a Gore-Tex (polytetrafluoroethylene) tube (graft) connecting the ascending aorta to the pulmonary artery, is a palliative operation for neonates with cyanotic congenital heart disease. These tubes often have an extended length, and therefore must be angulated to complete the connection to the posterior pulmonary arteries. Thrombosis of the graft is not uncommon and can be life-threatening. We have shown that a viscous fluid (such as blood) traversing a curve or bend in a small-caliber vessel or conduit can give rise to marked increases in wall shear stress, which is the major mechanical factor responsible for vascular thrombosis. Thus, the objective of this study was to use computational fluid dynamics to investigate whether wall shear stress (and shear rate) generated in angulated central aorta-to-pulmonary artery connections, in vivo, can be of magnitude and distribution to initiate platelet activation/aggregation, ultimately leading to thrombus formation. Anatomical features required to construct the computer-simulated blood flow pathways were verified from angiograms of central aortic shunts in patients. For the modeled central aortic shunts, we found wall shear stresses of (80–200 N/m2), with shear rates of (16,000–40,000/s), at sites of even modest curvature, to be high enough to cause platelet-mediated shunt thrombosis. The corresponding energy losses for the fluid transitions through the aorta-to-pulmonary connections constituted (70 %) of the incoming flow’s mechanical energy. The associated velocity fields within these shunts exhibited vortices, eddies, and flow stagnation/recirculation, which are thrombogenic in nature and conducive to energy dissipation. Angulation-induced, shear stress-mediated shunt thrombosis is insensitive to aspirin therapy alone. Thus, for patients with central aortic shunts of longer length and with angulation, aspirin alone will provide insufficient protection against clotting. These patients are at risk for shunt thrombosis and significant morbidity and mortality, unless their anticoagulation regimen includes additional antiplatelet medications.

[1]  N. Maalej,et al.  Increased shear stress overcomes the antithrombotic platelet inhibitory effect of aspirin in stenosed dog coronary arteries. , 1996, Circulation.

[2]  C. Byrum,et al.  Subclavian-pulmonary artery shunts with polytetrafluorethylene interposition grafts. , 1984, The Annals of thoracic surgery.

[3]  F. Midgley,et al.  Construction of interposition polytetrafluoroethylene ascending aorta-pulmonary artery shunt. , 1980, The Annals of thoracic surgery.

[4]  R. Brickman,et al.  Use of microporous expanded polytetrafluoroethylene grafts for aorta-pulmonary shunts in infants with complex cyanotic heart disease. A report of seven cases. , 1978, The Journal of thoracic and cardiovascular surgery.

[5]  R. Bartlett,et al.  Microporous expanded polytetrafluoroethylene arterial prosthesis for construction of aortopulmonary shunts: experimental and clinical results. , 1976, The Annals of thoracic surgery.

[6]  R. Ascuitto,et al.  Pressure Loss from Flow Energy Dissipation: Relevance to Fontan-Type Modifications , 2001, Pediatric Cardiology.

[7]  Hidesato Ito,et al.  Flow in curved pipes. , 1987 .

[8]  W. Gersony,et al.  Systemic‐Pulmonary Arterial Shunts in the First Year of Life , 1974, Circulation.

[9]  K. Schrör The basic pharmacology of ticlopidine and clopidogrel. , 1993, Platelets.

[10]  W. S. Ring,et al.  Central aorta-pulmonary artery shunts in neonates with complex cyanotic congenital heart disease. , 1987, The Journal of thoracic and cardiovascular surgery.

[11]  Anna Trykozko,et al.  The effects of graft geometry on the patency of a systemic-to-pulmonary shunt: a computational fluid dynamics study. , 2005, Artificial organs.

[12]  A. Wessel,et al.  Reduced frequency of occlusion of aorto-pulmonary shunts in infants receiving Aspirin , 1999, Cardiology in the Young.

[13]  H. Okabe,et al.  [Long-term patency after the Blalock-Taussig operation--comparison between classic and modified shunts]. , 1996, [Zasshi] [Journal]. Nihon Kyobu Geka Gakkai.

[14]  Y. Ueda,et al.  Three-Dimensional Simulation of the Blalock-Taussig Shunt Using Computational Fluid Dynamics , 2001, Surgery Today.

[15]  R. Anderson,et al.  Postoperative Hemodynamics in Children with Polytetrafluoroethylene Shunts , 1981, Circulation.

[16]  W. R. Dean LXXII. The stream-line motion of fluid in a curved pipe (Second paper) , 1928 .

[17]  K S Sakariassen,et al.  Shear-induced platelet activation and platelet microparticle formation at blood flow conditions as in arteries with a severe stenosis. , 1997, Arteriosclerosis, thrombosis, and vascular biology.

[18]  S. Mousa,et al.  PLATELETS IN HEALTH AND DISEASE : PLATELET GPIIB-IIIA STRUCTURE AND FUNCTION : RECENT ADVANCES IN ANTIPLATELET THERAPY , 1996 .

