Image quality of a large-area dynamic flat detector: comparison with a state-of-the-art II/TV system

The purpose of this research is to establish the technical and clinical image quality of a 30 X 40 cm2 dynamic flat detector (FD) compared to state-of-the-art IITV technology. A Trixell detector for vascular and RF applications is designed for a mixed use of fluoroscopy as well as exposure series and a range of radiographic applications. An RF system has been built which comprises both the FD as well as an IITV detector. This system enables a direct comparison of technical image quality measurements and patient images under exactly the same X-ray conditions. Image quality measurements comprise Detective Quantum Efficiency including transfer characteristics, Modulation Transfer Function, Noise Power Spectrum, lag, Low Frequency Drop and residual signals. Observation tests, using Threshold Contrast Detail Detectability (TCDD) techniques, are performed in order to confirm the results of the technical measurements. Results show a DQE (f) of the flat detector that is higher compared to IITV and above all constant over a wide dose range, the IITV DQE (f) drops at higher dose range due to fixed structure. Furthermore the Low Frequency Drop is substantially smaller in the FD-based system. The TCDD subjective tests show improved system performance in favor of the FD system.

[1]  Paul R. Granfors Performance characteristics of an amorphous silicon flat-panel x-ray imaging detector , 1999, Medical Imaging.

[2]  Norbert Jung,et al.  Technical and clinical results of an experimental flat dynamic (digital) x-ray image detector (FDXD) system with real-time corrections , 1998, Medical Imaging.

[3]  A. Cowen,et al.  Technical report: Initial experiences with an experimental solid-state universal digital X-ray image detector. , 1998, Clinical Radiology.

[4]  J A Rowlands,et al.  Origins of flare in x-ray image intensifiers. , 1990, Medical physics.

[5]  A R Cowen,et al.  A comprehensive physical image quality evaluation of a selenium based digital x-ray imaging system for thorax radiography. , 1998, Medical physics.

[6]  Falko Busse,et al.  Methodology to measure fundamental performance parameters of x-ray detectors , 2001, SPIE Medical Imaging.

[7]  Falko Busse,et al.  Technical and clinical assessments of an experimental flat dynamic x-ray image detector system , 1999, Medical Imaging.

[8]  Norbert Conrads,et al.  Fluoroscopic x-ray imaging with amorphous silicon thin-film arrays , 1994, Medical Imaging.

[9]  Shuanghe Shi,et al.  Clinical comparison of a new digital x-ray detector system with a conventional screen film system , 1999, Medical Imaging.

[10]  Thierry Ducourant,et al.  New CsI/a-Si 17" x 17" x-ray flat-panel detector provides superior detectivity and immediate direct digital output for general radiography systems , 1998, Medical Imaging.

[11]  Ralf Schmidt,et al.  Correction of amplifier nonlinearity, offset, gain, temporal artifacts, and defects for flat-panel digital imaging devices , 2002, SPIE Medical Imaging.