Approach to modelling of microstrip-microslot applicator
暂无分享,去创建一个
Microstrip-slot applicators are used for biomedical applications, but their modelling is not easy. We present an approach which considers this type of structure on a lossy medium as a succession of microstrip lines with tuning septums of different sizes. Using a classical spectral-domain approach (SDA) model, calculations of the resonant frequency have been performed and compared with experimental measurements for different simple structures of microstrip-slot applicators.
[1] J. R. James,et al. New Low-Profile Applicators for Local Heating of Tissues , 1984, IEEE Transactions on Biomedical Engineering.
[2] S. Toutain,et al. Dispersion characteristics and energy configurations of a microstrip line with tuning septums laid on a lossy material , 1983 .
[3] M. A. Stuchly,et al. A New Microstrip Radiator For Medical Applications , 1980 .
[4] M. Chive,et al. Microstrip microslot antennas for biomedical applications: frequency analysis of different parameters of this type of applicator , 1985 .