Impact of the MLC on the MRI field distortion of a prototype MRI-linac.

PURPOSE To cope with intrafraction tumor motion, integrated MRI-linac systems for real-time image guidance are currently under development. The multileaf collimator (MLC) is a key component in every state-of-the-art radiotherapy treatment system, allowing for accurate field shaping and tumor tracking. This work quantifies the magnetic impact of a widely used MLC on the MRI field homogeneity for such a modality. METHODS The finite element method was employed to model a MRI-linac assembly comprised of a 1.0 T split-bore MRI magnet and the key ferromagnetic components of a Varian Millennium 120 MLC, namely, the leaves and motors. Full 3D magnetic field maps of the system were generated. From these field maps, the peak-to-peak distortion within the MRI imaging volume was evaluated over a 30 cm diameter sphere volume (DSV) around the isocenter and compared to a maximum preshim inhomogeneity of 300 μT. Five parametric studies were performed: (1) The source-to-isocenter distance (SID) was varied from 100 to 200 cm, to span the range of a compact system to that with lower magnetic coupling. (2) The MLC model was changed from leaves only to leaves with motors, to determine the contribution to the total distortion caused by MLC leaves and motors separately. (3) The system was configured in the inline or perpendicular orientation, i.e., the linac treatment beam was oriented parallel or perpendicular to the magnetic field direction. (4) The treatment field size was varied from 0 × 0 to 20×20 cm(2), to span the range of clinical treatment fields. (5) The coil currents were scaled linearly to produce magnetic field strengths B0 of 0.5, 1.0, and 1.5 T, to estimate how the MLC impact changes with B0. RESULTS (1) The MLC-induced MRI field distortion fell continuously with increasing SID. (2) MLC leaves and motors were found to contribute to the distortion in approximately equal measure. (3) Due to faster falloff of the fringe field, the field distortion was generally smaller in the perpendicular beam orientation. The peak-to-peak DSV distortion was below 300 μT at SID≥130 cm (perpendicular) and SID≥140 cm (inline) for the 1.0 T design. (4) The simulation of different treatment fields was identified to cause dynamic changes in the field distribution. However, the estimated residual distortion was below 1.2 mm geometric distortion at SID≥120 cm (perpendicular) and SID≥130 cm (inline) for a 10 mT/m frequency-encoding gradient. (5) Due to magnetic saturation of the MLC materials, the field distortion remained constant at B0>1.0 T. CONCLUSIONS This work shows that the MRI field distortions caused by the MLC cannot be ignored and must be thoroughly investigated for any MRI-linac system. The numeric distortion values obtained for our 1.0 T magnet may vary for other magnet designs with substantially different fringe fields, however the concept of modest increases in the SID to reduce the distortion to a shimmable level is generally applicable.

[1]  B W Raaymakers,et al.  Magnetic-field-induced dose effects in MR-guided radiotherapy systems: dependence on the magnetic field strength , 2008, Physics in medicine and biology.

[2]  Stuart Crozier,et al.  Flanged-edge transverse gradient coil design for a hybrid LINAC-MRI system. , 2013, Journal of magnetic resonance.

[3]  J Yun,et al.  Brushed permanent magnet DC MLC motor operation in an external magnetic field. , 2010, Medical physics.

[4]  Sasa Mutic,et al.  A Device for Realtime 3D Image-Guided IMRT , 2005 .

[5]  Dirk Verellen,et al.  A (short) history of image-guided radiotherapy. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[6]  B. Fallone,et al.  Radio frequency noise from clinical linear accelerators , 2009, Physics in medicine and biology.

[7]  B. Fallone,et al.  An integrated 6 MV linear accelerator model from electron gun to dose in a water tank. , 2010, Medical physics.

[8]  B. Fallone,et al.  First MR images obtained during megavoltage photon irradiation from a prototype integrated linac-MR system. , 2009, Medical physics.

[9]  B G Fallone,et al.  Magnetic decoupling of the linac in a low field biplanar linac-MR system. , 2010, Medical physics.

[10]  E. W. Shrigley Medical Physics , 1944, British medical journal.

[11]  K. Niayesh,et al.  Finite element methode analysis of performance of inductive saturable-core fault current limiter , 2011, 2011 1st International Conference on Electric Power Equipment - Switching Technology.

[12]  J G M Kok,et al.  Integrating a 1.5 T MRI scanner with a 6 MV accelerator: proof of concept , 2009, Physics in medicine and biology.

[13]  L. Kochian Author to whom correspondence should be addressed , 2006 .

[14]  S P M Crijns,et al.  Proof of concept of MRI-guided tracked radiation delivery: tracking one-dimensional motion , 2012, Physics in medicine and biology.

[15]  Hans-Peter Meinzer,et al.  Quantification of lung tumor volume and rotation at 3D dynamic parallel MR imaging with view sharing: preliminary results. , 2006, Radiology.

[16]  Ben Burkea Radiation induced current in the RF coils of integrated linac-MR systems : The effect of buildup and magnetic field , 2012 .

[17]  B G Fallone,et al.  Magnetic shielding investigation for a 6 MV in-line linac within the parallel configuration of a linac-MR system. , 2012, Medical physics.

[18]  Dirk Verellen,et al.  Innovations in image-guided radiotherapy , 2008, Nature Reviews Cancer.

[19]  Steve B. Jiang,et al.  Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation. , 2002, Physics in medicine and biology.

[20]  Satyapal Rathee,et al.  First demonstration of intrafractional tumor-tracked irradiation using 2D phantom MR images on a prototype linac-MR. , 2013, Medical physics.

[21]  Satyapal Rathee,et al.  TU‐C‐M100F‐01: Development of a Linac‐MRI System for Real‐Time ART , 2007 .

[22]  Rebecca Fahrig,et al.  A study of the effect of in-line and perpendicular magnetic fields on beam characteristics of electron guns in medical linear accelerators. , 2011, Medical physics.

[23]  James M. Galvin THE MULTILEAF COLLIMATOR-A COMPLETE GUIDE , 1999 .

[24]  Paul J Keall,et al.  An analysis of thoracic and abdominal tumour motion for stereotactic body radiotherapy patients , 2008, Physics in medicine and biology.