Validation and application of high performance liquid chromatographic method for the estimation of metoclopramide hydrochloride in plasma.

The objective of this study was validation of a reverse phase HPLC method for the estimation of metoclopramide HCl in plasma already validated for determination of metoclopramide HCl in tablets dosage form. A reverse chromatographic method was used for estimation of metoclopramide HCl with the mobile phase of acetonitrile, 20mM potassium dihydrogen phosphate buffer solution (pH 3.0 adjusted with orthophosphoric acid) in the ratio of 40: 60. The column used was Waters C18 3.9×300mm µBondapak (RP). The flow rate of the mobile phase was 2ml/ minute. The detector was set at the wavelength of 275nm. This method validated in plasma and was found to be linear, with correlation coefficient (R2), value of 0.9988, in the range of 48 ng/ml-0.25ng/ml. The method modified was accurate, precise, sensitive and showed good stability results. The % RSD of the retention time and peak area of metoclopramide HCl was 0.19% and 1.44% respectively. All the parameters such as specificity, linearity, range, accuracy, precision, system suitability, solution stability, detection and quantification limits were evaluated to validate this method and were found within the acceptance limits. The method can be effectively used for estimation of metoclopramide HCl in plasma.

[1]  N. Yamagishi,et al.  Simple and sensitive method for measurement of metoclopramide in cattle plasma by LC-MS/MS using a multimode chromatography. , 2013, The Journal of veterinary medical science.

[2]  J. Inamadugu,et al.  Sensitive and selective liquid chromatography-tandem mass spectrometry method for the determination of metoclopramide in human plasma: application to a bioequivalence study. , 2010, Biomedical chromatography : BMC.

[3]  M. Yan,et al.  Determination of metoclopramide in human plasma by LC-ESI-MS and its application to bioequivalance studies. , 2010, Journal of chromatography. B, Analytical technologies in the biomedical and life sciences.

[4]  M. Javanbakht,et al.  Novel molecularly imprinted polymers for the selective extraction and determination of metoclopramide in human serum and urine samples using high-performance liquid chromatography. , 2009, Journal of chromatography. B, Analytical technologies in the biomedical and life sciences.

[5]  Tatsuro Sekiya,et al.  [Residue analysis of metoclopramide in bovine and swine tissues by HPLC and administration experiment]. , 2006, Shokuhin eiseigaku zasshi. Journal of the Food Hygienic Society of Japan.

[6]  L. Vlase,et al.  Pharmacokinetic interaction between fluoxetine and metoclopramide in healthy volunteers , 2006, Biopharmaceutics & drug disposition.

[7]  A. Lister 7 - Validation of HPLC Methods in Pharmaceutical Analysis , 2005 .

[8]  G. Amidon,et al.  Molecular properties of WHO essential drugs and provisional biopharmaceutical classification. , 2004, Molecular pharmaceutics.

[9]  M. De Amici,et al.  Synthesis of a metabolite of metoclopramide and its detection in human urine. , 1994, Farmaco.

[10]  A. Mutlib,et al.  Determination of metoclopramide and two of its metabolites using a sensitive and selective gas chromatographic-mass spectrometric assay. , 1994, Journal of chromatography. B, Biomedical applications.

[11]  D. Buss,et al.  A Rapid Liquid Chromatographic Method for the Determination of Metoclopramide in Human Plasma , 1990, Therapeutic drug monitoring.

[12]  M. Nieder,et al.  High performance liquid chromatographic determination of metoclopramide in plasma and urine and its application to biopharmaceutical investigations , 1987 .

[13]  E. Mcgovern,et al.  EVALUATION OF A HIGH PRESSURE LIQUID CHROMATOGRAPHIC TECHNIQUE FOR METOCLOPRAMIDE ANALYSIS , 1984, Journal of clinical and hospital pharmacy.

[14]  D. Bateman,et al.  The pharmacokinetics of metoclopramide in man with observations in the dog. , 1980, British journal of clinical pharmacology.

[15]  D. S. Smith,et al.  Current Concepts in Diabetic Gastroparesis , 2012, Drugs.

[16]  V. Kadam,et al.  Development and Validation of Stability Indicating HPLC Method for Estimation of Metoclopramide Hydrochloride From A Novel Formulation , 2009 .

[17]  Jordan L. Cohen,et al.  High Pressure Liquid Chromatographic Analysis of Metoclopramide in Serum , 2004, Pharmaceutical Research.

[18]  G. Williams,et al.  Analysis of Metoclopramide and Related Compounds in Tablets by Liquid Chromatography , 1989 .

[19]  W. Behrendt,et al.  Bioavailability of controlled-release metoclopramide. 1st communication: single dose study. , 1987, Arzneimittel-Forschung.