Comparison of the Bone Anchored Hearing Aid Implantable Hearing Device with Contralateral Routing of Offside Signal Amplification in the Rehabilitation of Unilateral Deafness

Objective Monaural hearing imposes constraints under many listening conditions. The authors compared the effects of a semi-implantable bone conductor, the Entific bone anchored hearing aid, with conventional contralateral routing of offside signal amplification to assess rehabilitative benefit in adults with unilateral deafness. Design Prospective trials of subjects with unilateral deafness using benefit surveys, source identification testing, and hearing in noise testing. Setting Tertiary referral center, outpatient surgical and audiologic services. Patients Adults with unilateral deafness (pure tone average >90 dB, SD <20%) after acoustic neuroma excision (n = 7), meningitis (n = 1), sudden sensorineural hearing loss (n = 1), and sudden sensorineural hearing loss with chronic suppurative otitis media (n = 1). Entry criteria included normal hearing in the contralateral ear (pure tone average <25 dB, SD >80%). Intervention Subjects were fitted with contralateral routing of offside signal amplification devices for 1 month and tested with contralateral routing of offside signal before mastoid implantation of the deaf ear, fitting, and testing for bone anchored hearing aid. Outcome measures Subjects' assessment of experience with their devices and patterns of use, 2) source azimuth identification in noise test, and 3) speech discrimination in quiet and in noise under conditions of noise-front, noise-to-normal-ear, and noise-to-deaf-ear. Results There was consistent satisfaction with bone anchored hearing aid implantation and amplification, and poor acceptance of contralateral routing of offside signal amplification. Sound localization was poor at baseline and with both bone anchored hearing aid and contralateral routing of offside signal. Relative to baseline, contralateral routing of offside signal and bone anchored hearing aid produced significantly better speech recognition in noise under most conditions. The bone anchored hearing aid enabled significantly better speech recognition than contralateral routing of offside signal in quiet and in a composite of noise conditions. The advantages may relate to averting the interference of speech signals delivered to the better ear, as occurs with conventional contralateral routing of offside signal amplification. Conclusions Preliminary data in subjects with normal monaural hearing indicate that vibromechanical stimulation with the bone anchored hearing aid overcomes some of the negative head shadow effects in unilateral deafness. The bone anchored hearing aid system, when placed on the side of a deaf ear, yields greater benefit in subjects with normal monaural hearing than does contralateral routing of offside signal amplification. It seems that this rehabilitative approach can expand the sound field of monaural listeners in further enhancing speech understanding. Observations suggest that further understanding of bone conduction as implemented in transcranial stimulation will guide further options for patients with monaural hearing. Longer follow-up will help to determine whether communicative skill improvements with the bone anchored hearing aid outweigh the disadvantages of implantation surgery, costs, and device maintenance.

[1]  C. Hollenbeak,et al.  The Minimum Speech Test Battery in Profound Unilateral Hearing Loss , 2001, Otology and Neurotology.

[2]  C A Syms,et al.  Hearing Rehabilitation Using the BAHA Bone-Anchored Hearing Aid: Results in 40 Patients , 2001, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.

[3]  D E Hayes,et al.  Bone-conduction amplification with completely-in-the-canal hearing aids. , 1998, Journal of the American Academy of Audiology.

[4]  Michelsen,et al.  Directional sound processing and interaural sound transmission in a small and a large grasshopper , 1995, The Journal of experimental biology.

[5]  Robyn M. Cox,et al.  The Abbreviated Profile of Hearing Aid Benefit , 1995, Ear and hearing.

[6]  C. Cremers,et al.  A one-stage surgical procedure for placement of percutaneous implants for the bone-anchored hearing aid , 1994, The Journal of Laryngology & Otology.

[7]  S. Soli,et al.  Development of the Hearing in Noise Test for the measurement of speech reception thresholds in quiet and in noise. , 1994, The Journal of the Acoustical Society of America.

[8]  D. Welling,et al.  Unilateral Sensorineural Hearing Loss Rehabilitation , 1991, Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery.

[9]  Peder Carlsson,et al.  Ten years of experience with the Swedish bone-anchored hearing system. , 1990, The Annals of otology, rhinology & laryngology. Supplement.

[10]  B. Håkansson,et al.  Percutaneous Vs. Transcutaneous Transducers for Hearing by Direct Bone Conduction , 1990, Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery.

[11]  D. Gibson Interaural crosstalk in the cat , 1982, Hearing Research.

[12]  A Tjellström,et al.  Osseointegrated titanium implants in the temporal bone. A clinical study on bone-anchored hearing aids. , 1981, The American journal of otology.

[13]  J. Tonndorf A new concept of bone conduction. , 1968, Archives of otolaryngology.

[14]  S Stenfelt,et al.  Vibration characteristics of bone conducted sound in vitro. , 2000, The Journal of the Acoustical Society of America.

[15]  Anders Brandt On Sound Transmission Characteristics of the Human Skull in vivo , 1989 .

[16]  A Tjellström,et al.  Hearing thresholds with direct bone conduction versus conventional bone conduction. , 1984, Scandinavian audiology.

[17]  P. Branemark,et al.  Analysis of the mechanical impedance of bone-anchored hearing aids. , 1980, Acta oto-laryngologica.

[18]  R. Butler The monaural localization of tonal stimuli , 1971 .

[19]  Fritz Genz,et al.  Über einen Fall von Cylindrom des weichen Gaumens , 1938 .

[20]  Georg v. Békésy,et al.  Zur Theorie des Hörens bei der Schallaufnahme durch Knochenleitung , 1932 .