In Vitro hydrodynamic evaluation of prosthetic polymer heart valves in steady flow

Hydrodynamic comparison of two polymer valves with two mechanical valves is presented. The valves were perfused in a steady flow system, and comparisons between the valves were made on the transvalvular pressure distribution and drop, opening area, and leakage volume. Particular emphasis was placed on a slit-type bileaflet polymer valve which was newly designed and fabricated in our research group. The results showed that the functional characteristics of a slit-type bileaflet polymer valve compared favorably with that of mechanical valves. This valve may be a viable and inexpensive alternative, especially for short-term use in TAH or VAD systems.