Improved Semi-Automated 3D Kinematic Estimation of Total Knee Arthroplasty Using X-Ray Fluoroscopy

Purpose:To materialize 3D kinematic analysis of total knee arthroplasty (TKA), 2D/3D registration techniques, which use X-ray fluoroscopic images and the knee implants CAD, have been applied to clinical cases. However, most conventional methods have needed time-consuming and labor-intensive manual operations in some process. In previous study, we addressed a manual operations problem when setting initial pose of implants model for 2D/3D registration, and reported a semi-automated initial pose estimation method based on an interpolation technique. However, this method still required appropriate initial pose estimation of the model with manual operations for some X-ray images (key frames). Additionally, in the situation like fast knee motion and use of low frame rate, good registration results were not obtained because of the large displacement between each frame silhouette. To overcome these problems, this study proposes an improved semi-automated 3D kinematic estimation method.Methods:Our 2D/3D registrati...