Modeling of a charge coupled device based optical tomographic instrumentation system for particle sizing

Abstract This research investigates the use of charge coupled device (CCD) linear image sensors in an optical tomographic instrumentation system used for sizing particles. Four CCD linear image sensors are configured around an octagonal shaped flow pipe for a four projections system. The measurement system used four CCD linear image sensors consisting of 2048 pixels with a pixel size of 14 by 14 μm. Hence, a high-resolution system is produced. Three mathematical models based on the effects due to particles, light sources and diffraction are discussed. The models are used to simulate the actual process in order to understand the limitations of the designed system.

[1]  W. G. Driscoll,et al.  Handbook of optics , 1978 .

[2]  Yordan Kostov,et al.  Low-cost optical instrumentation for biomedical measurements , 2000 .

[3]  R. Cubeddu,et al.  Optical Tomography , 1998, Technical Digest. 1998 EQEC. European Quantum Electronics Conference (Cat. No.98TH8326).

[4]  R. S. Longhurst Geometrical and Physical Optics , 1967 .

[5]  R. Abdul. Rahim A tomographic imaging system for pneumatic conveyors using optical fibres , 1996 .

[6]  R. A. Williams,et al.  Chapter 1 – Introduction to process tomography , 1995 .

[7]  Robert West In industry, seeing is believing , 2003 .

[8]  Richard A Williams,et al.  Process tomography: a European innovation and its applications , 1996 .

[9]  Michel P. Bonin,et al.  Laser-based techniques for particle-size measurement: A review of sizing methods and their industrial applications , 1996 .

[10]  K H Elliott,et al.  The use of commercial CCD cameras as linear detectors in the physics undergraduate teaching laboratory , 1998 .

[11]  R. G. Green,et al.  Fibre optic lens modelling for optical tomography , 1999 .

[12]  T. Allen Particle Size Measurement , 1965, Nature.

[13]  Brian S. Hoyle,et al.  Electrical capacitance tomography for flow imaging: system model for development of image reconstruction algorithms and design of primary sensors , 1992 .

[14]  Joaquim Ferreira,et al.  An overview of electromagnetic inductance tomography: Description of three different systems , 1996 .

[15]  W. H. Boyes,et al.  Instrumentation reference book , 2003 .

[16]  Wuqiang Yang,et al.  An image-reconstruction algorithm based on Landweber's iteration method for electrical-capacitance tomography , 1999 .

[17]  Koichi Hishida,et al.  Simultaneous Measurement of Velocity and Equivalent Diameter on Non‐Spherical Particles , 1994 .

[18]  R. G. Green,et al.  Sensor systems for lightly loaded pneumatic conveyors , 1998 .