[19]  Wei Yin,et al.  Effects of physiologically relevant dynamic shear stress on platelet complement activation , 2011, Platelets.

[20]  J. Berkson,et al.  THE RELATION BETWEEN VISCOSITY OF THE BLOOD AND THE RELATIVE VOLUME OF ERYTHROCYTES (HEMATOCRIT VALUE) , 1935 .

[21]  W. R. Dean,et al.  Note on the motion of fluid in a curved pipe , 1959 .

[22]  S. Yuan,et al.  The Blalock‐Taussig Shunt , 2009, Journal of cardiac surgery.

[23]  W. R. Dean XVI. Note on the motion of fluid in a curved pipe , 1927 .

[24]  J J Amato,et al.  Systemic-pulmonary polytetrafluoroethylene shunts in palliative operations for congenital heart disease. Revival of the central shunt. , 1988, The Journal of thoracic and cardiovascular surgery.

[25]  J. Hathcock Flow effects on coagulation and thrombosis. , 2006, Arteriosclerosis, thrombosis, and vascular biology.

[26]  J. S. Davidson Anastomosis Between the Ascending Aorta and the Main Pulmonary Artery in the Tetralogy of Fallot , 1955, Thorax.

[27]  J. Newburger,et al.  Clopidogrel in infants with systemic-to-pulmonary-artery shunts. , 2013, The New England journal of medicine.

[28]  Y Komiyama,et al.  High shear stress can initiate both platelet aggregation and shedding of procoagulant containing microparticles. , 1996, Blood.

[29]  A. Federici,et al.  Activation-independent platelet adhesion and aggregation under elevated shear stress. , 2005, Blood.

[30]  Sarah L. Waters,et al.  Steady flows in pipes with finite curvature , 2005 .

[31]  L. Edmunds,et al.  A comparison of Blalock-Taussig, Waterston, and polytetrafluoroethylene shunts in children less than two weeks of age. , 1984, The Annals of thoracic surgery.

[32]  J. O'brien,et al.  Shear-induced platelet aggregation , 1990, The Lancet.

[33]  R. Ecker,et al.  Intrapericardial Aortico‐Pulmonary Artery Shunt , 1963, Circulation.

[34]  S. Pappelbaum,et al.  Systemic-pulmonary shunts in infants and children. Early and late results. , 1984, The Journal of thoracic and cardiovascular surgery.

[35]  B. Cowan,et al.  Wall Shear Stress is the Primary Mechanism of Energy Loss in the Fontan Connection , 2006, Pediatric Cardiology.

[36]  Y. Ikeda,et al.  Characterization of the Unique Mechanism Mediating the Shear-dependent Binding of Soluble von Willebrand Factor to Platelets (*) , 1995, The Journal of Biological Chemistry.

[37]  F. Migliavacca,et al.  Computational model of the fluid dynamics in systemic-to-pulmonary shunts. , 2000, Journal of biomechanics.

[38]  S. Uchiyama,et al.  Effects of Dipyridamole and Aspirin on Shear-Induced Platelet Aggregation in Whole Blood and Platelet-Rich Plasma , 2002, Cerebrovascular Diseases.

[39]  T. Gardner,et al.  Systemic-Pulmonary shunts inneonates and infants using microporous expanded polytetrafluoroethylene: immediate and late results. , 1980, The Annals of thoracic surgery.

[40]  J. Moake,et al.  Direct demonstration of radiolabeled von willebrand factor binding to platelet glycoprotein Ib and IIb-IIIa in the presence of shear stress , 1995, Annals of Biomedical Engineering.

[41]  F. Moll,et al.  Heparin immobilization reduces thrombogenicity of small-caliber expanded polytetrafluoroethylene grafts. , 2006, Journal of vascular surgery.

[42]  M. A. Ali Khan,et al.  Results of 546 Blalock-Taussig shunts performed in 478 patients , 1998, Cardiology in the Young.

[43]  R. Califf,et al.  Clinical Outcomes of Palliative Surgery Including a Systemic-to-Pulmonary Artery Shunt in Infants With Cyanotic Congenital Heart Disease: Does Aspirin Make a Difference? , 2007, Circulation.

[44]  Ted B. Martonen,et al.  Simulations of Flow in Curved Tubes , 1997 .

[45]  V. Willman,et al.  The Blalock-Taussig shunt in the neonate. , 1978, The Annals of thoracic surgery.

[46]  D. Birchall,et al.  Analysis of haemodynamic factors involved in carotid atherosclerosis using computational fluid dynamics. , 2009, The British journal of radiology.

[47]  J. Moake,et al.  Platelets and shear stress. , 1996, Blood.

[48]  N. Kleiman,et al.  Platelet glycoprotein IIb/IIIa antagonists. What are the relevant issues concerning their pharmacology and clinical use? , 1999, Circulation.

[49]  R. Guyton,et al.  The Blalock-Taussig shunt. Low risk, effective palliation, and pulmonary artery growth. , 1983, The Journal of thoracic and cardiovascular surgery.