A new computational fluid dynamics method for in-depth investigation of flow dynamics in roller pump systems.

Roller pumps are commonly used in circulatory assist devices to deliver blood, but the inherent high mechanical stresses (especially wall shear stress) may cause considerable damage to cells. Conventional experimental approaches to evaluate and reduce device-induced cell damage require considerable effort and resources. In this work, we describe the use of a new computational fluid dynamics method to more effectively study roller pump systems. A generalized parametric model for the fluid field in a typical roller pump system is presented first, and analytical formulations of the moving boundary are then derived. Based on the model and formulations, the dynamic geometry and mesh of the fluid field can be updated automatically according to the time-dependent roller positions. The described method successfully simulated the pulsing flow generated by the pump, offering a convenient way to visualize the inherent flow pattern and to assess shear-induced cell damage. Moreover, the highly reconfigurable model and the semiautomated simulation process extend the usefulness of the presented method to a wider range of applications. Comparison studies were conducted, and valuable indications about the detailed effects of structural parameters and operational conditions on the produced wall shear stress were obtained. Given the good consistency between the simulated results and the existing experimental data, the presented method displays promising potential to more effectively guide the development of improved roller pump systems which produce less mechanical damage to cells.

[1]  G. Noon,et al.  Reduction of blood trauma in roller pumps for long-term perfusion , 1985, World Journal of Surgery.

[2]  Daniel L. Timms,et al.  Axial Magnetic Bearing Development for the BiVACOR Rotary BiVAD/TAH , 2010, IEEE Transactions on Biomedical Engineering.

[3]  H Reul,et al.  Assessment of hemolysis related quantities in a microaxial blood pump by computational fluid dynamics. , 2001, Artificial organs.

[4]  Bartley P Griffith,et al.  Computational and experimental evaluation of the fluid dynamics and hemocompatibility of the CentriMag blood pump. , 2006, Artificial organs.

[5]  Pingan Du,et al.  A dilution-filtration system for removing cryoprotective agents. , 2011, Journal of biomechanical engineering.

[6]  C Bludszuweit,et al.  Three-dimensional numerical prediction of stress loading of blood particles in a centrifugal pump. , 1995, Artificial organs.

[7]  D. Ku BLOOD FLOW IN ARTERIES , 1997 .

[8]  C Bludszuweit,et al.  Model for a general mechanical blood damage prediction. , 1995, Artificial organs.

[9]  W. Buurman,et al.  Hemolysis is associated with acute kidney injury during major aortic surgery. , 2010, Kidney international.

[10]  H. Reul,et al.  Blood pumps for circulatory support , 2000, Perfusion.

[11]  Yukihiko Nose,et al.  Comparison of Centrifugal and Roller Pump Hemolysis Rates at Low Flow. , 1996, Artificial organs.

[12]  Maik Foltan,et al.  Prospective randomized clinical study of arterial pumps used for routine on pump coronary bypass grafting. , 2011, Artificial organs.

[13]  J. Bergström,et al.  Fluid-Structure Interaction Analysis of a Peristaltic Pump , 2011 .

[14]  James F. Antaki,et al.  Quantitative Evaluation of Blood Damage in a Centrifugal VAD by Computational Fluid Dynamics , 2004 .

[15]  A. Bhutta,et al.  Hemolysis During Cardiac Extracorporeal Membrane Oxygenation: A Case-Control Comparison of Roller Pumps and Centrifugal Pumps in a Pediatric Population , 2011, ASAIO journal.

[16]  Mehdi Behbahani,et al.  A review of computational fluid dynamics analysis of blood pumps , 2009, European Journal of Applied Mathematics.

[17]  Akif Ündar,et al.  A nonocclusive, inexpensive pediatric pulsatile roller pump for cardiopulmonary bypass, extracorporeal life support, and left/right ventricular assist systems. , 2013, Artificial organs.

[18]  K A Crane,et al.  Further decrease in subclinical hemolysis utilizing 12.7 mm tubing in the arterial roller head. , 1983, The Annals of thoracic surgery.

[19]  James F Antaki,et al.  Computational fluid dynamics analysis of blade tip clearances on hemodynamic performance and blood damage in a centrifugal ventricular assist device. , 2009, Artificial organs.

[20]  J. Shelton,et al.  Investigation and quantification of the blood trauma caused by the combined dynamic forces experienced during cardiopulmonary bypass , 2000, Perfusion.

[21]  H MacGregor,et al.  Effects of centrifugal and roller pumps on survival of autologous red cells in cardiopulmonary bypass surgery , 2006, Perfusion.

[22]  Nigel H. Lovell,et al.  Parameter-Optimized Model of Cardiovascular–Rotary Blood Pump Interactions , 2010, IEEE Transactions on Biomedical Engineering.

[23]  Xiaoyu Luo,et al.  Blood flow and damage by the roller pumps during cardiopulmonary bypass , 2005 .

[24]  Tohru Takaseya,et al.  Non-occlusive condition with the Better-Header roller pump: impacts of flow dynamics and hemolysis. , 2004, Annals of thoracic and cardiovascular surgery : official journal of the Association of Thoracic and Cardiovascular Surgeons of Asia.