Independent quality assurance of a helical tomotherapy machine using the dose magnifying glass.

PURPOSE Helical tomotherapy is a complex delivery technique, integrating CT image guidance and intensity modulated radiotherapy in a single system. The integration of the CT detector ring on the gantry not only allows patient position verification but is also often used to perform various QA procedures. This convenience lacks the rigor of a machine-independent QA process. METHODS In this article, a Si strip detector, known as the Dose Magnifying Glass (DMG), was used to perform machine-independent QA measurements of the multileaf collimator alignment, leaf open time threshold, and leaf fluence output factor (LFOF). RESULTS The DMG measurements showed good agreements with EDR2 film for the MLC alignment test while the CT detector agrees well with DMG measurements for leaf open time threshold and LFOF measurements. The leaf open time threshold was found to be approximately 20 ms. The LFOF measured with the DMG agreed within error with the CT detector measured LFOF. CONCLUSIONS The DMG with its 0.2 mm spatial resolution coupled to TERA ASIC allowed real-time high temporal resolution measurements of the tomotherapy leaf movement. In conclusion, DMG was shown to be a suitable tool for machine-independent QA of a tomotherapy unit.

[1]  Niko Papanikolaou,et al.  QA for helical tomotherapy: Report of the AAPM Task Group 148a). , 2010, Medical physics.

[2]  M Petasecca,et al.  A silicon strip detector dose magnifying glass for IMRT dosimetry. , 2010, Medical physics.

[3]  D. Verellen,et al.  A diagnostic tool for basic daily quality assurance of a tomotherapy Hi•Art machine , 2009, Journal of applied clinical medical physics.

[4]  J. Balog,et al.  Helical tomotherapy quality assurance. , 2008, International journal of radiation oncology, biology, physics.

[5]  Tim Holmes,et al.  Helical tomotherapy dynamic quality assurance. , 2006, Medical physics.

[6]  T R Mackie,et al.  History of tomotherapy , 2006, Physics in medicine and biology.

[7]  T R Mackie,et al.  Quality assurance of a helical tomotherapy machine. , 2004, Physics in medicine and biology.

[8]  Robert Jeraj,et al.  Radiation characteristics of helical tomotherapy. , 2004, Medical physics.

[9]  John Balog,et al.  Clinical helical tomotherapy commissioning dosimetry. , 2003, Medical physics.

[10]  Robert Jeraj,et al.  Benchmarking beam alignment for a clinical helical tomotherapy device. , 2003, Medical physics.

[11]  T R Mackie,et al.  On the verification of the incident energy fluence in tomotherapy IMRT. , 2001, Physics in medicine and biology.

[12]  T R Mackie,et al.  Characterization of the output for helical delivery of intensity modulated slit beams. , 1999, Medical physics.