Ultra-low flow rate measurement techniques

Abstract This paper outlines the state-of-the-art techniques to measure ultra-low flow rates down to 5 nL/min that have been developed by eight European organisations in an EMPIR collaborative research project. The techniques include gravimetric, optical interface tracking, optical pendant drop, micro Particle Image Velocimetry (μPIV) and displacement methods. The measurement principles, experimental setup, results and measurement uncertainties are summarised. The techniques and uncertainties developed are currently being validated through an intercomparison exercise using three transfer standards. This comparison will test steady flows and fast changing flow rates. The new techniques developed will be used to test and calibrate medical infusion devices, micro- and nano-fluidic systems and other ultra-low flow rate devices such as perfusion systems for organ-on-a-chip applications.

[1]  Rui Martins,et al.  Primary standards for measuring flow rates from 100 nl/min to 1 ml/min – gravimetric principle , 2015, Biomedizinische Technik. Biomedical engineering.

[2]  Norbert Stoll,et al.  An Optical Approach for the Determination of Droplet Volumes in Nanodispensing , 2009, Journal of automated methods & management in chemistry.

[3]  Elsa Batista,et al.  Development of an experimental setup for microflow measurement using interferometry , 2020, Flow Measurement and Instrumentation.

[4]  Peter Lucas,et al.  Metrology for drug delivery , 2015, Biomedizinische Technik. Biomedical engineering.

[5]  J. Alves e Sousa,et al.  New EMPIR project – Metrology for Drug Delivery , 2020 .