Influence of mechanical trauma of blood and hemolysis on PFA-100 testing

Although the appropriate quality of samples is essential for platelet function testing, information is lacking on interference from mechanical trauma of blood and hemolysis on PFA-100 analyzer. Citrated blood collected from nine healthy volunteers was divided into three aliquots. The first aliquot (‘A’) was processed without further manipulation, whereas the second and third were subjected to mechanical trauma by two (‘aliquot B’) or four passages (‘aliquot C’) through a very fine needle (30 gauge) to produce hemolysis and cell trauma mimicking poor sample collection. Samples were tested on PFA-100 and Advia 2120, and plasma then separated and tested for lactate dehydrogenase (LDH) and hemolysis index. Negligible hemolysis was present in aliquot A (hemolysis index 0.2 ± 0.1, cell-free hemoglobin 0–0.5 g/l), whereas an increasing amount was present in aliquots B (hemolysis index of 13.1 ± 1.8, cell-free hemoglobin 6.0–6.5 g/l) and C (hemolysis index 24.0 ± 1.1, cell-free hemoglobin 11.5–12.0 g/l). Increases in LDH, and concomitant reductions in platelet and red blood cell counts were observed in aliquots B and C. In hemolyzed aliquots B, four out of nine samples yielded ‘flow obstruction’ with both PFA-100 agonist cartridges, whereas the closure times were dramatically prolonged in the remaining five samples. In hemolyzed aliquots C, flow obstruction was recorded in six of nine samples for collagen and ADP and all samples for collagen and epinephrine, whereas closure times of collagen and ADP in the remaining three samples were dramatically prolonged. Mechanical trauma of blood causing hemolysis makes PFA-100 testing unreliable. When flow obstructions are observed, the potential presence of hemolysis should be investigated.

[1]  G. Lippi,et al.  Prevalence of hemolytic specimens referred for arterial blood gas analysis , 2011, Clinical chemistry and laboratory medicine.

[2]  E. Favaloro,et al.  External quality assurance for the PFA‐100® , 2011, Journal of thrombosis and haemostasis : JTH.

[3]  G. Lippi,et al.  Shortened activated partial thromboplastin time: causes and management , 2010, Blood coagulation & fibrinolysis : an international journal in haemostasis and thrombosis.

[4]  G. Lippi,et al.  Contemporary platelet function testing , 2010, Clinical chemistry and laboratory medicine.

[5]  G. Lippi,et al.  Laboratory reporting of hemostasis assays: the final post-analytical opportunity to reduce errors of clinical diagnosis in hemostasis? , 2010, Clinical chemistry and laboratory medicine.

[6]  E. Favaloro,et al.  Platelet Function Testing: Auditing Local Practice and Broader Implications , 2010, American Society for Clinical Laboratory Science.

[7]  G. Lippi,et al.  Laboratory assessment and perioperative management of patients on antiplatelet therapy: from the bench to the bedside. , 2009, Clinica chimica acta; international journal of clinical chemistry.

[8]  G. Lippi,et al.  Survey on the prevalence of hemolytic specimens in an academic hospital according to collection facility: opportunities for quality improvement , 2009, Clinical chemistry and laboratory medicine.

[9]  G. Fritsma,et al.  Whole blood platelet aggregometry and platelet function testing. , 2009, Seminars in thrombosis and hemostasis.

[10]  E. Favaloro Internal quality control and external quality assurance of platelet function tests. , 2009, Seminars in thrombosis and hemostasis.

[11]  Paul J. Harrison,et al.  Screening tests of platelet function: update on their appropriate uses for diagnostic testing. , 2009, Seminars in thrombosis and hemostasis.

[12]  Mario Plebani,et al.  Multicenter evaluation of the hemolysis index in automated clinical chemistry systems , 2009, Clinical chemistry and laboratory medicine.

[13]  Emmanuel J Favaloro,et al.  Clinical Utility of the PFA-100 , 2008, Seminars in thrombosis and hemostasis.

[14]  G. Lippi,et al.  Preanalytical and Postanalytical Variables: The Leading Causes of Diagnostic Error in Hemostasis? , 2008, Seminars in thrombosis and hemostasis.

[15]  Young-Uk Cho,et al.  Variables that affect platelet function analyzer-100 (PFA-100) closure times and establishment of reference intervals in Korean adults. , 2008, Annals of clinical and laboratory science.

[16]  Mario Plebani,et al.  Haemolysis: an overview of the leading cause of unsuitable specimens in clinical laboratories , 2008, Clinical chemistry and laboratory medicine.

[17]  G. Lippi,et al.  Prevalence and type of pre-analytical problems for inpatients samples in coagulation laboratory. , 2008, Journal of evaluation in clinical practice.

[18]  P. Vest,et al.  [Evaluation of the Unicel DXC 600 (Beckman Coulter) and comparison with Integra 800 (Roche Diagnostics)]. , 2007, Annales de biologie clinique.

[19]  E. Duncan,et al.  Utility of the PFA-100 as a screening test of platelet function: an audit of haemostasis laboratories in Australia and New Zealand , 2007, Blood Coagulation and Fibrinolysis.

[20]  Giuseppe Lippi,et al.  Influence of the needle bore size on platelet count and routine coagulation testing , 2006, Blood coagulation & fibrinolysis : an international journal in haemostasis and thrombosis.

[21]  G. Lippi,et al.  Quality and reliability of routine coagulation testing: can we trust that sample? , 2006, Blood coagulation & fibrinolysis : an international journal in haemostasis and thrombosis.

[22]  G. Lippi,et al.  Preanalytical variability: the dark side of the moon in laboratory testing , 2006, Clinical chemistry and laboratory medicine.

[23]  G. Lippi,et al.  Short-term venous stasis influences routine coagulation testing , 2005, Blood coagulation & fibrinolysis : an international journal in haemostasis and thrombosis.

[24]  G. Dimeski Effects of hemolysis on the Roche ammonia method for Hitachi analyzers. , 2004, Clinical chemistry.

[25]  V. Blanchette,et al.  Use of the PFA-100® in the Assessment of Primary, Platelet-Related Hemostasis in a Pediatric Setting , 1998, Seminars in thrombosis and hemostasis.

[26]  J. Barbenel,et al.  A Critical Investigation into the Existence of Circulating Platelet Aggregates , 1986, Thrombosis and Haemostasis.

[27]  S. Meites Letter: Reproducibly simulating hemolysis, for evaluating its interference with chemical methods. , 1973, Clinical chemistry.

[28]  J. Lovelock,et al.  The haemolysis of human red blood-cells by freezing and thawing. , 1953, Biochimica et biophysica acta.