Radiation dose of non-enhanced chest CT can be reduced 40% by using iterative reconstruction in image space.

AIM To evaluate the image quality and dose reduction capability of non-enhanced chest computed tomography (CT) examinations using iterative reconstruction in image space (IRIS). MATERIALS AND METHODS A CT water phantom was scanned at 120 kV/150 mAs and 100 kV/270 mAs as the reference, and the tube current was decreased in 10% intervals down to 40% of the reference value. Image noise was evaluated and compared between filtered back-projection (FBP) and IRIS reconstructed data. In the patient study, 90 patients underwent non-enhanced chest CT examinations; the patients were randomly assigned into three groups: group A (n=30) standard dose protocol, 120 kV/110 mAs; group B (n=30) low dose, 100 kV/110 mAs; group C (n=30) low dose, 120 kV/67 mAs. All images were reconstructed by FBP and IRIS algorithm using matched kernels of B30 and I30. The objective image noise (OIN), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) of the aorta and latissimus dorsi were measured. The subjective image quality and noise were scored using a three-point scale by two experienced radiologists. The results of the subjective and objective image assessment were compared between groups B and C (low dose) IRIS and group A (standard dose) FBP. RESULTS The phantom study showed comparable image noise between the scans using 60% dose with IRIS and 100% dose with FBP for both 120 and 100 kV. In the patient study, groups A, B, and C had effective dose of 3.81 ± 0.43, 2.40 ± 0.19, and 2.41 ± 0.15 mSv. IRIS significantly improved the OIN, SNR, and CNR compared with FBP for the same patient. The OIN, SNR, and CNR using IRIS in group B and C were improved or comparable to those in group A using FBP. No significant difference was found in subjective image quality and noise between groups B and C using IRIS and group A using FBP. CONCLUSION Compared with FBP, IRIS can maintain or improve image quality on unenhanced chest CT image reconstruction while saving 40% radiation dose.

[1]  A. Beelen,et al.  Interobserver agreement on MRI evaluation of patients with cervical radiculopathy. , 2011, Clinical radiology.

[2]  L. Goldman Principles of CT and CT Technology* , 2007, Journal of Nuclear Medicine Technology.

[3]  G. Hounsfield Computerized transverse axial scanning (tomography): Part I. Description of system. 1973. , 1973, The British journal of radiology.

[4]  G. Hounsfield Computerized transverse axial scanning (tomography). 1. Description of system. , 1973, The British journal of radiology.

[5]  D. Sahani,et al.  Reducing Abdominal CT Radiation Dose With Adaptive Statistical Iterative Reconstruction Technique , 2010, Investigative radiology.

[6]  B. Vastag Lung screening study to test popular CT scans. , 2002, JAMA.

[7]  David J Brenner,et al.  Radiation risks potentially associated with low-dose CT screening of adult smokers for lung cancer. , 2004, Radiology.

[8]  T. Mulkens,et al.  Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. , 2005, Radiology.

[9]  D. Frush,et al.  Helical CT in children: technical considerations and body applications. , 1998, Radiology.

[10]  E. Samei,et al.  Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm--initial clinical experience. , 2010, Radiology.

[11]  J. Remy,et al.  Chest computed tomography using iterative reconstruction vs filtered back projection (Part 2): image quality of low-dose CT examinations in 80 patients , 2011, European Radiology.

[12]  J. Remy,et al.  Chest computed tomography using iterative reconstruction vs filtered back projection (Part 1): evaluation of image noise reduction in 32 patients , 2011, European Radiology.

[13]  G. Bongartz,et al.  Diagnostic performance of low-dose CT for the detection of urolithiasis: a meta-analysis. , 2008, AJR. American journal of roentgenology.

[14]  A. Einstein,et al.  Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. , 2007, JAMA.

[15]  M. Kalra,et al.  Strategies for CT radiation dose optimization. , 2004, Radiology.

[16]  L. Hayman,et al.  Pregnancy and women radiologists. , 1982, Radiology.

[17]  A. Beck,et al.  Feasibility of MDCT Colonography in ultra-low-dose technique in the detection of colorectal lesions: comparison with high-resolution video colonoscopy. , 2004, AJR. American journal of roentgenology.

[18]  S. J. Golding,et al.  The Quality Criteria for Computed Tomography , 1998 .

[19]  D. Frush,et al.  Dose reduction in paediatric MDCT: general principles. , 2007, Clinical radiology.

[20]  Mannudeep K. Kalra,et al.  In-Plane Shielding for CT: Effect of Off-Centering, Automatic Exposure Control and Shield-to-Surface Distance , 2009, Korean journal of radiology.

[21]  Natalie N. Braun,et al.  Strategies for reducing radiation dose in CT. , 2009, Radiologic clinics of North America.

[22]  Fuminari Tatsugami,et al.  Evaluation of a body mass index-adapted protocol for low-dose 64-MDCT coronary angiography with prospective ECG triggering. , 2009, AJR. American journal of roentgenology.

[23]  Alvin C. Silva,et al.  Iterative Reconstruction Technique for Reducing Body Radiation Dose at Ct: Feasibility Study Hara Et Al. Ct Iterative Reconstruction Technique Gastrointestinal Imaging Original Research , 2022 .

[24]  J. Hua,et al.  Prospective ECG triggering versus low-dose retrospective ECG-gated 128-channel CT coronary angiography: comparison of image quality and radiation dose. , 2010, Clinical radiology.

[25]  C. Heyer,et al.  Image quality and radiation exposure at pulmonary CT angiography with 100- or 120-kVp protocol: prospective randomized study. , 2007, Radiology.

[26]  Simon Wildermuth,et al.  Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose , 2008, European Radiology.

[27]  W. Kalender,et al.  Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT angiography using dual source CT: technique and initial experience , 2009, European Radiology.

[28]  M. Kalra,et al.  Radiation Dose Reduction With Chest Computed Tomography Using Adaptive Statistical Iterative Reconstruction Technique: Initial Experience , 2010, Journal of computer assisted tomography.

[29]  M. L. R. D. Christenson,et al.  Image Quality and Radiation Exposure at Pulmonary CT Angiography with 100- or 120-kVp Protocol: Prospective Randomized Study , 2008 .

[30]  Jiang Hsieh,et al.  Diffuse lung disease: CT of the chest with adaptive statistical iterative reconstruction technique. , 2010, Radiology.

[31]  Takeshi Nakaura,et al.  Abdominal CT with low tube voltage: preliminary observations about radiation dose, contrast enhancement, image quality, and noise. , 2005, Radiology.

[32]  L. Wagner,et al.  Exposure of the pregnant patient to diagnostic radiations : a guide to medical management , 1985 .