Vibration isolator carrying atomic force microscope’s head

Abstract For high-resolution imaging in harsh environments this paper proposes a vibration isolation system that actively positions the head of an atomic force microscope (AFM) to maintain the vertical distance to the sample. On the moving platform carrying the AFM head, a displacement sensor is installed to detect the vibrations between the probe and the sample that impair the imaging quality. The detected vibrations are rejected by vertically moving the platform with feedback control. For the motion, flexure-guided Lorentz actuators are designed, such that the resulting suspension mode occurs around the major spectrum of the floor vibrations. By feedback control design, the high gain of the suspension mode is used to increase the open-loop gain for better vibration rejection. The experimental results demonstrate that the vibration isolation system can reject 99.3% of the vibrations. As a result, AFM imaging of nanoscale features is successfully performed in a vibrational environment.

[1]  S. Manalis,et al.  Atomic force microscopy with inherent disturbance suppression for nanostructure imaging , 2006, Nanotechnology.

[2]  L. Jabben Mechatronic design of a magnetically suspended rotating platform , 2007 .

[3]  Georg Schitter,et al.  Atomic force microscopy using voice coil actuators for vibration isolation , 2015, 2015 IEEE International Conference on Advanced Intelligent Mechatronics (AIM).

[4]  Georg Schitter,et al.  Auto-tuning PI controller for surface tracking in atomic force microscopy - a practical approach , 2016, 2016 American Control Conference (ACC).

[5]  W. Graham Richards,et al.  Art of electronics , 1983, Nature.

[6]  A.J. Fleming Nanopositioning System With Force Feedback for High-Performance Tracking and Vibration Control , 2010, IEEE/ASME Transactions on Mechatronics.

[7]  Gil U. Lee,et al.  Scanning probe microscopy. , 2010, Current opinion in chemical biology.

[8]  Jinchun Hu,et al.  A negative stiffness vibration isolator using magnetic spring combined with rubber membrane , 2013 .

[9]  Hal Amick,et al.  Evolving criteria for research facilities: vibration , 2005, SPIE Optics + Photonics.

[10]  Kyihwan Park,et al.  Note: vibration reduction control of an atomic force microscope using an additional cantilever. , 2011, The Review of scientific instruments.

[11]  Andrew J. Fleming,et al.  Dual-Stage Vertical Feedback for High-Speed Scanning Probe Microscopy , 2011, IEEE Transactions on Control Systems Technology.

[12]  Shingo Ito,et al.  Flexure design for precision positioning using low-stiffness actuators , 2016 .

[13]  N. Zhang,et al.  Experimental study of train-induced vibrations of environments and buildings , 2005 .

[14]  Mitsuo Hirata,et al.  High-speed precision motion control , 2017 .

[15]  K. Leang,et al.  Design and Control of a Three-Axis Serial-Kinematic High-Bandwidth Nanopositioner , 2012, IEEE/ASME Transactions on Mechatronics.

[16]  Georg Schitter,et al.  Comparison and Classification of High-Precision Actuators Based on Stiffness Influencing Vibration Isolation , 2016, IEEE/ASME Transactions on Mechatronics.

[17]  Stefan Kuiper,et al.  Model-based feedback controller design for dual actuated atomic force microscopy , 2012 .

[18]  S O R Moheimani,et al.  Invited review article: high-speed flexure-guided nanopositioning: mechanical design and control issues. , 2012, The Review of scientific instruments.

[19]  Georg Schitter,et al.  Eliminating mechanical perturbations in scanning probe microscopy , 2002 .

[20]  Jacqueline A. Cutroni,et al.  Rigid design of fast scanning probe microscopes using finite element analysis. , 2004, Ultramicroscopy.

[21]  J. H. Rainer,et al.  Control of traffic-induced vibration in buildings using vehicle suspension systems , 1996 .

[22]  Georg Schitter,et al.  Extending the Range of Geophones by Negative Impedance Converter , 2016 .

[23]  P. Hansma,et al.  Atomic force microscopy , 1990, Nature.

[24]  M Maarten Steinbuch,et al.  Model-based spatial feedforward for over-actuated motion systems , 2014 .

[25]  Andrew J. Fleming,et al.  A review of nanometer resolution position sensors: Operation and performance , 2013 .

[26]  John Lygeros,et al.  Dual-Stage Nanopositioning for High-Speed Scanning Probe Microscopy , 2014, IEEE/ASME Transactions on Mechatronics.

[27]  Christophe Collette,et al.  Sensor fusion methods for high performance active vibration isolation systems , 2015 .

[28]  Kazuhiro Kogure,et al.  Application of Atomic Force Microscopy to Observation of Marine Bacteria , 2004 .

[29]  G. G. Stokes "J." , 1890, The New Yale Book of Quotations.

[30]  Andrew J Fleming,et al.  Compact ultra-fast vertical nanopositioner for improving scanning probe microscope scan speed. , 2011, The Review of scientific instruments.

[31]  No-Cheol Park,et al.  Optimal Design of Moving-Magnet Type Actuators for Optical Disk Drives Considering Effect of Coil Electromagnet , 2009, IEEE Transactions on Magnetics.

[32]  Lu Xia,et al.  Lead and lag compensators with complex poles and zeros design formulas for modeling and loop shaping , 2007, IEEE Control Systems.

[33]  Andrea Jeremie,et al.  Sub-nanometer active seismic isolator control , 2013 .

[34]  Shingo Ito,et al.  Low-stiffness dual stage actuator for long rage positioning with nanometer resolution , 2015 .