Wear analysis in anatomical and reversed shoulder prostheses

This paper describes the development of a computational model to calculate wear rates in total shoulder prostheses, for a 5–150 degrees arm abduction. Anatomical keeled and pegged prosthesis as well as reversed prosthesis were the studied implants. The bone models were built based on computed tomography (CT) images and using a computer aided design-based modelling pipeline. The finite element method was used to solve the contact problem between the surface of the polyethylene (PE) components and the corresponding articular component. The aim of this work was to determine linear and volumetric PE wear, for several radial mismatches, in conditions of pathological (rheumatoid arthritis) and non-pathological bone. Results showed that contact pressures and linear wear developed in anatomical prosthesis were higher than those visualised in reversed prosthesis. However, anatomical prosthesis exhibited a better volumetric wear performance. Moreover, our findings indicated higher values of volumetric wear in higher congruent models and on pathological bone conditions.

[1]  T D Brown,et al.  Finite element analysis of acetabular wear. Validation, and backing and fixation effects. , 1997, Clinical orthopaedics and related research.

[2]  A. Edidin,et al.  Influence of mechanical behavior on the wear of 4 clinically relevant polymeric biomaterials in a hip simulator. , 2000, The Journal of arthroplasty.

[3]  C. M. Agrawal,et al.  Isolation and characterization of polyethylene wear debris associated with osteolysis following total shoulder arthroplasty. , 1999, The Journal of bone and joint surgery. American volume.

[4]  F. V. D. van der Helm Analysis of the kinematic and dynamic behavior of the shoulder mechanism. , 1994, Journal of biomechanics.

[5]  Kamiar Aminian,et al.  Arm position during daily activity. , 2008, Gait & posture.

[6]  J. Levy,et al.  Reverse shoulder prosthesis for acute four-part fracture: tuberosity fixation using a horseshoe graft. , 2011, Journal of orthopaedic trauma.

[7]  T. Wright,et al.  The glenoid in shoulder arthroplasty. , 2009, Journal of shoulder and elbow surgery.

[8]  C A Rockwood,et al.  Complications of total shoulder-replacement arthroplasty. , 1996, The Journal of bone and joint surgery. American volume.

[9]  Paulo R Fernandes,et al.  Computational hip joint simulator for wear and heat generation. , 2007, Journal of biomechanics.

[10]  Kamal I. Bohsali,et al.  Complications of total shoulder arthroplasty. , 2006, The Journal of bone and joint surgery. American volume.

[11]  Pascal Boileau,et al.  Grammont reverse prosthesis: design, rationale, and biomechanics. , 2005, Journal of shoulder and elbow surgery.

[12]  P J Prendergast,et al.  Contact stresses in the glenoid component in total shoulder arthroplasty , 2003, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[13]  F. Linde,et al.  X-ray quantitative computed tomography: the relations to physical properties of proximal tibial trabecular bone specimens. , 1989, Journal of biomechanics.

[14]  Frederick A Matsen,et al.  The complex characteristics of 282 unsatisfactory shoulder arthroplasties. , 2007, Journal of shoulder and elbow surgery.

[15]  John Fisher,et al.  The role of macrophages in osteolysis of total joint replacement. , 2005, Biomaterials.

[16]  F. C. T. Helm,et al.  Analysis of the kinematic and dynamic behavior of the shoulder mechanism , 1994 .

[17]  The biomechanics of total shoulder arthroplasty , 2002 .

[18]  Andreas Kontaxis,et al.  Adaptation of scapula lateral rotation after reverse anatomy shoulder replacement , 2008, Computer methods in biomechanics and biomedical engineering.

[19]  J. Lynch,et al.  Glenoid component failure in total shoulder arthroplasty. , 2008, The Journal of bone and joint surgery. American volume.

[20]  G. Walch,et al.  Scapular notching in reverse shoulder arthroplasty. , 2008, Journal of shoulder and elbow surgery.

[21]  Jove Graham,et al.  Retrieved glenoid components: a classification system for surface damage analysis. , 2002, The Journal of arthroplasty.

[22]  J. O. Søjbjerg,et al.  Bone strength and material properties of the glenoid. , 1997, Journal of shoulder and elbow surgery.

[23]  F A Matsen,et al.  Observations on retrieved polyethylene glenoid components. , 2001, The Journal of arthroplasty.

[24]  Mark Taylor,et al.  Wear in the prosthetic shoulder: association with design parameters. , 2007, Journal of biomechanical engineering.

[25]  T. Wright,et al.  Observations on retrieved glenoid components from total shoulder arthroplasty. , 2009, Journal of shoulder and elbow surgery.

[26]  G R Johnson,et al.  The biomechanics of reverse anatomy shoulder replacement--a modelling study. , 2009, Clinical biomechanics.

[27]  S. Cowin,et al.  On the dependence of the elasticity and strength of cancellous bone on apparent density. , 1988, Journal of biomechanics.

[28]  A Terrier,et al.  Comparison of polyethylene wear in anatomical and reversed shoulder prostheses. , 2009, The Journal of bone and joint surgery. British volume.

[29]  J. Archard Contact and Rubbing of Flat Surfaces , 1953 .

[30]  L. Hubert,et al.  Reverse shoulder arthroplasty for the treatment of three- and four-part fractures of the proximal humerus in the elderly: a prospective review of 43 cases with a short-term follow-up. , 2007, The Journal of bone and joint surgery. British volume.

[31]  F T Ballmer,et al.  Observations on retrieved glenoid components. , 2003, The Journal of arthroplasty.

[32]  Gilles Walch,et al.  Reverse total shoulder arthroplasty: a review of results according to etiology. , 2007, The Journal of bone and joint surgery. American volume.

[33]  P J Prendergast,et al.  Three-dimensional finite element analysis of glenoid replacement prostheses: a comparison of keeled and pegged anchorage systems. , 2000, Journal of biomechanical engineering.

[34]  D. Burr,et al.  Stiffness of compact bone: effects of porosity and density. , 1988, Journal of biomechanics.

[35]  P R Fernandes,et al.  Multilayered learning architecture applied to humanoid locomotion , 2009, Computer methods in biomechanics and biomedical engineering.