Adaptive FES Assistance Using a Novel Gait Phase Detection Approach

This paper presents an adaptive knee-joint based Functional Electrical Stimulation (FES)method to correct the foot drop of paretic patients during swing phase. The rationale behind the adaptive FES is to amplify dorsiflexor stimulation in the late swing when it is most needed in order to face the increased plantar flexor co-contraction as gastrocnemius muscles are stretched by knee re-extension. To accurately detect the swing phase (i.e., toes off (TO)and initial contact (IC)), a novel algorithm is proposed by using a foot-mounted inertial measurement unit (IMU). The proposed strategy is verified by experiments conducted with three healthy subjects and three paretic patients. The experimental results show that highly accurate detection of TO/I C can be achieved under different walking speeds and foot contact conditions (normal and abnormal gaits). The clinical experimental results with paretic patients also reveal that similar effects on ankle dorsiflexion can be observed during mid and late swing using the proposed adaptive FES with respect to the classical FES method, while the adaptive FES used lower stimulation intensity.

[1]  Qingguo Li,et al.  Concurrent validation of Xsens MVN measurement of lower limb joint angular kinematics , 2013, Physiological measurement.

[2]  Kyoungchul Kong,et al.  Fast Gait Mode Detection and Assistive Torque Control of an Exoskeletal Robotic Orthosis for Walking Assistance , 2018, IEEE Transactions on Robotics.

[3]  T. Sinkjaer,et al.  A review of portable FES-based neural orthoses for the correction of drop foot , 2002, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[4]  P. Goldie,et al.  Deficit and change in gait velocity during rehabilitation after stroke. , 1996, Archives of physical medicine and rehabilitation.

[5]  G. Lyons,et al.  Evaluation of a Drop Foot Stimulator FES Intensity Envelope Matched to Tibialis Anterior Muscle Activity during Walking , 2001 .

[6]  G. Lyons,et al.  The development of a potential optimized stimulation intensity envelope for drop foot applications , 2003, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[7]  Liberson Wt,et al.  Functional electrotherapy: stimulation of the peroneal nerve synchronized with the swing phase of the gait of hemiplegic patients. , 1961, Archives of physical medicine and rehabilitation.

[8]  S McEwen,et al.  Disablement following stroke. , 1999, Disability and rehabilitation.

[9]  Qin Zhang,et al.  Dual predictive control of electrically stimulated muscle using biofeedback for drop foot correction , 2011, 2011 IEEE/RSJ International Conference on Intelligent Robots and Systems.

[10]  D. Pradon,et al.  A pilot study to investigate the combined use of Botulinum toxin type-a and ankle foot orthosis for the treatment of spastic foot in chronic hemiplegic patients. , 2011, Clinical biomechanics.

[11]  Thomas Seel,et al.  Experimental Evaluation of a Novel Inertial Sensor Based Realtime Gait Phase De-tection Algorithm , 2015 .

[12]  Giancarlo Ferrigno,et al.  A Novel Adaptive, Real-Time Algorithm to Detect Gait Events From Wearable Sensors , 2015, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[13]  Jörg Raisch,et al.  Iterative learning control of a drop foot neuroprosthesis — Generating physiological foot motion in paretic gait by automatic feedback control , 2016 .

[14]  R W Bohannon,et al.  Gait performance of hemiparetic stroke patients: selected variables. , 1987, Archives of physical medicine and rehabilitation.

[15]  Roger Pissard-Gibollet,et al.  Continuous gait cycle index estimation for electrical stimulation assisted foot drop correction , 2014, Journal of NeuroEngineering and Rehabilitation.

[16]  Yoichi Shimada,et al.  Clinical application of acceleration sensor to detect the swing phase of stroke gait in functional electrical stimulation. , 2005, The Tohoku journal of experimental medicine.

[17]  P H Veltink,et al.  Three dimensional inertial sensing of foot movements for automatic tuning of a two-channel implantable drop-foot stimulator. , 2003, Medical engineering & physics.

[18]  H. Ring,et al.  Functional electrical stimulation-induced neural changes and recovery after stroke. , 2006, Europa medicophysica.

[19]  Marco Donati,et al.  An enhanced estimate of initial contact and final contact instants of time using lower trunk inertial sensor data. , 2012, Gait & posture.

[20]  Katherine S. Rudolph,et al.  Novel Patterns of Functional Electrical Stimulation Have an Immediate Effect on Dorsiflexor Muscle Function During Gait for People Poststroke , 2010, Physical Therapy.

[21]  G M Lyons,et al.  Effect of walking speed changes on tibialis anterior EMG during healthy gait for FES envelope design in drop foot correction. , 2007, Journal of electromyography and kinesiology : official journal of the International Society of Electrophysiological Kinesiology.

[22]  G M Lyons,et al.  A system for the delivery of programmable, adaptive stimulation intensity envelopes for drop foot correction applications. , 2006, Medical engineering & physics.